Open Access 23-05-2016 | Rhythm Puzzle Answer

A fatal combination of situs inversus, pregnancy and cardiac arrest treated with an automated external defibrillator

Auteurs: S. Calle, M. De Leeuw, N. Mpotos, P. Calle, B. De Turck

Gepubliceerd in: Netherlands Heart Journal | Uitgave 11/2016

share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail
insite
ZOEKEN

Answer

The AED delivered wrongful shocks, probably because the tachyarrhythmia with QRS complexes of low amplitude was considered to be ventricular fibrillation. Evidence for this comes from the ECG strips from the first AED rhythm analysis, clearly showing QRS complexes of a similar shape and amplitude, but at a lower rate. Furthermore, these small QRS complexes were continuously found from the start of the resuscitation attempt until the fourth shock (data not shown).
This case illustrates that the specificity of shock/no-shock decisions by the AEDs is not 100 % [1, 2]. Cardiologists should also be aware that AED sensitivity figures are only within a 90–95 % range, and that external artefacts during the analysis process (such as chest compressions) decrease the accuracy significantly. This implies that shock/no-shock decisions should be scrutinised before being used for diagnostic/therapeutic decision making and prognostication.
Remarkably, no cause of death could be detected on autopsy. Therefore, further research on the cardiac arrest risk stratification in dextrocardia in general, and during pregnancy in particular, is needed.
A third lesson concerns the low amplitude of QRS complexes detected by the AED electrodes placed in the conventional sternal-apical position [3]. Therefore, it seems reasonable in known dextrocardia cases to change to bi-axillary electrodes placement or a mirror-like approach (i. e. placement to the left of the sternum and in the right mid-axillary line). The reduced QRS amplitude in this dextrocardia case also argues against the use of the above-mentioned mirror-like approach in a standard cardiac arrest patient with a medical device implanted below the right clavicle.
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail

Onze productaanbevelingen

Netherlands Heart Journal

Het Netherlands Heart Journal wordt uitgegeven in samenwerking met de Nederlandse Vereniging voor Cardiologie en de Nederlandse Hartstichting. Het tijdschrift is Engelstalig en wordt gratis beschikbaa ...

Literatuur
1.
go back to reference Calle PA, Mpotos N, Calle SP, Monsieurs KG. Inaccurate treatment decisions of automated external defibrillators used by emergency medical services personnel: incidence, cause and impact on outcome. Resuscitation. 2015;88:68–74.CrossRefPubMed Calle PA, Mpotos N, Calle SP, Monsieurs KG. Inaccurate treatment decisions of automated external defibrillators used by emergency medical services personnel: incidence, cause and impact on outcome. Resuscitation. 2015;88:68–74.CrossRefPubMed
2.
go back to reference Calle PA, Paepe P De, Van Sassenbroeck D, Monsieurs KG. External artifacts by advanced life support providers misleading automated external defibrillators. Resuscitation. 2008;79:482–89.CrossRefPubMed Calle PA, Paepe P De, Van Sassenbroeck D, Monsieurs KG. External artifacts by advanced life support providers misleading automated external defibrillators. Resuscitation. 2008;79:482–89.CrossRefPubMed
Metagegevens
Titel
A fatal combination of situs inversus, pregnancy and cardiac arrest treated with an automated external defibrillator
Auteurs
S. Calle
M. De Leeuw
N. Mpotos
P. Calle
B. De Turck
Publicatiedatum
23-05-2016
Uitgeverij
Bohn Stafleu van Loghum
Gepubliceerd in
Netherlands Heart Journal / Uitgave 11/2016
Print ISSN: 1568-5888
Elektronisch ISSN: 1876-6250
DOI
https://doi.org/10.1007/s12471-016-0852-4

Andere artikelen Uitgave 11/2016

Netherlands Heart Journal 11/2016 Naar de uitgave