Skip to main content
Top
Gepubliceerd in: Netherlands Heart Journal 7-8/2018

Open Access 12-03-2018 | Heart Beat

Inappropriate shock caused by P wave oversensing in an entirely subcutaneous ICD

Auteurs: B. A. Mulder, A. H. Maass, Y. Blaauw

Gepubliceerd in: Netherlands Heart Journal | Uitgave 7-8/2018

share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail
insite
ZOEKEN
A 28-year-old woman with familial hypertrophic cardiomyopathy caused by a mutation in the gene encoding MYH7 presented with an implantable cardioverter defibrillator (ICD) shock to the emergency department. In the past she suffered from non-sustained ventricular tachycardias as well as atrial flutter but she had not experienced an ICD shock before. She cycled home and felt dyspnoeic and shortly thereafter experienced a bang and intense burning sensation retrosternally. ICD interrogation revealed an inappropriate shock because of P wave oversensing. Fig. 1a shows her electrocardiogram with large P waves due to atrial enlargement. Fig. 1b shows inappropriate sensing with detection of both QRS complexes and P waves (arrows indicate sensed P waves). This triggered the ICD to deliver a shock as the calculated frequency exceeded the conditional shock zone. To prevent future shocks, the vector was manually selected and the SMART Pass filter was turned on. During exercise testing appropriate QRS sensing was observed up to 130 beats per minute.
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail

Onze productaanbevelingen

Netherlands Heart Journal

Het Netherlands Heart Journal wordt uitgegeven in samenwerking met de Nederlandse Vereniging voor Cardiologie en de Nederlandse Hartstichting. Het tijdschrift is Engelstalig en wordt gratis beschikbaa ...

Metagegevens
Titel
Inappropriate shock caused by P wave oversensing in an entirely subcutaneous ICD
Auteurs
B. A. Mulder
A. H. Maass
Y. Blaauw
Publicatiedatum
12-03-2018
Uitgeverij
Bohn Stafleu van Loghum
Gepubliceerd in
Netherlands Heart Journal / Uitgave 7-8/2018
Print ISSN: 1568-5888
Elektronisch ISSN: 1876-6250
DOI
https://doi.org/10.1007/s12471-018-1099-z

Andere artikelen Uitgave 7-8/2018

Netherlands Heart Journal 7-8/2018 Naar de uitgave