Open Access 10-10-2023 | Editor’s Comment

Decoding fractional flow reserve/instantaneous wave-free ratio discordance: is flow the answer?

Auteurs: Lokien X. van Nunen, Peter Damman

Gepubliceerd in: Netherlands Heart Journal | Uitgave 11/2023

share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail
insite
ZOEKEN
That functional testing to assess the haemodynamic significance of a particular coronary artery stenosis is superior to solely visual assessment is indisputable when making decisions regarding revascularisation. In past decades, both fractional flow reserve (FFR) and coronary flow reserve (CFR) have emerged as two prominent methods to assess haemodynamic significance. CFR is defined as the ratio between maximum and resting coronary flow and is measured using intracoronary Doppler or bolus thermodilution. While CFR has proven prognostic implications [1], its use is limited by the absence of a universally accepted normal value, technical challenges in obtaining good and reproducible measurements, and its dependence on stable heart rate and blood pressure. Finally, CFR is a reflection of flow in the entire coronary circulation, including both epicardial coronary arteries and microcirculation. FFR is defined as the ratio of distal to proximal coronary pressure under conditions of maximum hyperaemia. Therefore, assessing resting flow with its variable haemodynamics is not necessary. In addition, FFR purely reflects flow limitation in the epicardial coronary artery. Besides FFR, several non-hyperaemic pressure ratios have since been introduced in an attempt to abandon the need for pharmacologically induced hyperaemia, including the instantaneous wave-free ratio (iFR). We know that discordance between FFR and iFR values occurs in up to 20% of cases [2, 3].
In this light, we read with interest the article by Stegehuis et al. published in this issue of the Netherlands Heart Journal [4]. We commend the authors on this deep dive into coronary physiology, attempting to further explain discordant FFR and iFR cases using different methods to assess coronary flow and stenosis resistance in a post hoc analysis of the IDEAL registry and DEFINE-FLOW study [5, 6]. The authors found that iFR and FFR were discordant in 15% (n = 97) of all cases. In this discordant subgroup, hyperaemic flow velocity and coronary flow reserve measurements showed iFR to be more in line with intracoronary flow evaluation by Doppler wire compared to FFR. Therefore, the authors conclude iFR should be the preferred index to guide revascularisation in the case of FFR/iFR discordance.
The results presented in this study should be interpreted with caution. The statistical analysis of the different Doppler wire measurements was performed in all patients, including the concordant positive and negative FFR/iFR groups. The main results and their statistical significance seem to be largely attributable to the differences between the concordant FFR+(abnormal)/iFR+ and FFR−(normal)/iFR− groups. Nonetheless, the results are almost identical to those in previous reports [2, 3].
Some additional considerations have to be mentioned regarding the complex subject of discordant physiological results. First is the role of the microcirculation. In the FFR−/iFR+ group, the reasoning offered for discordance is a blunted hyperaemic response, caused by old age and the higher prevalence of diabetes in this subgroup. However, both age and diabetes are also well-established risk factors for microvascular dysfunction and diminished coronary autoregulation, potentially leading to a lower hyperaemic response and increased resting flow. This is corroborated by the low CFR found in this study. These changes result in a lower maximum achievable flow down a coronary artery, and thus lower pressure gradient and higher FFR across a particular stenosis. The higher FFR reflects a lower potential gain in maximal coronary flow if that particular stenosis was treated. Earlier studies have shown that in the presence of microvascular dysfunction FFR remains accurate for detecting lesions responsible for myocardial ischaemia [7].
Regarding FFR+/iFR− discordance, the authors propose that a supra-normal hyperaemic response in coronary flow leads to the low FFR/high iFR discordance. The normal CFR (mean 2.4) corroborates the negative iFR in these cases. In the literature, this FFR+/iFR− discordance is most prevalent in left main or proximal left anterior descending coronary arteries, possibly due to a greater increase from baseline to hyperaemic flow owing to the large amount of subtended myocardium [8]. Unfortunately, we are not informed with regard to the exact location of the discordant cases in this study. Furthermore, it should be noted that the original DEFINE-FLOW included in this analysis found that, in patients with abnormal FFR and preserved CFR, there was a significantly higher rate of target vessel failure with medical therapy compared to patients with normal FFR and CFR [6].
Overall, this study presents valuable insights in coronary flow measurements in FFR/iFR concordant and discordant groups. It emphasises the role of the microcirculation and assessment of coronary flow in the functional assessment of epicardial disease. Selective measurements of microvascular function (index of microcirculatory resistance/hyperaemic microcirculatory resistance, microvascular resistance reserve, absolute flow measurements) are important to further explain if (part of) the discordance is attributable to microvascular dysfunction.
Furthermore, FFR/iFR discordance is common. We emphasise the role of location and extent of disease in the assessment of ischaemic heart disease, and its effects on epicardial and microvascular haemodynamics to best treat patients’ complaints and improve prognosis.
Until further evidence is available on clinical outcomes after iFR versus FFR-based revascularisation with discordant values, we believe this decision should be based on the patient’s anginal complaints, complexity of the revascularisation procedure versus pharmacological options, operator discretion taking the limitation of binary cut-offs into account, and the clinical setting of acute or chronic coronary syndrome.

Conflict of interest

L.X. van Nunen declares that he has no competing interests. P. Damman received research grants and lecture fees from Philips and Abbott, and consultancy fees from Philips.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.
share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail

Onze productaanbevelingen

Netherlands Heart Journal

Het Netherlands Heart Journal wordt uitgegeven in samenwerking met de Nederlandse Vereniging voor Cardiologie en de Nederlandse Hartstichting. Het tijdschrift is Engelstalig en wordt gratis beschikbaa ...

Literatuur
1.
go back to reference Murthy VL, Naya M, Foster CR, et al. Association between coronary vascular dysfunction and cardiac mortality in patients with and without diabetes mellitus. Circulation. 2012;126:1858–68.CrossRefPubMedPubMedCentral Murthy VL, Naya M, Foster CR, et al. Association between coronary vascular dysfunction and cardiac mortality in patients with and without diabetes mellitus. Circulation. 2012;126:1858–68.CrossRefPubMedPubMedCentral
2.
go back to reference Cook CM, Jeremias A, Petraco R, et al. Fractional flow reserve/instantaneous wave-free ratio discordance in angiographically intermediate coronary stenoses: an analysis using Doppler-derived coronary flow measurements. JACC Cardiovasc Interv. 2017;10:2514–24.CrossRefPubMedPubMedCentral Cook CM, Jeremias A, Petraco R, et al. Fractional flow reserve/instantaneous wave-free ratio discordance in angiographically intermediate coronary stenoses: an analysis using Doppler-derived coronary flow measurements. JACC Cardiovasc Interv. 2017;10:2514–24.CrossRefPubMedPubMedCentral
3.
go back to reference Jeremias A, Maehara A, Généreux P, et al. Multicenter core laboratory comparison of the instantaneous wave-free ratio and resting Pd/Pa with fractional flow reserve: the RESOLVE study. J Am Coll Cardiol. 2014;63:1253–61.CrossRefPubMed Jeremias A, Maehara A, Généreux P, et al. Multicenter core laboratory comparison of the instantaneous wave-free ratio and resting Pd/Pa with fractional flow reserve: the RESOLVE study. J Am Coll Cardiol. 2014;63:1253–61.CrossRefPubMed
5.
go back to reference Nijjer SS, de Waard GA, Sen S, et al. Coronary pressure and flow relationships in humans: phasic analysis of normal and pathological vessels and the implications for stenosis assessment: a report from the Iberian-Dutch-English (IDEAL) collaborators. Eur Heart J. 2016;37:2069–80.CrossRefPubMed Nijjer SS, de Waard GA, Sen S, et al. Coronary pressure and flow relationships in humans: phasic analysis of normal and pathological vessels and the implications for stenosis assessment: a report from the Iberian-Dutch-English (IDEAL) collaborators. Eur Heart J. 2016;37:2069–80.CrossRefPubMed
6.
go back to reference Johnson NP, Matsuo H, Nakayama M, et al. Combined pressure and flow measurements to guide treatment of coronary stenoses. JACC Cardiovasc Interv. 2021;14:1904–13.CrossRefPubMed Johnson NP, Matsuo H, Nakayama M, et al. Combined pressure and flow measurements to guide treatment of coronary stenoses. JACC Cardiovasc Interv. 2021;14:1904–13.CrossRefPubMed
7.
go back to reference De Bruyne B, Pijls NH, Bartunek J, et al. Fractional flow reserve in patients with prior myocardial infarction. Circulation. 2001;104:157–62.CrossRefPubMed De Bruyne B, Pijls NH, Bartunek J, et al. Fractional flow reserve in patients with prior myocardial infarction. Circulation. 2001;104:157–62.CrossRefPubMed
8.
go back to reference CONTRAST Study Investigators, Kobayashi Y, Johnson NP, Berry C, et al. The influence of lesion location on the diagnostic accuracy of adenosine-free coronary pressure wire measurements. JACC Cardiovasc Interv. 2016;9:2390–9.CrossRef CONTRAST Study Investigators, Kobayashi Y, Johnson NP, Berry C, et al. The influence of lesion location on the diagnostic accuracy of adenosine-free coronary pressure wire measurements. JACC Cardiovasc Interv. 2016;9:2390–9.CrossRef
Metagegevens
Titel
Decoding fractional flow reserve/instantaneous wave-free ratio discordance: is flow the answer?
Auteurs
Lokien X. van Nunen
Peter Damman
Publicatiedatum
10-10-2023
Uitgeverij
Bohn Stafleu van Loghum
Gepubliceerd in
Netherlands Heart Journal / Uitgave 11/2023
Print ISSN: 1568-5888
Elektronisch ISSN: 1876-6250
DOI
https://doi.org/10.1007/s12471-023-01822-y

Andere artikelen Uitgave 11/2023

Netherlands Heart Journal 11/2023 Naar de uitgave