Gepubliceerd in: Netherlands Heart Journal 6/2010

01-06-2010 | Case report

Ventricular asystole due to inhibition of ventricular backup pacing by a dual chamber pacemaker: possible pitfalls of default settings

Auteurs: F. M. van de Sandt, R. Jansen, G. P. Kimman, J. H. Ruiter

Gepubliceerd in: Netherlands Heart Journal | Uitgave 6/2010

Log in om toegang te krijgen
share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail

Abstract

A 62-year-old patient presented with dizzy spells after her dual chamber pacemaker (Medtronic Enrhythm P1501DR), implanted for complete AV block, had been reprogrammed to deliver antitachycardia therapy (ATP) for paroxysmal atrial tachycardia. Her symptoms were caused by inhibition of ventricular backup pacing during ATP, leading to ventricular asystoles. Inhibition was the result of premature ventricular beats occurring prior to ATP: when ventricular backup pacing is left in the default setting, this pacemaker withholds backup pacing if any of the four preceding events is a sensed event. This case illustrates the possibly hazardous effects of default pacemaker settings, especially in pacemaker-dependent patients. (Neth Heart J 2010;18:323-6.)
Literatuur
1.
go back to reference Medtronic EnrhythmTM P1501DR Reference Manual. Medtronic EnrhythmTM P1501DR Reference Manual.
2.
go back to reference Ruiter JH, Barrett MJ, Weteling L, Jansen R. Inhibition of a dual chamber pacemaker by oversensing of far-field atrial depolarisation in a patient after His-bundle ablation. Neth Heart J. 2004;12:543-6. Ruiter JH, Barrett MJ, Weteling L, Jansen R. Inhibition of a dual chamber pacemaker by oversensing of far-field atrial depolarisation in a patient after His-bundle ablation. Neth Heart J. 2004;12:543-6.
3.
go back to reference Herweg B, Barold SS. Ventricular asystole during atrial antitachycardia pacing by implanted cardioverter-defibrillator. Pacing Clin Electrophysiol. 2005;28:432-3. Herweg B, Barold SS. Ventricular asystole during atrial antitachycardia pacing by implanted cardioverter-defibrillator. Pacing Clin Electrophysiol. 2005;28:432-3.
4.
go back to reference Navarrete A. Failure to pace: pacemaker malfunction? J Cardiovasc Electrophysiol. 2008;19:100-1. Navarrete A. Failure to pace: pacemaker malfunction? J Cardiovasc Electrophysiol. 2008;19:100-1.
Metagegevens
Titel
Ventricular asystole due to inhibition of ventricular backup pacing by a dual chamber pacemaker: possible pitfalls of default settings
Auteurs
F. M. van de Sandt
R. Jansen
G. P. Kimman
J. H. Ruiter
Publicatiedatum
01-06-2010
Uitgeverij
Bohn Stafleu van Loghum
Gepubliceerd in
Netherlands Heart Journal / Uitgave 6/2010
Print ISSN: 1568-5888
Elektronisch ISSN: 1876-6250
DOI
https://doi.org/10.1007/BF03091784

Andere artikelen Uitgave 6/2010

Netherlands Heart Journal 6/2010 Naar de uitgave

Interuniversity Cardiology Institute of the Netherlands

Einthoven Dissertation Prize 2010