Gepubliceerd in: Netherlands Heart Journal 2/2010

01-02-2010 | original article

An in vivo comparison of radiofrequency cardiac lesions formed by standard and magnetically steered 4 mm tip catheters

Gepubliceerd in: Netherlands Heart Journal | Uitgave 2/2010

Log in om toegang te krijgen
share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail

Abstract

Background. In vivo comparison of cardiac radiofrequency ablation lesions between standard and magnetically steered 4 mm tip catheters has never been reported.
Methods. High and low right atrium (RA) free wall, isthmus, right ventricle (RV) free wall and outflow tract lesions were studied macroscopically and microscopically five days after lesion formation in seven pigs. Shape, size, thrombus formation, and ablation parameters were compared. The effect of minimal, medium and high wall contact was assessed by a contact measurement utility for magnetic catheters.
Results. All 14 RA free wall lesions were transmural with a similar epicardial and endocardial surface area. In the RV, the epicardial area usually appeared to be smaller than the endocardial area with standard catheters. Isthmus lesions were difficult to assess transmurality. There was no difference in endocardial area: standard 39 mm2 (range 16 to 82 mm2) vs. magnetic 36 mm2 (range 23 to 111 mm2). If the catheter tip was perpendicular to the tissue, magnetic lesions were more often round or oval, while standard lesions were more often elongated (p<0.05). When the catheter tip was parallel to tissue, lesions always tended to be elongated. Microscopic characteristics were similar. The contact utility was not useful. Average impedance (p<0.0001) and energy delivered (p<0.05) were less with magnetic catheters.
Conclusion. Lesions from magnetically steered catheters are transmural of similar size, but with less variability than standard catheter lesions when the tip is perpendicular to the tissue. Magnetic lesions are associated with lower impedance and energy delivery. This suggests a more stable tip-to-tissue contact. (Neth Heart J 2010;18:66–71.)
Literatuur
1.
go back to reference Stagegaard N, Petersen HH, Chen X, Svendsen JH. Indication of the radiofrequency induced lesion size by pre-ablation measurements. Europace. 2005;7:525–34. Stagegaard N, Petersen HH, Chen X, Svendsen JH. Indication of the radiofrequency induced lesion size by pre-ablation measurements. Europace. 2005;7:525–34.
2.
go back to reference Kongsgaard E, Steen T, Jensen O, Aass, H, Amlie JP. Temperature guided radiofrequency catheter ablation of myocardium: comparison of catheter tip and tissue temperatures in vitro. Pacing Clin Electrophysiol. 1997;20:1252–60. Kongsgaard E, Steen T, Jensen O, Aass, H, Amlie JP. Temperature guided radiofrequency catheter ablation of myocardium: comparison of catheter tip and tissue temperatures in vitro. Pacing Clin Electrophysiol. 1997;20:1252–60.
3.
go back to reference Haines DE, Watson DD. Tissue heating during radiofrequency catheter ablation: a hermodynamic model and observations in isolated perfused and superfused canine right ventricular free wall. Pacing Clin Electrophysiol. 1989;12:962–76. Haines DE, Watson DD. Tissue heating during radiofrequency catheter ablation: a hermodynamic model and observations in isolated perfused and superfused canine right ventricular free wall. Pacing Clin Electrophysiol. 1989;12:962–76.
4.
go back to reference Petersen HH, Chen X, Pietersen A, Svendsen JH, Haunso S. Tissue temperatures and lesion size during irrigated tip catheter radiofrequency ablation: an in vitro comparison of temperaturecontrolled irrigated tip ablation, power-controlled irrigated tip ablation, and standard temperature-controlled ablation. Pacing Clin Electrophysiol. 2000;23:8–17. Petersen HH, Chen X, Pietersen A, Svendsen JH, Haunso S. Tissue temperatures and lesion size during irrigated tip catheter radiofrequency ablation: an in vitro comparison of temperaturecontrolled irrigated tip ablation, power-controlled irrigated tip ablation, and standard temperature-controlled ablation. Pacing Clin Electrophysiol. 2000;23:8–17.
5.
go back to reference Nakagawa H, Yamanashi WS, Pitha JV, Arruda M, Wang X, Ohtomo K, et al. Comparison of in vivo tissue temperature profile and lesion geometry for radiofrequency ablation with a saline-irrigated electrode versus temperature control in a canine thigh muscle preparation. Circulation. 1995;91:2264–73. Nakagawa H, Yamanashi WS, Pitha JV, Arruda M, Wang X, Ohtomo K, et al. Comparison of in vivo tissue temperature profile and lesion geometry for radiofrequency ablation with a saline-irrigated electrode versus temperature control in a canine thigh muscle preparation. Circulation. 1995;91:2264–73.
6.
go back to reference Petersen HH, Chen X, Pietersen A, Svendsen JH, Haunso S. Temperature-controlled radiofrequency ablation of cardiac tissue: an in vitro study of the impact of electrode orientation, electrode tissue contact pressure and external convective cooling. J Interv Card Electrophysiol. 1999;3:257–62. Petersen HH, Chen X, Pietersen A, Svendsen JH, Haunso S. Temperature-controlled radiofrequency ablation of cardiac tissue: an in vitro study of the impact of electrode orientation, electrode tissue contact pressure and external convective cooling. J Interv Card Electrophysiol. 1999;3:257–62.
7.
go back to reference Thornton AS, Janse P, Theuns DA, Scholten MF, Jordaens LJ. Magnetic navigation in AV nodal re-entrant tachycardia study: early results of ablation with one- and three-magnet catheters. Europace. 2006;8:225–30. Thornton AS, Janse P, Theuns DA, Scholten MF, Jordaens LJ. Magnetic navigation in AV nodal re-entrant tachycardia study: early results of ablation with one- and three-magnet catheters. Europace. 2006;8:225–30.
8.
go back to reference Al-Ahmad A, Grossman JD, Wang PJ. Early experience with a computerized robotically controlled catheter system. J Interv Card Electrophysiol. 2005;12:199–202. Al-Ahmad A, Grossman JD, Wang PJ. Early experience with a computerized robotically controlled catheter system. J Interv Card Electrophysiol. 2005;12:199–202.
9.
go back to reference Saliba W, Cummings JE, Oh S, Zhang Y, Mazgalev TN, Schweikert RA, et al. Novel robotic catheter remote control system: feasibility and safety of transseptal puncture and endocardial catheter navigation. J Cardiovasc Electrophysiol. 2006;17:1102–5. Saliba W, Cummings JE, Oh S, Zhang Y, Mazgalev TN, Schweikert RA, et al. Novel robotic catheter remote control system: feasibility and safety of transseptal puncture and endocardial catheter navigation. J Cardiovasc Electrophysiol. 2006;17:1102–5.
10.
go back to reference Ernst S, Ouyang F, Linder C, Hertting K, Stahl F, Chun J, et al. Initial experience with remote catheter ablation using a novel magnetic navigation system: magnetic remote catheter ablation. Circulation. 2004;109:1472–5. Ernst S, Ouyang F, Linder C, Hertting K, Stahl F, Chun J, et al. Initial experience with remote catheter ablation using a novel magnetic navigation system: magnetic remote catheter ablation. Circulation. 2004;109:1472–5.
11.
go back to reference Faddis MN, Blume W, Finney J, Hall A, Rauch J, Sell J, et al. Novel, magnetically guided catheter for endocardial mapping and radiofrequency catheter ablation. Circulation. 2002;106:2980–5. Faddis MN, Blume W, Finney J, Hall A, Rauch J, Sell J, et al. Novel, magnetically guided catheter for endocardial mapping and radiofrequency catheter ablation. Circulation. 2002;106:2980–5.
12.
go back to reference Faddis MN, Chen J, Osborn J, Talcott M, Cain ME, Lindsay BD. Magnetic guidance system for cardiac electrophysiology: a prospective trial of safety and efficacy in humans. J Am Coll Cardiol. 2003;42:1952–8. Faddis MN, Chen J, Osborn J, Talcott M, Cain ME, Lindsay BD. Magnetic guidance system for cardiac electrophysiology: a prospective trial of safety and efficacy in humans. J Am Coll Cardiol. 2003;42:1952–8.
13.
go back to reference Kerzner R, Sanchez JM, Osborn JL, Talcott M, Cain ME, Lindsay BD. Radiofrequency ablation of atrioventricular nodal reentrant tachycardia using a novel magnetic guidance system compared with a conventional approach. Heart Rhythm.. 2006;3:261–7. Kerzner R, Sanchez JM, Osborn JL, Talcott M, Cain ME, Lindsay BD. Radiofrequency ablation of atrioventricular nodal reentrant tachycardia using a novel magnetic guidance system compared with a conventional approach. Heart Rhythm.. 2006;3:261–7.
14.
go back to reference Pappone C, Vicedomini G, Manguso F, Gugliotta F, Mazzone P, Gulletta S, et al. Robotic magnetic navigation for atrial fibrillation ablation. J Am Coll Cardiol. 2006;47:1390–400. Pappone C, Vicedomini G, Manguso F, Gugliotta F, Mazzone P, Gulletta S, et al. Robotic magnetic navigation for atrial fibrillation ablation. J Am Coll Cardiol. 2006;47:1390–400.
15.
go back to reference Thornton AS, Jordaens LJ. Remote magnetic navigation for mapping and ablating right ventricular outflow tract tachycardia. Heart Rhythm.. 2006;3:691–6. Thornton AS, Jordaens LJ. Remote magnetic navigation for mapping and ablating right ventricular outflow tract tachycardia. Heart Rhythm.. 2006;3:691–6.
16.
go back to reference Thornton AS, Rivero-Ayerza M, Knops P, Jordaens LJ. Magnetic navigation in left-sided AV reentrant tachycardias: preliminary results of a retrograde approach. J Cardiovasc Electrophysiol. 2007;18:467–72. Thornton AS, Rivero-Ayerza M, Knops P, Jordaens LJ. Magnetic navigation in left-sided AV reentrant tachycardias: preliminary results of a retrograde approach. J Cardiovasc Electrophysiol. 2007;18:467–72.
Metagegevens
Titel
An in vivo comparison of radiofrequency cardiac lesions formed by standard and magnetically steered 4 mm tip catheters
Publicatiedatum
01-02-2010
Gepubliceerd in
Netherlands Heart Journal / Uitgave 2/2010
Print ISSN: 1568-5888
Elektronisch ISSN: 1876-6250
DOI
https://doi.org/10.1007/BF03091740

Andere artikelen Uitgave 2/2010

Netherlands Heart Journal 2/2010 Naar de uitgave