Introduction
Refractory cardiogenic shock has traditionally been associated with high mortality rates [
1,
2]. With the evolvement of venoarterial extracorporeal life support (ECLS) in the latter part of the previous century, the arsenal of treatment options was significantly extended. Early disappointing results [
3], however, rendered physicians reluctant to apply ECLS in clinical practice.
With the development of more advanced devices incorporating novel centrifugal pumps and biocompatible membranes [
4], interest in circulatory ECLS was renewed. Initially, circulatory ECLS was primarily used in paediatric patients with postcardiotomy cardiogenic shock [
5]. Subsequently, its use grew to encompass similar indications in the adult population [
6]. The preference for ECLS [
7] was catalysed by findings from the IABP-SHOCK-II trial demonstrating an absence of mortality benefit of the intra-aortic balloon pump in the setting of cardiogenic shock [
8].
In the Netherlands, a number of centres apply ECLS for circulatory and respiratory indications. The University Medical Centre Utrecht (UMCU) has been using ECLS since 2007 in several set-ups, traditionally in the context of cardiogenic shock. Over 12 years, we have seen considerable changes in patient population, indications, complications and outcomes. Except for individual case reports [
9], these data have not yet been summarised.
In order to reflect on and to learn from these first 12 years of ECLS in our centre, we performed a retrospective cohort study to describe these characteristics and their evolution over time.
Methods
A cohort study was performed by retrospectively collecting data from all patients who were supported with venoarterial ECLS for circulatory indications between April 2007, when the first patient was supported, and December 2018 at the UMCU. A waiver was obtained from the medical ethics committee for the retrieval of anonymised data. Information about demographics, comorbidities, ECLS indications, set-up and mortality was collected. Registration of comorbidities occurred on the basis of written diagnoses in the medical chart and included previous presence of diabetes mellitus, chronic kidney disease, peripheral artery disease, transient ischaemic attack (TIA) or ischaemic cerebrovascular accident (ICVA), acute (on chronic) heart failure, coronary artery disease, or previous revascularisation via coronary artery bypass grafting or percutaneous coronary intervention.
Survival status and death causes were recorded at 30 days, hospital discharge, and 1 year after ECLS initiation and at the last time of contact. Patients were censored at each time point when they were alive.
Indications
ECLS was initiated in patients with circulatory shock, which was defined as systolic blood pressure < 90 mm Hg and/or evidence of insufficient organ perfusion refractory to inotropic and vasopressor support. Circulatory indications included postcardiotomy cardiogenic shock (within 24 h after cardiac surgery), shock due to myocardial infarction, refractory ventricular fibrillation/tachycardia, acute on chronic heart failure, myocarditis or other causes such as pulmonary embolism, septic cardiomyopathy or left ventricular assist device (LVAD) failure. Finally, ECLS was used in patients after lung transplantation as elective extended measure to prevent pulmonary congestion in those with pre-existing pulmonary hypertension or in case of severe right ventricular failure.
The onset and conclusion of an ECLS run were defined by cannulation and decannulation, respectively. A new run was marked by the start of recannulation when a patient had been decannulated or when cannula positions were changed because of a change in ECLS indication (for example, a patient would have received venovenous ECLS for acute respiratory distress syndrome, but cannulas were changed to a venoarterial or venoarterial-venous configuration because of newly developed septic or cardiogenic shock) or due to complications resulting from the previous set-up (e.g. change of cannula position from femoral vein to jugular vein because of thrombosis). This definition was used to maximise the detection of complications and to learn from different ECLS strategies that were used in our clinical practice.
The ECLS systems used at the UMCU encompassed a variety of circuit set-ups, all including an extracorporeal centrifugal pump with or without an oxygenator, heater and cannulas. In our clinical practice, Permanent Life Support (Getinge Maquet, Rastatt, Germany), CentriMag (Levitronix-Thoratec-Abbott, Abbott Park, IL, USA) and Cardiohelp (Getinge Maquet) were used. Central cannulation was performed by a cardiothoracic surgeon in the operating theatre, whereas peripheral cannulation was primarily utilised by an interventional cardiologist and/or intensivist in the catheterisation laboratory.
Cannulation of the central aorta was performed using a 20–24–French (F) Elongated One-Piece Arterial Cannula (Medtronic, Minneapolis, MN, USA) or a 20‑F Jostra cannula in combination with an 8‑mm Dacron prosthesis. For right atrial cannulation, a 34- or 36‑F cannula was used; for peripheral cannulation purposes, a 21–25–F Maquet HLS multistage drainage cannula and a 15–19–F Bio-Medicus (Medtronic) arterial single stage cannula were used. An 8‑F Arrow sheath was inserted for selective distal perfusion in case of angiographic obliteration of peripheral arterial flow or signs of limb ischaemia. From 2018 on forward, a distal cannula was also put in place for prevention of ischaemia.
As an adjunct to ECLS, LV unloading strategies were applied when complications of high afterload were seen or anticipated upon ECLS initiation. These included: (a) presence of pulmonary oedema, (b) LV distention and/or virtual absence of LV ejection, (c) pulmonary capillary wedge pressures > 15 mm Hg, or (d) the anticipation of a high risk for development of pulmonary oedema during the ECLS run. An intra-aortic balloon pump was the main technique of choice. When theoretical considerations prior to placement led us to believe that venting capacity of the intra-aortic balloon pump was probably insufficient [
10], the left atrium or ventricle was directly vented.
Statistical analyses
The population was divided into six groups based on the year of presentation, with each group comprising a 2-year time frame. Sequential Organ Failure Assessment (SOFA) scores were calculated by summarising the scores from six subdomains (central nervous system, circulatory, respiratory, coagulation, renal and liver). The central nervous system domain score was missing in some cases due to usage of sedatives just before ECLS initiation. These missing values were imputed with the nearest value, a common technique used for imputation of missing SOFA scores at baseline [
11]. Survival After Veno-arterial Extracorporeal membrane oxygenation (SAVE) scores were calculated based on the validated formula published by Schmidt et al. [
12]. Acute Physiology and Chronic Health Evaluation (APACHE) IV scores were similarly calculated [
13].
Baseline and ECLS characteristics are presented as mean with standard deviation (SD) or as median with interquartile range (IQR), depending on the variable distribution. Baseline characteristics and survival were compared on a patient level and ECLS characteristics on the level of the ECLS run. Trends and differences across groups were compared using linear regression analyses and chi-squared tests for continuous and categorised data, respectively.
To analyse the association between year of first ECLS exposure and mortality, Cox proportional hazard models were constructed. A crude model was composed of year of ECLS initiation and 30-day, 1‑year or overall mortality. In a second model, age, sex and comorbidities were added as covariates. In a third and fourth model, SOFA scores and APACHE IV scores were added, respectively. In separate models, SAVE scores were added as covariate.
In order to compare the observed number of deaths during hospital stay with expected mortality figures as based on the SAVE sore, we computed standardised mortality ratios (SMRs). SMRs are calculated by dividing the observed in-hospital death rate by the hypothetical mortality rate as predicted by the SAVE score. The result is a ratio with 95% confidence interval for each year category.
P-values < 0.05 and confidence intervals not including 1 were considered statistically significant. Analyses were performed with RStudio: Integrated Development for R (RStudio, Inc., Boston, MA, VS).
Discussion
Our study, which is based on one of the largest ECLS cohorts in the Netherlands [
14‐
17], describes changes in case mix, indications, set-ups and outcomes over a 12-year period. Within this time frame, ECLS indications broadened and circuit set-ups shifted from central to mainly peripheral approaches. Meanwhile, the rate of successful weaning from ECLS increased and mortality rates diminished. This latter observation could potentially be explained by an altered, and possibly, improved patient selection.
Our data reflect the evolution of ECLS in our centre over time. At first, ECLS was primarily used for refractory postcardiotomy cardiogenic shock. During this phase, patients were centrally cannulated and mortality was high. After this initial phase, ECLS was increasingly applied as bridge to LVAD implantation or heart transplantation. After recognising that a considerable number of patients could be weaned from temporary support, the technique was increasingly deployed as bridge to recovery. With our overall weaning success rate (42%) being comparable to that of other centres, success rates increased up to 53% in the last year of inclusion. During this last phase, ECLS set-ups were refined with a rising use of distal cannulas and LV unloading devices.
The growth in application of LV unloading devices over time (2% to 23%) likely represents an increased recognition of: (a) the relatively high incidence of LV distention [
18], and (b) its potential negative impact on the recovery of the left ventricle and on patient recovery [
19,
20]. Despite an increase in application of LV unloading at our centre, the highest percentages in 2017–2018 still seem lower than that of other centres [
18]. This discrepancy could be explained by different thresholds for placement of an LV unloading device. In certain centres, insertion of ECLS is routinely accompanied by placement of an intra-aortic balloon pump [
21], while others only use the combined set-up in selected patients who developed pulmonary oedema [
22]. Optimal indications and timing for LV unloading remain for now unknown.
Similar to observations in other ECLS-treated populations [
12], and specifically in those after acute myocardial infarction [
23] or cardiac arrest [
24], we noted a statistically significant decrease in mortality rates over time. Because adjustment for SOFA, APACHE IV and SAVE scores largely abolished the observed association between year of inclusion and mortality, this improvement in mortality rates may reflect better patient selection rather than just evolvement of ECLS management. Nonsignificant changes attributable to better ECLS care may, however, not be excluded, especially since more patients could be weaned from support.
Comparisons between observed and expected mortality rates, as predicted by the SAVE score, suggested improved survival after 2012 as compared with international references. Although it is tempting to attribute this discrepancy solely to good clinical practice at our centre, it should be interpreted with caution for several reasons. First, the SAVE score was designed with data from 2003–2013. In the years thereafter, knowledge likely increased and possibly led to better management. Second, as ECLS has traditionally been applied in the context of LVAD and heart transplant care at the UMCU, patient selection could have biased direct comparisons.
Limitations
Although our analyses were based on one of the largest ECLS populations in the Netherlands, providing a unique historical overview of ECLS evolution, some limitations apply. First, the association between time of inclusion and overall mortality could have been subject to bias. Follow-up was naturally longer when patients were included earlier and, thereby, the chance of dying could be higher. Nevertheless, in-hospital, 30-day and 1‑year mortality significantly improved over time as well. Furthermore, it is important to consider that our study was performed in a retrospective way. Nevertheless, the primary goal of our study was to describe evolutions throughout history, for which our cohort provided excellent data.
Conclusion
ECLS is increasingly utilised at the UMCU for circulatory support. Over the course of 12 years, ECLS indications broadened and circuit set-ups shifted from centrally to peripherally cannulated approaches. Meanwhile, the weaning success rate from ECLS increased and mortality rates diminished. This latter observation could potentially be explained by altered, and possibly, improved patient selection.