Treatment efficacy
RDN is at present one of the most widely studied invasive approaches for the treatment of hypertension. With the first generation of RDN catheters, varying effect magnitudes of BP reduction in patients who underwent RDN were reported [
31‐
33]. Non-standardised medical treatment, changes in antihypertensive medication throughout the course of the trials and suboptimal denervation procedures were soon recognised as major factors complicating device-based antihypertensive therapy research [
34]. Addressing these limitations in improved study protocols and new denervation techniques, five proof-of-principle RCTs proved the overall efficacy and safety of RDN in patients on and off antihypertensive medication [
35‐
40].
Two RCTs were performed in patients taken off antihypertensive drugs in a well-controlled setting to determine treatment efficacy of RDN in the absence of any antihypertensive drug effects [
35,
37,
39]. These studies enrolled patients with uncontrolled, mild to moderate hypertension and a low cardiovascular risk [
35,
37,
39]. The SPYRAL HTN-OFF MED trial (
n = 331) investigated the effect of radiofrequency (RF) RDN (using the Medtronic Symplicity Spyral multi-electrode catheter (Medtronic, Galway, Ireland)) whereas the RADIANCE HTN SOLO trial (
n = 146) evaluated the effect of ultrasound (US) RDN (using the Paradise Renal Denervation System (ReCor Medical, Palo Alto, CA, USA)) [
35,
37,
39]. Both RDN techniques proved efficacious in achieving a significant drop in sABP (−3.9 to −6.3 mm Hg), dABP (−2.6 to −4.4 mm Hg), sOBP (−6.5 to −7.7 mm Hg) and dOBP (−4.1 to −4.9 mm Hg) at two to three months compared with a sham-control arm [
35,
39]. In the RADIANCE HTN SOLO trial, standardised antihypertensive drug treatment was introduced after assessment of the primary endpoint at two months. At six months, the efficacy of RDN was confirmed as patients who underwent RDN had lower sABP (−4.3 mm Hg) and sOBP (−3.7 mm Hg) on a lower burden of antihypertensive drugs compared with control patients [
41].
Next, RDN was evaluated with a similar level of scrutiny in patients on antihypertensive drugs in three RCTs [
36,
38,
40]. The DENERHTN trial (
n = 106) evaluated the effect of RDN in patients not meeting BP targets despite the use a standardised, triple-pill antihypertensive drug regimen (indapamide, ramipril (or irbesartan) and amlodipine) [
36]. Patients were randomised to RDN (using the Medtronic Symplicity Flex uni-electrode RF catheter) plus standardised stepped antihypertensive treatment (SSAHT) or SSAHT alone [
36]. The SSAHT uptitration scheme consisted of spironolactone, bisoprolol, prazosin and rilmenidine, accordingly [
36]. RDN on top of SSAHT resulted in an additional reduction in sABP (−5.9 mm Hg) compared with SSAHT alone at six months [
36]. These findings were confirmed by the preliminary results of the SPYRAL HTN-ON MED (
n = 80) trial that investigated the effect of RF-RDN (using the Medtronic Symplicity Spyral multi-electrode catheter) in patients not meeting BP targets while on a non-standardised, stable regimen of one to three antihypertensive drugs [
38]. RDN proved to effectively lower sABP (−7.0 mm Hg), dABP (−4.3 mm Hg), sOBP (−6.6 mm Hg) and dOBP (−4.2 mm Hg) at six months as compared with sham-control [
38]. These results were confirmed in the RADIANCE TRIO trial which evaluated the efficacy of US RDN (using the Paradise Renal Denervation System) in patients on a standardised triple-pill antihypertensive drug regimen (amlodipine, valsartan and hydrochlorothiazide) [
40]. The study demonstrated a significant reduction in sABP (−4.5 mm Hg) and sOBP (−7.0 mm Hg) two months after RDN compared with sham-control, whereas dABP and dOBP did not differ between both groups [
40]. In contrast, no significant reduction in sABP post RDN compared with sham-control was observed in the REQUIRE trial which refrained from standardising antihypertensive therapy and adherence testing [
42]. When comparing US-RDN to RF-RDN treatment, the RADIOSOUND study demonstrated that US-RDN results in similar BP reductions as RF-RDN including any accessory renal arteries [
43].
Whereas significant mean BP reductions have been observed post RDN, the treatment effect in individual patients showed substantial heterogeneity, with approximately one out of three patients exhibiting no significant BP response to RDN [
35,
38,
44,
45]. Unfortunately, as of to date, consistent predictors of treatment response have not yet been identified [
28].
Whereas more pragmatic trials with more lenient entry criteria are needed, the global SYMPLICITY registry reported significant and sustained BP reductions in real-world patients [
46].
Treatment durability
Demonstration of durability of the BP response following interventional procedures is challenged by changes in medications, coexisting illness and patient behaviour (e.g. weight loss, exercise and diet). Animal studies suggest the potential of renal nerve regeneration over time [
47,
48]. However, renal hormone excretion was only partially restored [
47]. Whereas evidence on this phenomenon lacks in humans, registry studies confirmed a durable sABP-lowering effect of −8.0 and −20.9 mm Hg up until three and five years after RDN respectively [
46,
49]. In the RADIANCE HTN SOLO study, sOBP was lower in the RDN group (−5.9 vs. −4.3 mm Hg) while on less medication compared with the control group 12 months post randomisation [
50]. The 3‑year results of the SPYRAL HTN-ON MED study demonstrated a persistent reduction in sABP of −10.0 mm Hg post RDN as compared with sham-control, which could not be explained by differences in prescribed antihypertensive drug regimen or therapy adherence [
51].
Cost-effectiveness
Several studies on health decision modelling have reported modelled incremental cost-effectiveness ratios (ICER) for RDN ranging from € 1,474 to € 6,573 per quality-adjusted life year (QALY) [
56‐
58]. The only Dutch study investigating the cost-effectiveness of RDN demonstrated the lowest ICER from all studies (€ 1,474 per QALY) [
58]. The latter shows RDN is considered a cost-effective treatment for all common willingness-to-pay thresholds in the Netherlands [
59]. As all published evidence on cost-effectiveness on RDN is currently based on first generation trials, the cost-effectiveness analyses of second generation trials are eagerly awaited.
Patient preference
In patients who require antihypertensive treatment, there is a profound interest for non-pharmacological, invasive treatment options over taking drugs on a daily base. About 8.2% of all patients in the United States would be willing to trade two years off their lives to avoid taking any drugs for cardiovascular prevention [
60]. For RDN specifically, 28% of all drug-treated uncontrolled hypertensive German patients would prefer RDN over further intensification of drug therapy [
61]. In a recent multi-country (including the Netherlands) 15-day social media campaign recruiting hypertensive patients for a novel RDN trial, 12,000 individuals clicked on the advertisement which resulted in over 400 registrations for that particular trial [
62].
Ongoing studies
At present, several new studies investigating the safety and efficacy of different RDN technologies are ongoing. For radiofrequency RDN, the SPYRAL ON-MED study will focus on the effect of RDN on top of antihypertensive therapy (ClinicalTrials.gov Identifier: NCT02439775). With respect to ultrasound RDN, the RADIANCE II pivotal study will focus on the treatment effect in absence of antihypertensive medication (ClinicalTrials.gov Identifier: NCT03614260). In parallel, the TARGET BP I (ClinicalTrials.gov Identifier: NCT02910414) and TARGET BP OFF-MED (ClinicalTrials.gov Identifier: NCT03503773) trials currently investigate RDN using perivascular alcohol infusion in patients on and off antihypertensive medication respectively.
Topics to be addressed in future research
In the upcoming years, a shift from proof-of-concept trials to pragmatic real-world RDN studies might be expected. Registries including large numbers of patients will reveal more information about long-term efficacy and safety of the technology. To allow for poolability of long-term data from different studies, trials will have to be designed in a standardised fashion with respect to inclusion criteria and outcome measures. As such, the upcoming Hypertension Academic Research Consortium (HARC) statement will provide further guidance on the matter [
63]. In parallel, more insights into predictors of RDN success are required to facilitate adequate selection of patients who are most likely to benefit from treatment. Previous studies have identified nighttime sABP as well as its variability, 24-hour dABP, 24-hour heart rate, pulse wave velocity, central pulse pressure, gender and a history of diabetes mellitus as predictors of the treatment effect of RDN, but caution has to be applied as these predictors were mostly detected in retrospective post-hoc analyses [
64‐
69]. Finally, there is a growing interest in measuring and defining technical and procedural success. As such, renal artery nerve mapping has shown to be a safe and feasible technique [
70,
71]. This procedure allows for measuring the effect of renal nerve stimulation on BP pre-RDN, which was shown to be correlated to decrease in ABP afterwards [
72].
Evidence for treatment indications outside of hypertension
This current consensus statement focusses on RDN as a well-investigated, promising therapy for patients with hypertension. In parallel, the safety and efficacy of RDN for alternative indications and conditions associated with sympathetic overactivity, such as kidney failure, kidney-related pain syndromes, atrial fibrillation, ventricular arrhythmias, heart failure, insulin resistance, metabolic syndrome and OSAS, have been studied [
63]. The consortium believes more data is needed to decide on the role of RDN in the treatment of the diseases mentioned above and advises to refrain from RDN treatment in patients with these conditions outside well-controlled study settings.