Abstract

The clinical course of a patient with a left ventricular assist device is described. A total of 6 weeks after device insertion, the lactate dehydrogenase (LDH) level increased to 2801 U/l despite adding low-molecular-weight heparin to acenocoumarol and aspirin. Pump thrombosis was suspected but unconfirmed by computed tomography. Increased pump power requirement did not occur. Instituting unfractionated heparin caused a drop in the LDH level. After discontinuing heparin, the LDH levels rose to 5529 U/l whereupon pump replacement was performed. LDH levels, combined with clinical deterioration and right heart catheterisation, led to the diagnosis of pump thrombosis.

Introduction

Implantation of a permanent left ventricular assist device (LVAD) in a patient with cardiogenic shock must be avoided. Short-term external left ventricular support and appropriate pharmacological treatment can be a feasible alternative [1, 2]. Mechanical circulatory support by an implantable device is, however, the current therapy of choice for patients who deteriorate clinically while on the waiting list for a heart transplant [3]. Well-known complications after insertion of an LVAD are bleeding, right ventricular dysfunction, (driveline) infection, arrhythmias and pump thrombosis [4]. The incidence of pump thrombosis was reported in a range of 1.4–4 % in 2010 [5]. We report a case of early postoperative pump thrombosis.
A 46-year-old man with dilated cardiomyopathy and end-stage heart failure, despite optimal medical and implantable cardioverter defibrillator therapy, was accepted on the Eurotransplant waiting list for heart transplant on May 2013. In July 2013, due to clinical deterioration, he received—in a non-emergency setting—an LVAD type HeartMate II (Thoratec Corp. Pleasanton CA, USA) as a bridge to transplantation [Fig. 1, label 1]. Postoperative recovery was without complications, and the patient was discharged to a cardiac rehabilitation centre 19 days after insertion of the device. In all, 27 days after implantation, the level of lactate dehydrogenase (LDH) increased from 254 U/l (pre-VAD) to 704 and 1050 U/l [Fig. 1, label 2]. This raised the suspicion of pump thrombosis and low-molecular-weight heparin was added to International Normalized Ratio (INR)-guided (1.5–2.5) acenocoumarol and 100 mg of aspirin daily. However, the LDH levels further increased to 2801 U/l [Fig. 1, label 3]. The patient was hospitalised. Although computed tomography (CT) using contrast showed no signs of obstruction in the VAD circuit and LVAD power requirement did not increase [Fig. 1, label 3], unfractionated heparin (target Activated Partial Thromoplastin Time (APTT) level 40–60 s) was given intravenously. This led to a drop in the LDH level to 778 U/l [Fig. 1, label 4]. The patient was discharged, but within a few days he complained of fatigue, dyspnoea and reduced exercise tolerance. On re-admission, echocardiography showed an aortic valve opening at each contraction, which was observed more often than in previous echocardiographies. There were no signs of aortic valve regurgitation. Right ventricular function was impaired, indicated by a tricuspid annulus plane systolic excursion of 10 mm. A velocity of the tricuspid annular systolic motion s’RV of 5.3 cm/s was more impaired than earlier. Pump power requirement remained unaltered. The LDH level rose to 1967 U/l [Fig. 1, label 5]. No lung embolisms or LVAD obstructions were found by repeated contrast CT. Right heart catheterisation showed a cardiac output of 2.96 l/min, which was less than the pre-LVAD calculation of 4.45 l/min. Besides our suspicion of LVAD thrombosis, we ruled out right ventricular overload by reducing pump speed with 400 rotations/min. Sildenafil and milrinone were added to the medication in an attempt to improve right ventricular function but did not lead to clinical or biochemical improvement. In contrast, LDH rose to 5529 U/l [Fig. 1, label 6]. Since we had no other possible explanation than pump thrombosis, we replaced the LVAD next day using the subcostal approach and found a thrombus localised on the bearing of the inlet cannula (Fig. 2). Clopidogrel replaced aspirin, and the target INR level was raised to 2.5–3.5 by means of home monitoring. The patient recovered and was discharged first to the cardiac rehabilitation centre and then home. Currently (28 April 2014) the LDH level is 404 U/l.

Discussion

Our patient demonstrates that the diagnosis of LVAD thrombus can be difficult because of discrepancies in biochemical, echocardiographic and LVAD diagnostics. First, the initially raised LDH concentrations could have also been a marker of haemolysis instead of thrombosis since the haemoglobin decreased from 7.7 to 4.4 mmol/l. Second, the pump power requirement was more or less constant in the interval between LVAD insertion and exchange. Third, subsequent echocardiography (eight recordings between first LVAD and the replacement) revealed stable findings. In particular, the frequency of aortic valve opening did not increase during follow-up, except directly before LVAD replacement. Finally, repeated contrast CTs did not support our clinical- and biochemical-based suspicion of LVAD thrombosis.
Increased power requirement of the pump can be pathognomonic for an impeller thrombus but is absent when a thrombus is localised at a HeartMate II inlet bearing ball [6]. We did not apply the suggested ramp test [7]. An unexpected, abrupt increase in incidence of LVAD thrombosis was reported recently [8]. In 2011, the incidence of confirmed pump thrombosis was 2.2 %. By January 2013, this had increased to 8.4 % (95 % confidence interval 5–13.9). In addition, the median time from implantation to thrombosis was 18.6 months, but is now much shorter, i.e. 2.7 months. Six weeks after device insertion, a sharp rise in LDH level closely suggested confirmed pump thrombosis. The exact cause of the increased rate of thrombosis remains unclear. After confirmation of the diagnosis, thrombolytic therapy might be an option [9] but pump replacement can be performed with low mortality [10]. It is crucial to monitor LDH levels, among other parameters, to track pump thrombosis in time [11] and apply the algorithm for the diagnosis and management of suspected pump thrombus [12].
In summary: in a LVAD patient receiving adequate oral anticoagulation and antiplatelet therapy despite no increase in pump power requirement, in the case of LDH increase and decrease after intravenous anticoagulation, the diagnosis of pump thrombosis located at the inlet bearing ball is highly likely. Consequently, only pump exchange is probably life-saving.
Funding
None.
Conflict of interest
None declared.
Open Access This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail

Onze productaanbevelingen

Netherlands Heart Journal

Het Netherlands Heart Journal wordt uitgegeven in samenwerking met de Nederlandse Vereniging voor Cardiologie en de Nederlandse Hartstichting. Het tijdschrift is Engelstalig en wordt gratis beschikbaa ...

Literatuur
1.
go back to reference Brugts JJ, Manintveld O, Constantinescu A, et al. Preventing LVAD implantation by early short-term mechanical support and prolonged inodilator therapy: a case series with acute refractory cardiogenic shock treated with veno-arterial extracorporeal membrane oxygenation and optimised medical strategy. Neth Heart J. 2014;22:176–81.PubMedCentralPubMedCrossRef Brugts JJ, Manintveld O, Constantinescu A, et al. Preventing LVAD implantation by early short-term mechanical support and prolonged inodilator therapy: a case series with acute refractory cardiogenic shock treated with veno-arterial extracorporeal membrane oxygenation and optimised medical strategy. Neth Heart J. 2014;22:176–81.PubMedCentralPubMedCrossRef
2.
3.
go back to reference Feldman D, Pamboukian SV, Teuteberg, et al. The 2013 International Society for Heart and Lung Transplantation guidelines for mechanical circulatory support: executive summary. J Heart Lung Transplant. 2013;32:157–87. Feldman D, Pamboukian SV, Teuteberg, et al. The 2013 International Society for Heart and Lung Transplantation guidelines for mechanical circulatory support: executive summary. J Heart Lung Transplant. 2013;32:157–87.
4.
go back to reference Felix SEA, Martina JR, Kirkels JH, et al. Continuous-flow left ventricular assist device support in patients with advanced heart failure: points of interest for the daily management. Eur J Heart Fail. 2012;14:351–6. Felix SEA, Martina JR, Kirkels JH, et al. Continuous-flow left ventricular assist device support in patients with advanced heart failure: points of interest for the daily management. Eur J Heart Fail. 2012;14:351–6.
5.
go back to reference Slaughter MS, Pagani FD, Rogers JG, et al. Clinical management of continuous-flow left ventricular assist devices in advanced heart failure. J Heart Lung Transplant. 2010;29:S1–39.PubMedCrossRef Slaughter MS, Pagani FD, Rogers JG, et al. Clinical management of continuous-flow left ventricular assist devices in advanced heart failure. J Heart Lung Transplant. 2010;29:S1–39.PubMedCrossRef
6.
go back to reference Mokadam NA, Andrus S, Ungerleider A. Thrombus formation in a Heart Mate II. Eur J Cardiothorac Surg. 2011;39:414.PubMedCrossRef Mokadam NA, Andrus S, Ungerleider A. Thrombus formation in a Heart Mate II. Eur J Cardiothorac Surg. 2011;39:414.PubMedCrossRef
7.
go back to reference Uriel N, Morrison KA, Garan AR, et al. Development of a novel echocardiography ramp test for speed optimization and diagnosis of device thrombosis in continuous-flow left ventricular assist devices. J Am Coll Cardiol. 2012;60:1764–75.PubMedCentralPubMedCrossRef Uriel N, Morrison KA, Garan AR, et al. Development of a novel echocardiography ramp test for speed optimization and diagnosis of device thrombosis in continuous-flow left ventricular assist devices. J Am Coll Cardiol. 2012;60:1764–75.PubMedCentralPubMedCrossRef
8.
go back to reference Starling RC, Moazami N, Silvestry SC, et al. Unexpected abrupt increase in left ventricular assist device thrombosis. N Engl J Med. 2014;370:33–40.PubMedCrossRef Starling RC, Moazami N, Silvestry SC, et al. Unexpected abrupt increase in left ventricular assist device thrombosis. N Engl J Med. 2014;370:33–40.PubMedCrossRef
9.
go back to reference Schlendorf K, Patel CB, Gehrig T, et al. Thrombolytic therapy for refractory hemolysis related to thrombosis of continuous flow ventricular assist devices. J Card Fail. 2014. doi:10.1016/ j.cardfail2013.12.011. Schlendorf K, Patel CB, Gehrig T, et al. Thrombolytic therapy for refractory hemolysis related to thrombosis of continuous flow ventricular assist devices. J Card Fail. 2014. doi:10.1016/ j.cardfail2013.12.011.
10.
go back to reference Stulak JM, Cowger J, Haft JW, Romano MA, Aaronson KD, Pagani FD. Device exchange after primary left ventricular assist device implantation: indications and outcomes. Ann Thorac Surg. 2013;95:1262–8.PubMedCrossRef Stulak JM, Cowger J, Haft JW, Romano MA, Aaronson KD, Pagani FD. Device exchange after primary left ventricular assist device implantation: indications and outcomes. Ann Thorac Surg. 2013;95:1262–8.PubMedCrossRef
11.
go back to reference Shah P, Mehta VM, Cowger JA, Aaronson KD, Pagani FD. Diagnosis of hemolysis and device thrombosis with lactate dehydrogenase during left ventricular assist device support. J Heart Lung Transplant. 2014;33:102–4.PubMedCrossRef Shah P, Mehta VM, Cowger JA, Aaronson KD, Pagani FD. Diagnosis of hemolysis and device thrombosis with lactate dehydrogenase during left ventricular assist device support. J Heart Lung Transplant. 2014;33:102–4.PubMedCrossRef
12.
go back to reference Goldstein DJ, John R, Salerno C, et al. Algorithm for the diagnosis and management of suspected pump thrombus. J Heart Lung Transplant. 2013;32:667–70.PubMedCrossRef Goldstein DJ, John R, Salerno C, et al. Algorithm for the diagnosis and management of suspected pump thrombus. J Heart Lung Transplant. 2013;32:667–70.PubMedCrossRef
Metagegevens
Titel
The diagnosis of left ventricular assist device thrombosis
Auteurs
H.Z.R. Gerds
J. Brügemann
M. Rienstra
M.E. Erasmus
Publicatiedatum
01-07-2015
Uitgeverij
Bohn Stafleu van Loghum
Gepubliceerd in
Netherlands Heart Journal / Uitgave 7-8/2015
Print ISSN: 1568-5888
Elektronisch ISSN: 1876-6250
DOI
https://doi.org/10.1007/s12471-015-0705-6

Andere artikelen Uitgave 7-8/2015

Netherlands Heart Journal 7-8/2015 Naar de uitgave