Background
Transcatheter aortic valve implantation (TAVI) is a safe and effective treatment for inoperable, intermediate- or high-risk patients with severe symptomatic aortic stenosis and has been associated with excellent clinical outcomes [
1,
2]. A remaining problem is post-TAVI aortic regurgitation (AR). Moderate or severe AR after TAVI is associated with an increased risk of mortality [
3]. Therefore, reducing AR after TAVI remains an important target to optimise treatment. Wide ranges of AR rates have been reported, partly due to non-standardised and non-quantitative measurements [
3‐
5], resulting in under-or overestimation of AR [
6‐
8] as well as inter-observer variability. So far, quantitative phase-contrast magnetic resonance imaging (MRI) has proved to be the most accurate method for assessing AR and the best method to improve the understanding of the relation between AR severity and clinical outcomes [
9,
10].
Over the past decade, several improvements to address AR have been introduced in different valves with different expansion systems. Two often-used systems are the self-expandable valve (SEV) by Medtronic (CoreValve, Medtronic, Minneapolis, MN, USA) and the balloon-expandable valve (BEV) by Edwards (Sapien 3, Edwards Lifesciences, Irvine, CA, USA). The latter has been specifically designed with the aim of reducing paravalvular AR by an additional outer skirt. Numerous studies have been published on the safety, efficacy and post-procedural AR of the SEV and the latest-generation BEV [
11‐
13]. To our knowledge, however, no randomised comparison between the two types of valve is available that investigates the differences in quantitatively measured AR and clinical outcomes.
Therefore, we conducted a prospective randomised trial to assess the safety and efficacy of the latest-generation BEV and first-generation SEV with respect to clinical outcomes, and AR as determined by MRI.
Methods
The study design and patient selection of the Edwards Sapien 3 periprosthetic leakage evaluation versus Medtronic CoreValve in transfemoral aortic valve implantation (ELECT) trial have been described previously [
14]. Briefly, this study was an investigator-initiated, single-centre, prospective, two-arm randomised controlled trial comparing the latest-generation BEV (Sapien 3) and the first-generation SEV (CoreValve), with the intention of including 106 patients according to block randomisation. The CoreValve was chosen for comparison, because of its widespread use at the time the trial was initiated. The study was approved by the local ethics committee, in accordance with the ethical principles of the Declaration of Helsinki, and written informed consent was obtained from all patients.
Primary endpoint was severity of post-procedural AR, quantitatively assessed by MRI, performed 4–5 days after the index procedure. The severity of AR with respect to regurgitant fraction was graded according to previously published definitions: none/trace <8%, mild 9–20%, moderate 21–39% and severe >40% [
15]. As these definitions were based on patients without a TAVI valve, the flow values used for the regurgitant fraction were measured at the level of the left ventricular outflow tract for the aortic outflow volume, and at the level of the aorta ascending for the regurgitant volume. The aim was to obtain the most reliable measurements for patients post-TAVI according to Kooistra et al. [
16], whose study reported accurate quantification of AR by MRI velocity mapping in the presence of a prosthetic valve, when corrected for a systematic error and when the appropriate MRI slice position is used. Their findings were applied to the present study. For the regurgitant volume, a linear regression correction was applied [for Sapien 3: (measured flow −3.3823)/0.9634; for CoreValve: (measured flow −0.3903)/1.0392], to achieve most accurate AR results based on ex vivo phantom experiments [
16]. The regurgitant fraction was calculated by aortic outflow volume/corrected aortic regurgitant volume × 100% offline using PACS (Sectra PACS, version 17.3, Sectra Workstations IDS7, Linköping, Sweden).
Secondary clinical endpoints at 30 days and 1 year have been described previously and were defined as per the Second Valve Academic Research Consortium (VARC-2) definitions [
14,
17]. All patients were actively followed up for 1 year, and had clinical visits and transthoracic echocardiographic (TTE) evaluations at 6 months and 1 year. Quality of life was assessed on average 1 month before and 1 year after the index procedure by the Short-Form-36 and EuroQol visual analogue scale, and new cerebral lesions were measured by cerebral MRI (performed at day 4–5 after the index procedure).
All endpoints were analysed on an intention-to-treat basis. Continuous variables were presented as means ± standard deviation or medians (interquartile range) and compared using the two-sided unpaired Student
t-test or Mann-Whitney U test, depending on data distribution. Categorical variables were presented as counts and percentages and compared using Fisher’s exact test. Time-to-event curves were constructed using Kaplan-Meier curves and compared with the log-rank test. The paired
t-test, repeated-measures analysis of variance and Wilcoxon signed rank test were used to assess the change in effective regurgitant orifice area and mean aortic gradient over time between and within groups. The correlations between continuous and ordinal variables were assessed using Spearman’s correlation coefficient. Sample size calculation has been described previously [
14]. In summary, the calculation yielded a sample size of 49 patients in each arm to detect a difference between BEV over SEV with respect to AR with a power of 80%.
All tests were two-sided, and a p-value of less than 0.05 was regarded to be statistically significant. Data were analysed using IBM SPSS Statistics software version 21 (IBM Corp., Armonk, NY, USA).
Discussion
The ELECT trial is the first randomised controlled trial that compares quantitative AR as measured by MRI and clinical outcomes between the latest-generation balloon-expandable valve and a widely used first-generation self-expandable valve.
In this study, patients with a BEV had lower AR rates post-TAVI. Furthermore, clinical outcomes at 1 year showed a lower mortality rate, fewer strokes and MACCE, and a shorter hospital stay for the BEV group.
A distinguishing feature of the investigated latest-generation BEV to reduce post-TAVI AR is the additional skirt around the valve. Indeed, a lower AR rate for the new BEV was observed compared with the SEV. In recent studies of the latest-generation BEV, similar AR rates to those in our study have been reported (none/trace 56–74% [
11,
12,
18]; moderate/severe 0–3.5% [
11,
12,
18]). These AR rates are lower than those of the previous BEV generation (none/trace 29–66% [
19‐
21]; moderate/severe 5–23% [
3,
19‐
22]). The very low AR rates for BEV observed in the present study might reflect the effectiveness of the new skirt of this valve. Other potential factors influencing the reduced AR and complications might be the increased operator experience and the use of multi-detector computed tomography for adequate valve sizing.
The major strength of the present study is the fact that the AR severity was quantitatively assessed by MRI. A problem of previous studies is the wide range of AR rates, which are partly due to non-standardised and non-quantitative AR measurements that remain imprecise due to under- or overestimation of AR as well as inter-observer variability [
4,
6‐
8]. Since MRI is currently considered to be the most accurate method available with low intra- and inter-observer variability and high accuracy, resulting in good reproducibility [
9,
10,
16,
23,
24], the assessment and comparison of AR severity between the two valves has been assessed in this study as accurately as possible. The accuracy of AR is important, since AR is a frequent remaining problem that is associated with an increased 30-day and 1‑year mortality [
3].
To investigate the association between quantitatively measured AR severity and clinical outcomes, NYHA class improvement was used, since the mortality rate was too low to study a correlation. No significant association was found in our study between AR severity and NYHA class improvement. This might be due to the low AR rates, with mostly none/trace or mild AR severity. However, this lack of association between mild AR and clinical outcomes is in accordance with the FRANCE‑2 Registry and the PARTNER II SAPIEN 3 trial [
25,
26].
The secondary clinical endpoints of our randomised trial showed a significantly lower mortality rate at 1 year with no mortality at all in the BEV group. The mortality rate of 19% for the SEV group is in line with other studies of this first-generation valve [
13,
19,
22,
27,
28]. The low mortality rate associated with the BEV might be explained by the reduction in complications that are associated with increased mortality, i.e. reduction in AR severity, strokes and life-threatening bleeding.
The striking low rates of stroke and life-threatening bleeding in the BEV group might be caused by many patient- or procedure-related factors. However, these could also be related to the low delivery profile (14 or 16 Fr), the lower crossing profile of the Sapien 3 system and a better distal flexion of the delivery catheter, resulting in fewer aortic arch injuries and life-threatening bleedings in patients receiving this latest-generation BEV. The overall low complication and mortality rates of the BEV are also reflected in the composite endpoints according to the VARC‑2 criteria, with a high device success rate, better early safety and a lower rate of MACCE. The observed difference in clinical outcomes between the SEV and BEV groups in our study might represent the difference in valve generation. Studies of the newer generations of the SEV have found lower complication rates, which are in line with the latest-generation BEV with a 1-year all-cause mortality rate and stroke rate around 8.5% and 3.5% respectively[
29,
30].
Study limitations
The results of the clinical outcomes in the present study should be interpreted in the context of its small sample size and the lack of power to detect differences in endpoints—due to low recruitment—between the valves, resulting in wider confidence intervals and possible underestimation of differences. Nevertheless, the reported differences that were statistically significant can be considered to be a real difference and not a coincidence, since the statistical type I error (false-positive statistically significant result; α = 0.05) is not related to the power of a test. However, as stated by Button et al. [
31], it is possible that the magnitude of the found effect is exaggerated due to the underpowered endpoint, which might explain the high stroke rate in the SEV group. Secondly, in this study, the first-generation CoreValve was used, which is nowadays not employed anymore. However, at the time the trial was initiated, these were the two most common valves used commercially.
Clinical implications and future perspectives
The results of this study show a high-quality comparison between two valve types regarding AR severity as quantitatively assessed by MRI. Although the use of MRI for the quantification of AR is not feasible in daily clinical practice due to its high costs, low availability and certain contra-indications (e.g. in patients with claustrophobia or < 6 weeks after implantation of pacemaker leads), it is of additive value as regards improving the accuracy and reproducibility of AR as well as reducing variability with respect to AR severity. In addition, studies with quantitative AR assessment are needed in order to improve the accuracy of AR estimation and to better understand the relationship between AR severity and clinical outcomes as well as differences between next-generation valves. Therefore, measuring AR by MRI should be considered in future studies comparing valves. Finally, this small study could be a preamble to a new randomised comparison between current-generation TAVI valves, but probably requires a larger sample size because of smaller differences in technical and clinical outcomes.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit