Pressure measurement devices: from technical assessment to clinical performance

Tijdschrift:
Journal of Foot and Ankle Research > bijlage 1/2012
Auteurs:
Claudia Giacomozzi, Moreno D’Amico, Piero Roncoletta

Background

Technical assessment of pressure measurement devices (PMDs) should guarantee for their appropriate use in the clinics. The study aims at proving the validity of the assessment methodology ISS proposed [ 1], and at quantifying the impact of PMD performance on clinical assessment.

Materials and methods

Three commercial PMDs were first assessed and then compared during barefoot walking: PMDa and PMDb - resistive technology, 1sens/cm 2 – were assessed on-site, while PMDc – capacitive technology, 4sens/cm 2 - was tested on-the-bench and on-site [ 1]. The PMDs were aligned on the floor to capture successive at-regimen steps of the left foot of one trained volunteer; 10 complete steps were acquired in both directions for each PMD; data were temporally normalised and averaged; main kinetic parameters were extracted.

Results

Preliminary results (Table 1 and Figure 1): i) PMDc resulted accurate and was used as a reference; ii) PMDa was found inaccurate on-site and delivered unreliable gait data; iii) PMDb was found accurate on-site but performed significantly worse than PMDc during gait.
Table 1
Results from the on-the-bench and on-site assessment, and with respect to some clinically relevant parameters.
PMD under test
ISS Full technical assessment
ISS On-site partial assessment
“gait” assessment: Peak pressure (kPa)
“gait” assessment: Mean pressure (kPa))
“gait” assessment: Integral (kPa*s)[ 2]
a
not performed
error >10% at 250kPa
100 (4)**
80 (2)**
39 (2)**
b
not performed
error < 5% at 250kPa
266 (12)*
191 (8)*
85 (9)*
C
accuracy error < 5% up to 1200kPa
error < 5% at 250kPa
744 (137)
367 (17)
152 (23)
* statistically different from PMDc corresponding data (p<0.05, also verified with respect to the ± 5% maximum error); ** statistically different from PMDb and PMDc corresponding data (p<0.05, also verified with respect to the ± 5% maximum error)

Conclusions

To conclude: i) on-site assessment up to 250kPa proved to be necessary but not sufficient to guarantee for a good PMD performance during gait; ii) a thorough on-the-bench assessment is effective and recommended; iii) use of PMDb data might be misleading in research and risky in the clinics. The study is going on with the comparison among other commercial PMDs and under a wide range of testing conditions.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Onze productaanbevelingen

BSL Podotherapeut Totaal

Binnen de bundel kunt u gebruik maken van boeken, tijdschriften, e-learnings, web-tv's en uitlegvideo's. BSL Podotherapeut Totaal is overal toegankelijk; via uw PC, tablet of smartphone.

Literatuur
Over dit artikel

Andere artikelen bijlage 1/2012

Journal of Foot and Ankle Research 1/2012 Naar de uitgave