Preliminary results of a biomechanics driven design of a total ankle prosthesis

Auteurs: A Leardini, JJ O'Connor, F Catani, M Romagnoli, S Giannini

Gepubliceerd in: Journal of Foot and Ankle Research | bijlage 1/2008

share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail
insite
ZOEKEN

Introduction

A new design of total ankle replacement [1] was developed according to extensive prior biomechanical research [25]. A linkage-based mathematical model was used to design for the first time ligament-compatible shapes of the prosthesis components in the sagittal plane (Fig. 1). The radius of the metal talar component in the sagittal plane is about 50% longer than that of the normal talus, the metal tibial component is spherically convex. A fully conforming meniscal bearing is interposed between them and free to move. Experiments in cadaver specimens confirmed the mathematical prediction that the bearing moves forwards on both metal components during dorsiflexion and backwards during plantarflexion. FEA models [6] and experiments with a wear simulator [7] confirmed that the risk of wear is minimised. Preliminary clinical results are here reported as support to the biomechanical claims.

Methods

Between July 2003 and May 2007, the prosthesis was implanted in 189 patients at 7 hospitals in Northern Italy. Mean age was 59.4 years. The diagnosis was post-traumatic OA in 79.3%, primary OA in 6.9%, RA in 8.0%. At one hospital, range of motion was measured in the operating theatre before and after implantation in 90 ankles, and meniscal motion in lateral radiograms at maximum plantar- and dorsi-flexion in 30 ankles.

Results

At September 2007, the mean follow-up was 21 months. The mean pre-operative AOFAS score of 41.1 rose to 80.2, 79.7, 77.9, and 79.0 respectively at 18, 24, 36, 48 months. Mean dorsi-flexion increased after implantation from 0.1° to 9.7°, plantar-flexion from 15.1° to 24.6°. From radiographic measurements, the range of full motion, 14° – 53°, was significantly correlated to the range of bearing movement on the tibial component, 2–11 mm (r2 = 0.37, p < 0.0005), as predicted. Two revision operations had been performed, respectively for obvious surgical and indication errors.
There were no device related revisions (loosening, fracture, dislocation). The Kaplan-Meier survival rate (components removal as end-point) at 4 years was 97% (Confidence interval 92–100).

Conclusion

The shapes of the three components are compatible with physiologic ankle mobility and with the natural role of the ligaments. The survival rate at four years compares well with multi-centre 5-year rates published by the Swedish (531 cases, survival 78%), Norwegian (257, 89%) and New Zealand (202, 86%) registries and with a recent meta-analysis [8]. The early clinical results have demonstrated safety and efficacy and encourage a widening of the clinical trial.
Open AccessThis article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail

Onze productaanbevelingen

BSL Podotherapeut Totaal

Binnen de bundel kunt u gebruik maken van boeken, tijdschriften, e-learnings, web-tv's en uitlegvideo's. BSL Podotherapeut Totaal is overal toegankelijk; via uw PC, tablet of smartphone.

Literatuur
1.
go back to reference Leardini A, et al: Clin Orth Rel Res. 2004, 424: 39-46. 10.1097/01.blo.0000132246.26172.b7.CrossRef Leardini A, et al: Clin Orth Rel Res. 2004, 424: 39-46. 10.1097/01.blo.0000132246.26172.b7.CrossRef
4.
go back to reference Leardini A, et al: Med Biol Eng Comp. 2001, 39: 168-175. 10.1007/BF02344799.CrossRef Leardini A, et al: Med Biol Eng Comp. 2001, 39: 168-175. 10.1007/BF02344799.CrossRef
5.
go back to reference Leardini A, et al: Gait Post. 2002, 15: 220-229. 10.1016/S0966-6362(01)00153-9.CrossRef Leardini A, et al: Gait Post. 2002, 15: 220-229. 10.1016/S0966-6362(01)00153-9.CrossRef
8.
go back to reference Stengel D, et al: Arch Orth Trau Surg. 2005, 125: 109-119. 10.1007/s00402-004-0765-3.CrossRef Stengel D, et al: Arch Orth Trau Surg. 2005, 125: 109-119. 10.1007/s00402-004-0765-3.CrossRef
Metagegevens
Titel
Preliminary results of a biomechanics driven design of a total ankle prosthesis
Auteurs
A Leardini
JJ O'Connor
F Catani
M Romagnoli
S Giannini
Publicatiedatum
01-09-2008
Uitgeverij
BioMed Central
Gepubliceerd in
Journal of Foot and Ankle Research / Uitgave bijlage 1/2008
Elektronisch ISSN: 1757-1146
DOI
https://doi.org/10.1186/1757-1146-1-S1-O8

Andere artikelen bijlage 1/2008

Journal of Foot and Ankle Research 1/2008 Naar de uitgave