Magnetic resonance imaging (MRI) is the fastest growing standard of care in diagnostic imaging. Because of its unique properties such as noninvasive accurate assessment of left ventricular (LV) volumes and regional LV wall motion abnormalities, quantification of flow and tissue characterisation, MRI utilisation in the US is still growing at approximately 3 % per year [1]. Moreover, indications for MRI may expand due to innovative techniques such as magnetic resonance angiography (MRA) or the use of targeted contrast agents.
This increased usage of MRI is paralleled by an annual growth of de novo pacemakers (>700,000) and implantable cardio-defibrillators (>200,000) [2]. Since many recipients of pacemakers and ICDs have substantial comorbidities, it is estimated that up to 75 % will develop an indication for MRI examination during further life [3, 4].
The hazards of MRI examination in patients with pacemakers and ICDs are a consequence of the techniques used [5]. First the static magnetic field may yield mechanical forces on ferromagnetic components and may have unpredictable magnetic sensor activation. Second, the modulated radiofrequency (RF) field may cause heating of cardiac tissue adjacent to the lead tip, may induce life-threatening arrhythmias and may interact with the device leading to over- or under-sensing. Third, the gradient magnetic field may induce life-threatening arrhythmias or may cause over- or under-sensing. Finally, combined field effects can alter device function or lead to electronic reset. Electrical reset may have major implications since it can cause pacemaker inhibition or induce fatal tachyarrhythmias. Currently, MRI conditional CIEDs have become available in which the risk of an electronic reset is significantly reduced by adaptations in the device. Moreover, lead design has been improved to avoid tip heating.
In this issue of the Netherlands Heart Journal Van der Graaf and co-workers give an excellent overview of the current status of MRI in patients with cardiac implantable electronic devices (CIED) [6]. They underline that although safety in these patients who are scheduled for MRI examination has traditionally been regarded a major issue, the recently published European Society of Cardiology (ESC) guidelines express a different view [7]. These 2013 guidelines state that MRI can be safely performed irrespective of the properties of the CIED, the type of MRI examination or patient characteristics such as presence or absence of underlying rhythm, as long as safety restrictions are met. Differences in risk between chest and non-chest MRI are not mentioned. Moreover, the absolute number of patients on which these guidelines are based is limited. A total of seven studies each report on small numbers of chest MRIs. For example, in the study by Sommer and co-workers on MRI in 51 patients only 5 had chest scanning, and Naehle and colleagues reported on 18 CIED patients but only 8 patients had examinations of the chest [8, 9]. In a larger study, Mollerus evaluated 127 MRI scans in 103 patients of which 62 were chest MRIs [10]. No threshold elevation was found, but both sensing amplitudes and pacing impedances significantly changed. By far the largest study was conducted by Nazarian and co-workers in 438 patients with pacemakers and ICDs [11]. In this study, pacemaker-dependent patients with an ICD were excluded. In 89 patients chest MRI was performed. Three of the 438 patients had an electrical reset (power-on-reset) leading to occasional pacing inhibition in 2 patients. There were no symptomatic events in these non-pacemaker dependant individuals. In the studies mentioned CRT devices are strongly underrepresented.
Thus, current guidelines concerning MRI in patients with CIED are based on a limited number of relatively small studies in patients with a wide range of devices and leads, typically excluding pacemaker-dependent patients as well as patients with CRT devices. The evidence for safety of chest MRI is based on less than 200 patients and potentially associated arrhythmias are still reported. Although the recent guidelines have become more lenient and major complications have rarely been reported, risk assessment in patients with CIED is not straightforward. We would therefore advocate a prudent approach in scheduling patients with CIED for MRI examination, especially those with older type or CRT devices and when using newer (3T) MRI suites. MRI provides mesmerising image quality, but mesmerisation of CIED should be avoided.

Funding

None.

Conflict of interests

None declared.
Open Access This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail

Onze productaanbevelingen

Netherlands Heart Journal

Het Netherlands Heart Journal wordt uitgegeven in samenwerking met de Nederlandse Vereniging voor Cardiologie en de Nederlandse Hartstichting. Het tijdschrift is Engelstalig en wordt gratis beschikbaa ...

Literatuur
2.
go back to reference Mond HG, Proclemer A. The 11th world survey of cardiac pacing and implantable cardio-defibrillator: calendar year 2009–2011. Pacing Clin Electrophysiol. 2011;34:1013–27.PubMedCrossRef Mond HG, Proclemer A. The 11th world survey of cardiac pacing and implantable cardio-defibrillator: calendar year 2009–2011. Pacing Clin Electrophysiol. 2011;34:1013–27.PubMedCrossRef
3.
go back to reference Kalin R, Stanton MS. Current clinical issues for MRI scanning of pacemaker and defibrillator patients. Pacing Clin Electrophysiol. 2005;28:326–8.PubMedCrossRef Kalin R, Stanton MS. Current clinical issues for MRI scanning of pacemaker and defibrillator patients. Pacing Clin Electrophysiol. 2005;28:326–8.PubMedCrossRef
4.
go back to reference Levine GN, Gomes AS, Arai EA, et al. American Heart Association committee on diagnostic and interventional cardiac catheterization. Circulation. 2007;116:2878–91.PubMedCrossRef Levine GN, Gomes AS, Arai EA, et al. American Heart Association committee on diagnostic and interventional cardiac catheterization. Circulation. 2007;116:2878–91.PubMedCrossRef
5.
go back to reference Kanal E, Barcovich AJ, Bell C, et al. ACR guidance document on MR safe practices: 2013. J Magn Reson Imaging. 2013;37:501–30.PubMedCrossRef Kanal E, Barcovich AJ, Bell C, et al. ACR guidance document on MR safe practices: 2013. J Magn Reson Imaging. 2013;37:501–30.PubMedCrossRef
7.
go back to reference Brignole M, Auricchio A, Baron-Esquivias G, et al. 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). Europace. 2013;15:1070–118.PubMedCrossRef Brignole M, Auricchio A, Baron-Esquivias G, et al. 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). Europace. 2013;15:1070–118.PubMedCrossRef
8.
go back to reference Sommer T, Vahlhaus C, Lauck G, et al. MRI and cardiac pacemakers: in-vitro evaluation and in vivo studies in 51 patients at 0.5T. Radiology. 2000;215:869–79.PubMedCrossRef Sommer T, Vahlhaus C, Lauck G, et al. MRI and cardiac pacemakers: in-vitro evaluation and in vivo studies in 51 patients at 0.5T. Radiology. 2000;215:869–79.PubMedCrossRef
9.
go back to reference Naehle CP, Strach K, Thomas D, et al. Magnetic resonance imaging at 1.5-T in patients with implantable cardioverter-defibrillators. J Am Coll Cardiol. 2009;4(54):549–55.CrossRef Naehle CP, Strach K, Thomas D, et al. Magnetic resonance imaging at 1.5-T in patients with implantable cardioverter-defibrillators. J Am Coll Cardiol. 2009;4(54):549–55.CrossRef
10.
go back to reference Mollerus M, Albin G, Lipinski M, et al. Magnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption rate restrictions. Europace. 2010;12:947–51.PubMedCrossRef Mollerus M, Albin G, Lipinski M, et al. Magnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption rate restrictions. Europace. 2010;12:947–51.PubMedCrossRef
11.
go back to reference Nazarian S, Hansford R, Roguin A, et al. A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med. 2011;155:415–24.PubMedCrossRef Nazarian S, Hansford R, Roguin A, et al. A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med. 2011;155:415–24.PubMedCrossRef
Metagegevens
Titel
Magnetic resonance imaging and devices: a mesmerising combination
Auteurs
C. P. Allaart
C. C. de Cock
Publicatiedatum
01-06-2014
Uitgeverij
Bohn Stafleu van Loghum
Gepubliceerd in
Netherlands Heart Journal / Uitgave 6/2014
Print ISSN: 1568-5888
Elektronisch ISSN: 1876-6250
DOI
https://doi.org/10.1007/s12471-014-0562-8

Andere artikelen Uitgave 6/2014

Netherlands Heart Journal 6/2014 Naar de uitgave

Letter to the Editor

Epilepsy and noncompaction