Gepubliceerd in: Netherlands Heart Journal 3/2013

01-03-2013 | Original Article

Long-term performance of the St Jude Riata 1580–1582 ICD lead family

Auteurs: S. D. A. Valk, D. A. M. J. Theuns, L. Jordaens

Gepubliceerd in: Netherlands Heart Journal | Uitgave 3/2013

Log in om toegang te krijgen
share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail

Abstract

Objective

Safety concerns about the Riata ICD shock lead were recently raised, with insulation failure due to conductor externalisation. Its incidence and presentation were assessed, and predictors of insulation failure and lead survival of the Riata 1580–1582 were studied, retrospectively, before the official recall.

Methods

All 374 patients at the Erasmus Medical Center between July 2003 and December 2007 with a 1580, 1581 or 1582 shock lead.

Results

The majority of the patients were male (78 %), with a median age of 60 years (IQR 52–70); primary prevention in 61 %. Median follow-up was 60.3 months (IQR 35.5–73.2), with 117 (31 %) patients dying. Electrical abnormalities (mainly noise, 65 %) were observed in 20/257 patients (7.8 %). Definite conductor externalisation was confirmed with fluoroscopy or chest X-ray in 16 patients, and in one after extraction. One patient presented with a drop in the high-voltage impedance trend with a short circuit of the ICD system during defibrillation testing, and needed to be shocked externally. In 8 more patients, conductor externalisation was found during an elective procedure. No predictors of externalisation could be found, except for the use of single coil (p = 0.02). Median time to conductor externalisation was 5 years (IQR 3.1–6.2). Lead externalisation was observed in 5.4 % (95 % CI 3.1–9.3) at 5 years and 22.7 % (95 % CI 13.6–36.6) at 8 years.

Conclusion

A high incidence of insulation defects associated with conductor externalisation in the Riata ICD lead family is observed. The mode of presentation is diverse. This type of insulation failure can lead to failure of therapy delivery.
Literatuur
1.
go back to reference Epstein AE, DiMarco JP, Ellenbogen KA, et al. American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices); American Association for Thoracic Surgery; Society of Thoracic Surgeons. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. Circulation. 2008;117(21):e350–408.PubMedCrossRef Epstein AE, DiMarco JP, Ellenbogen KA, et al. American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices); American Association for Thoracic Surgery; Society of Thoracic Surgeons. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. Circulation. 2008;117(21):e350–408.PubMedCrossRef
2.
go back to reference Cannom DS, Prystowsky EN. The evolution of the implantable cardioverter defibrillator. NASPE history series. Pacing Clin Electropysiol. 2004;27:419–31.CrossRef Cannom DS, Prystowsky EN. The evolution of the implantable cardioverter defibrillator. NASPE history series. Pacing Clin Electropysiol. 2004;27:419–31.CrossRef
3.
go back to reference Camm AJ, Nisam S. European utilization of the implantable defibrillator: has 10 years changed the ‘enigma’? Europace. 2010;12:1063–9.CrossRef Camm AJ, Nisam S. European utilization of the implantable defibrillator: has 10 years changed the ‘enigma’? Europace. 2010;12:1063–9.CrossRef
4.
go back to reference Hauser RG. The growing mismatch between patient longevity and the service life of implantable cardioverter-defibrillators. J Am Coll Cardiol. 2005;45:2022–5.PubMedCrossRef Hauser RG. The growing mismatch between patient longevity and the service life of implantable cardioverter-defibrillators. J Am Coll Cardiol. 2005;45:2022–5.PubMedCrossRef
5.
go back to reference Kleemann T, Becker T, Doenges K, et al. Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years. Circulation. 2007;115:2474–80.PubMedCrossRef Kleemann T, Becker T, Doenges K, et al. Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years. Circulation. 2007;115:2474–80.PubMedCrossRef
6.
go back to reference Gradaus R, Breithardt G, Böcker D. ICD leads: design and chronic dysfunctions. Pacing Clin Electrophysiol. 2003;26(2 Pt 1):649–57.PubMedCrossRef Gradaus R, Breithardt G, Böcker D. ICD leads: design and chronic dysfunctions. Pacing Clin Electrophysiol. 2003;26(2 Pt 1):649–57.PubMedCrossRef
7.
go back to reference Hauser RG, Hayes DL. Increasing hazard of Sprint Fidelis implantable cardioverter-defibrillator lead failure. Hear Rhythm. 2009;6:605–10.CrossRef Hauser RG, Hayes DL. Increasing hazard of Sprint Fidelis implantable cardioverter-defibrillator lead failure. Hear Rhythm. 2009;6:605–10.CrossRef
8.
go back to reference Birnie DH, Parkash R, Exner DV, et al. Clinical predictors of Fidelis lead failure: report from the Canadian Heart Rhythm Society Device Committee. Circulation. 2012;125:1217–25.PubMedCrossRef Birnie DH, Parkash R, Exner DV, et al. Clinical predictors of Fidelis lead failure: report from the Canadian Heart Rhythm Society Device Committee. Circulation. 2012;125:1217–25.PubMedCrossRef
9.
go back to reference Valk S, Luijten R, Jordaens L. Insulation damage in a shock wire: an unexpected fluoroscopic image. Pacing Clin Electrophysiol. 2010;33:770–2.PubMedCrossRef Valk S, Luijten R, Jordaens L. Insulation damage in a shock wire: an unexpected fluoroscopic image. Pacing Clin Electrophysiol. 2010;33:770–2.PubMedCrossRef
10.
go back to reference Porterfield JG, Porterfield LM, Kuck KH, et al. Clinical performance of the St. Jude Medical Riata defibrillation lead in a large patient population. J Cardiovasc Electrophysiol. 2010;21:551–6.PubMedCrossRef Porterfield JG, Porterfield LM, Kuck KH, et al. Clinical performance of the St. Jude Medical Riata defibrillation lead in a large patient population. J Cardiovasc Electrophysiol. 2010;21:551–6.PubMedCrossRef
11.
go back to reference Chester KM. Important product information: St Jude Medical Riata and Riata ST silicone endocardial leads. Sylmar California: St Jude Medical; December 15, 2010. Chester KM. Important product information: St Jude Medical Riata and Riata ST silicone endocardial leads. Sylmar California: St Jude Medical; December 15, 2010.
12.
go back to reference Carlson M, Tsung P. Medical devise advisory. Sylmar California: St Jude Medical; November,28, 2011. Carlson M, Tsung P. Medical devise advisory. Sylmar California: St Jude Medical; November,28, 2011.
13.
go back to reference Epstein AE, Baker 2nd JH, Beau SL, et al. Performance of the St. Jude Medical Riata leads. Hear Rhythm. 2009;6:204–9.CrossRef Epstein AE, Baker 2nd JH, Beau SL, et al. Performance of the St. Jude Medical Riata leads. Hear Rhythm. 2009;6:204–9.CrossRef
14.
go back to reference Erkapic D, Duray GZ, Bauernfeind T, et al. Insulation defects of thin high-voltage ICD leads: an underestimated problem? J Cardiovasc Electrophysiol. 2011;22:1018–22.PubMedCrossRef Erkapic D, Duray GZ, Bauernfeind T, et al. Insulation defects of thin high-voltage ICD leads: an underestimated problem? J Cardiovasc Electrophysiol. 2011;22:1018–22.PubMedCrossRef
15.
go back to reference Schmutz M, Delacrétaz E, Schwick N, et al. Prevalence of asymptomatic and electrically undetectable intracardiac inside-out abrasion in silicon-coated Riata® and Riata® ST implantable cardioverter-defibrillator leads. Int J Cardiol. 2012 Jan 9. [Epub ahead of print]. Schmutz M, Delacrétaz E, Schwick N, et al. Prevalence of asymptomatic and electrically undetectable intracardiac inside-out abrasion in silicon-coated Riata® and Riata® ST implantable cardioverter-defibrillator leads. Int J Cardiol. 2012 Jan 9. [Epub ahead of print].
16.
go back to reference Mehta D, Nayak HM, Singson M, et al. Late complications in patients with pectoral defibrillator implants with transvenous defibrillator lead systems: high incidence of insulation breakdown. Pacing Clin Electrophysiol. 1998;21:1893–900.PubMedCrossRef Mehta D, Nayak HM, Singson M, et al. Late complications in patients with pectoral defibrillator implants with transvenous defibrillator lead systems: high incidence of insulation breakdown. Pacing Clin Electrophysiol. 1998;21:1893–900.PubMedCrossRef
17.
go back to reference Duray GZ, Israel CW, Schmitt J, et al. Implantable cardioverter-defibrillator lead disintegration at the level of the tricuspid valve. Hear Rhythm. 2008;5:1224–5.CrossRef Duray GZ, Israel CW, Schmitt J, et al. Implantable cardioverter-defibrillator lead disintegration at the level of the tricuspid valve. Hear Rhythm. 2008;5:1224–5.CrossRef
18.
go back to reference Hauser RG, Abdelhadi R, McGriff D, et al. Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads. Hear Rhythm. 2012;9:1227–35.CrossRef Hauser RG, Abdelhadi R, McGriff D, et al. Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads. Hear Rhythm. 2012;9:1227–35.CrossRef
19.
go back to reference Krebsbach A, Alhumaid F, Henrikson CA, et al. Premature Failure of a Riata Defibrillator Lead Without Impedance Change or Inappropriate Sensing: A Case Report and Review of the Literature. J Cardiovasc Electrophysiol. 2011;22:1070–2.PubMedCrossRef Krebsbach A, Alhumaid F, Henrikson CA, et al. Premature Failure of a Riata Defibrillator Lead Without Impedance Change or Inappropriate Sensing: A Case Report and Review of the Literature. J Cardiovasc Electrophysiol. 2011;22:1070–2.PubMedCrossRef
20.
go back to reference Parvathaneni SV, Ellis CR, Rottman JN. High prevalence of insulation failure with externalized cables in St Jude Medical Riata family ICD leads: Fluoroscopic grading scale and correlation to extracted leads. Hear Rhythm. 2012;9:1218–24.CrossRef Parvathaneni SV, Ellis CR, Rottman JN. High prevalence of insulation failure with externalized cables in St Jude Medical Riata family ICD leads: Fluoroscopic grading scale and correlation to extracted leads. Hear Rhythm. 2012;9:1218–24.CrossRef
21.
go back to reference Jalal Z, Derval N, Ploux S, et al. Unusual failure of a multilumen, small-diameter implantable cardioverter-defibrillator lead. Hear Rhythm. 2009;7:1166–7.CrossRef Jalal Z, Derval N, Ploux S, et al. Unusual failure of a multilumen, small-diameter implantable cardioverter-defibrillator lead. Hear Rhythm. 2009;7:1166–7.CrossRef
22.
go back to reference Richards MW, Warren CE, Anderson MH. Late failure of a single-coil transvenous implantable cardioverter-defibrillator lead associated with conductor separation. Europace. 2010;12:1191–2.PubMedCrossRef Richards MW, Warren CE, Anderson MH. Late failure of a single-coil transvenous implantable cardioverter-defibrillator lead associated with conductor separation. Europace. 2010;12:1191–2.PubMedCrossRef
23.
go back to reference Hauser RG, McGriff D, Retel LK. Riata implantable cardioverter-defibrillator lead failure: analysis of explanted leads with a unique insulation defect. Hear Rhythm. 2012;9:742–9.CrossRef Hauser RG, McGriff D, Retel LK. Riata implantable cardioverter-defibrillator lead failure: analysis of explanted leads with a unique insulation defect. Hear Rhythm. 2012;9:742–9.CrossRef
Metagegevens
Titel
Long-term performance of the St Jude Riata 1580–1582 ICD lead family
Auteurs
S. D. A. Valk
D. A. M. J. Theuns
L. Jordaens
Publicatiedatum
01-03-2013
Uitgeverij
Bohn Stafleu van Loghum
Gepubliceerd in
Netherlands Heart Journal / Uitgave 3/2013
Print ISSN: 1568-5888
Elektronisch ISSN: 1876-6250
DOI
https://doi.org/10.1007/s12471-012-0341-3

Andere artikelen Uitgave 3/2013

Netherlands Heart Journal 3/2013 Naar de uitgave

Rhythm Puzzle - Answer

Palpitations, should one worry?