The Effects of Varying Ankle Foot Orthosis Stiffness on Gait in Children with Spastic Cerebral Palsy Who Walk with Excessive Knee Flexion

  • Yvette L. Kerkum ,

    y.kerkum@vumc.nl

    Affiliation Department of Rehabilitation Medicine, MOVE Research Institute Amsterdam, VU University Medical Center, Amsterdam, The Netherlands

  • Annemieke I. Buizer,

    Affiliation Department of Rehabilitation Medicine, MOVE Research Institute Amsterdam, VU University Medical Center, Amsterdam, The Netherlands

  • Josien C. van den Noort,

    Affiliation Department of Rehabilitation Medicine, MOVE Research Institute Amsterdam, VU University Medical Center, Amsterdam, The Netherlands

  • Jules G. Becher,

    Affiliation Department of Rehabilitation Medicine, MOVE Research Institute Amsterdam, VU University Medical Center, Amsterdam, The Netherlands

  • Jaap Harlaar ,

    ‡ These authors are joint senior authors on this work.

    Affiliation Department of Rehabilitation Medicine, MOVE Research Institute Amsterdam, VU University Medical Center, Amsterdam, The Netherlands

  • Merel-Anne Brehm

    ‡ These authors are joint senior authors on this work.

    Affiliations Department of Rehabilitation Medicine, MOVE Research Institute Amsterdam, VU University Medical Center, Amsterdam, The Netherlands, Department of Rehabilitation, Academic Medical Center, University of Amsterdam, Amsterdam, The Netherlands

Abstract

Introduction

Rigid Ankle-Foot Orthoses (AFOs) are commonly prescribed to counteract excessive knee flexion during the stance phase of gait in children with cerebral palsy (CP). While rigid AFOs may normalize knee kinematics and kinetics effectively, it has the disadvantage of impeding push-off power. A spring-like AFO may enhance push-off power, which may come at the cost of reducing the knee flexion less effectively. Optimizing this trade-off between enhancing push-off power and normalizing knee flexion in stance is expected to maximize gait efficiency. This study investigated the effects of varying AFO stiffness on gait biomechanics and efficiency in children with CP who walk with excessive knee flexion in stance. Fifteen children with spastic CP (11 boys, 10±2 years) were prescribed with a ventral shell spring-hinged AFO (vAFO). The hinge was set into a rigid, or spring-like setting, using both a stiff and flexible performance. At baseline (i.e. shoes-only) and for each vAFO, a 3D-gait analysis and 6-minute walk test with breath-gas analysis were performed at comfortable speed. Lower limb joint kinematics and kinetics were calculated. From the 6-minute walk test, walking speed and the net energy cost were determined. A generalized estimation equation (p<0.05) was used to analyze the effects of different conditions. Compared to shoes-only, all vAFOs improved the knee angle and net moment similarly. Ankle power generation and work were preserved only by the spring-like vAFOs. All vAFOs decreased the net energy cost compared to shoes-only, but no differences were found between vAFOs, showing that the effects of spring-like vAFOs to promote push-off power did not lead to greater reductions in walking energy cost. These findings suggest that, in this specific group of children with spastic CP, the vAFO stiffness that maximizes gait efficiency is primarily determined by its effect on knee kinematics and kinetics rather than by its effect on push-off power.

Trial Registration

Dutch Trial Register NTR3418

Introduction

Gait in children with Cerebral Palsy (CP) is often characterized by abnormal gait biomechanics, such as excessive knee flexion during stance. Associated with such gait deviations, an elevated walking energy cost is often observed [13], which may contribute to activity limitations [4,5]. To treat these gait-related problems in CP, Ankle Foot Orthoses (AFOs) are commonly prescribed.

When prescribing an AFO, the specific gait deviations and functional deficits of the patient should be clearly identified, such that these can be optimally addressed by the design and mechanical properties of the AFO [6]. A rigid ventral shell AFO (vAFO) is typically used for children who walk with excessive knee flexion in stance [6]; a gait pattern that is particularly energy consuming [3,7,8]. Mechanically, a vAFO aims to shift the ground reaction force more anterior relative to the knee, which reduces the external flexion moment. This is expected to reduce knee flexion and decrease the elevated internal knee extensor moment during stance [6]. Accordingly, this may reduce walking energy cost [9,10].

Although a vAFO may be effective in reducing knee flexion and subsequent walking energy cost, its high stiffness has the disadvantage of impeding ankle range of motion. Ankle range of motion during gait has been shown to be a key kinematic factor in gait efficiency [11,12]. In fact, a reduced ankle range of motion during gait, especially towards plantar flexion, limits push-off power about the ankle, which almost always leads to an increased walking energy cost [11,13]. Besides, a common strategy to compensate for reduced push-off power is to deliver work around the hip [1417], which may also increase walking energy cost [14,18].

The metabolic penalty of limiting the ankle push-off power may be reduced by applying spring-like AFOs. These AFOs allow dorsiflexion in the beginning of stance phase, thereby storing energy within the AFO. This energy can be returned in pre-swing, which may support push-off power, therewith enhancing gait efficiency in terms of walking energy cost [19,20]. Considering the key role of ankle range of motion during gait, an AFO that would additionally allow plantar flexion in late stance might support push-off power and gait efficiency even further [21,22].

The efficacy of spring-like AFOs to improve gait is however partly dependent on their stiffness. This has been shown in simulation models [19], as well as in studies in healthy adults [21] and in adult patient populations [2328], where results indicated that changing the AFO stiffness significantly affected knee and ankle kinematics and kinetics, as well as walking energy cost. Results also indicated that the reduction in walking energy cost could be improved by choosing the appropriate AFO stiffness [26]. Such stiffness-based maximization of gait efficiency may also apply to children with CP [29], which is relevant considering that AFOs are not always effective in terms of reducing walking energy cost in these children [9,10,30], while this is an important goal of AFO prescription [31]. However, the effects of different degrees of AFO stiffness on gait biomechanics and walking energy cost have not previously been reported in this patient group.

The aim of this study was to investigate the effects of varying vAFO stiffness on lower limb joint kinematics and kinetics and walking energy cost in children with spastic CP whose gait pattern is characterized by excessive knee flexion in stance. Stiffer vAFOs were expected to normalize knee flexion most effectively, though at the expense of obstructing ankle range of motion and push-off power. Contrarily, the less stiff vAFOs were expected to enhance push-off power, but to be less effective in counteracting knee flexion. We hypothesized that the optimal vAFO stiffness (i.e. at which walking energy cost would be lowest), would be defined by a trade-off between improving knee kinematics and kinetics, and enhancing ankle push-off power.

Methods

Participants

Institutional review board (Medisch Ethische Toetsingscommissie Vrije Universiteit medisch centrum) approvals were obtained prior to the start of the study (approval number NL37910.029.11). Both parents of all participants and participants above 12 years old provided written informed consent. All measurements were performed in accordance to the Declaration of Helsinki.

Data used in the study were collected in the context of the AFO-CP trial [32] (S1 Protocol). Participants in the AFO-CP trial were recruited from the rehabilitation department of a university hospital in the Netherlands and its affiliated rehabilitation centers (Fig 1). Children diagnosed with spastic CP and aged between 6 and 14 years old were included. Other inclusion criteria were a Gross Motor Function Classification System (GMFCS) [33] level I, II or III, and a barefoot gait pattern that was characterized by excessive knee flexion in (mid)stance (i.e. more than 10° in midstance). Children with ankle plantar flexion contractures, knee flexion contractures and/or hip flexion contractures of more than 10° were excluded.

thumbnail
Fig 1. Trial flow diagram.

Abbreviations: vAFO, ventral shell Ankle Foot Othosis.

https://doi.org/10.1371/journal.pone.0142878.g001

Materials

Participants were prescribed with a new vAFO (Fig 2), which was designed with a ventral shell and rigid footplate. The vAFOs were made of pre-preg carbon fiber, and manufactured with an integrated ankle hinge (Neuro Swing®, Fior & Gentz, Lüneburg, Germany). This hinge allows mechanical characteristics (e.g. stiffness) to be varied within the same orthosis, as it holds a shaft for dorsal and plantar flexion in which pre-compressed springs with different mechanical properties can be inserted [29]. The hinge is available in different sizes, each accompanied with a spring package covering a range of stiffness degrees. The size of the hinge (14mm or 16mm) was individually determined following a standard prescription protocol (Fior & Gentz, Lüneburg, Germany), which is based on weight and height of the child. For this study, the hinge was set into three stiffness configurations: i) rigid, ii) stiff and iii) flexible. For the rigid configuration, the hinge’s spring-like properties were eliminated, aiming to act like a conventional rigid vAFO with limited range of motion. For the stiff and flexible configurations, stiffness towards dorsiflexion was varied by applying the stiffest available spring and one less stiff spring in the hinge’s ventral shaft. These springs were expected to sufficiently improve gait in children with spastic CP who walk with excessive knee flexion [29] The most compliant spring available was used towards plantar flexion for both configurations.

thumbnail
Fig 2. Picture of the spring-hinged ventral shell Ankle-Foot Orthosis, including possible adjustments using the hinge.

The hinge allows: A, the stiffness to be varied towards dorsal flexion and plantar flexion; B, adjustment of the alignment of the ventral shell with respect to the foot; C, the range of motion to be varied, although this is also dependent of the spring inserted (stiffer springs allow less range of motion). Figs adapted from Fior & Gentz.

https://doi.org/10.1371/journal.pone.0142878.g002

The vAFOs were worn in combination with the children’s own shoes (i.e. shoes with flat, flexible soles), referred to as the AFO-footwear combination. For shoes-only measurements, children were instructed to wear shoes that they normally used when walking without vAFOs.

Procedure

At the start of each measurement session (i.e. vAFO stiffness evaluation), measures of bare foot height [m] and weight [kg] were determined using an electronic scale (DGI 250D, KERN DE v. 3.3 10/2004, Kern & Sohn GmbH, Balingen, Germany). Leg length (from the trochanter head to the lateral malleolus) was measured while the participant was standing upright, with knees extended as much as possible. Stiffness of the new vAFO was randomly (i.e. block-randomized) set into one of the three configurations. After setting the hinge, the AFO-footwear combination was tuned following a common clinical protocol [34].

Each vAFO stiffness configuration was worn for an acclimatization period of four weeks, after which efficacy of that vAFO was evaluated. This evaluation consisted of a 3D gait analysis to measure gait biomechanics and a 6-minute walk test to measure walking energy cost. Additionally, the mechanical properties of the vAFO [35] and various shoe parameters were assessed [36]. Next, the hinge was set into the second stiffness configuration and the procedure was repeated until all three stiffness configurations were evaluated. For the shoes-only (i.e. baseline) condition, the same set of measurements was performed, and these were conducted during the assessment of the second vAFO stiffness configuration.

Measurements

Gait biomechanics.

Gait analyses were performed in our gait laboratory. Participants were instructed to walk up and down a 10m-walkway with integrated force plate (OR6-5-1000, AMTI, Watertown, USA) at a comfortable walking speed. Kinematic data were collected using an optoelectronic motion capture system (OptoTrak 3020, Northern Digital, Waterloo, Canada). Technical clusters of three markers were rigidly attached to the trunk, pelvis, thighs, shanks (including the vAFOs’ ventral shell) and feet, and anatomically calibrated by probing 32 bony landmarks [37]. The bony markers of the foot (i.e. calcaneus and metatarsal joints I and IV) were probed on the shoe, and horizontally aligned in the sagittal plane of the foot while the foot was flat on the ground. The foot segments included the vAFO’s foot part and shoe, where no movement between these components was assumed. Segment movements were tracked (sample frequency: 100 Hz) and synchronized with force plate data (sample frequency: 1000 Hz). Data collection trials were repeated until three strides with correct foot placement (i.e. within the borders of the force plate) of the most affected leg were recorded.

Walking energy cost.

A portable breath gas-analysis system (Metamax 3B, Cortex Biophysik, Leipzig, Germany) was used to record breath-by-breath oxygen uptake (VO2) and carbon dioxide production (VCO2) values. Each measurement started with a rest test. Participants were seated, while the equipment was put on and the facemask was fitted. Then participants sat down quietly watching a movie for six minutes. They were instructed not to talk or laugh during the measurement. After completion of the rest test, participants performed a 6-minute walk test at comfortable walking speed on a 40m indoor oval track, which has been shown to be a sufficiently sensitive and reliable protocol for energy cost measurements in children with CP [1,38].

AFO properties and shoe parameters.

The vAFO’s mechanical properties were measured using the Bi-articular Reciprocal Universal Compliance Estimator (BRUCE) [35], which is an instrument to measure AFO mechanical properties. Each vAFO was placed into the BRUCE, such that the rotation axis of the hinge was aligned with the “ankle” axis of the BRUCE. The ventral shaft of the vAFO was fixated to the BRUCE by a Velcro strap. The vAFO was then manually pushed towards dorsiflexion and plantar flexion, while the exerted moment and ankle angle were continuously recorded. For each movement direction, measurements were repeated three times [29].

Shoe parameters (i.e. the height of the shoe sole and the heel-sole differential angle) were obtained with the Vertical Inclinometer on a Rail (VICTOR) [36], a dedicated instrument to define these parameters.

Data processing

Gait biomechanics.

Optoelectronic marker data and force plate data of the three recorded trials were analyzed using custom-made software (Bodymech, www.bodymech.nl) based on MATLAB version R2011a (The Mathworks, Natick, USA). For each trial, initial contact and toe-off in the gait cycle of the ipsilateral leg were determined using force-plate data, while foot angular velocity was used to determine the gait events of the contralateral leg [39]. Based on these gait events, relevant phases of the gait cycle were determined (Fig 3). Furthermore, walking speed [m·s-1] was determined and averaged over three trials.

thumbnail
Fig 3. Representation of relevant phases of the gait cycle.

Phases of the gait cycle were defined as i) stance: initial contact to toe-off; ii) step: initial contact to contralateral initial contact; iii) single support (SS): contralateral toe-off to contralateral initial contact. Definitions of specific gait events and mean timing [%gait cycle]: i) contralateral toe-off (cTO) [11%]; ii) midstance (MSt): the moment that the malleolus marker of the contralateral leg passed the malleolus marker of the ipsilateral leg [33%]; iii) contralateral initial contact (cIC) [50%]; iv) toe-off (TO) [64%]; v) timing of minimal knee flexion angle during single support (peak knee extension angle) (TKEpk): [38%]. Abbreviations: cTO, contralateral toe-off; cIC, contralateral initial contact; IC, initial contact; TKEpk, timing of peak knee extension angle; MSt, midstance; SS, single support; TO, toe-off.

https://doi.org/10.1371/journal.pone.0142878.g003

3D lower limb joint flexion-extension angles [deg] were calculated from the anatomically calibrated cluster marker data, according to ISB anatomical frames [37,40]. As bony landmarks of the foot were probed on the shoe, a coordinate frame of the shoe (including vAFO and foot) was calculated. Shoe parameters (i.e. height of the shoe sole and the heel-sole differential angle) obtained using VICTOR [36], were used to correct the measured ankle angle for the offset between probing positions and the actual position of the bony landmarks, and for orientation of the foot in the shoe, which is dependent on the heel sole differential. Lower limb net joint flexion-extension moments [Nm·kg-1] were calculated with force plate data using inverse dynamics [41], expressed with respect to the proximal segment frame [42], and normalized to body weight. Also lower limb joint powers [W·kg-1] and work [J·kg-1] (i.e. integral of net ankle power) were calculated.

From the mean joint angles, moments and powers as a function of the gait cycle, we determined specific relevant parameters, primarily at the knee and ankle joints. At the knee joint, these included the knee flexion-extension angle and moment at midstance, peak knee extension angle during single support (KEpk) and the knee moment at timing of KEpk. Ankle joint parameters included range of motion (RoM) during the stride and peak power generation during push-off, where push-off was defined as the period in late stance and pre-swing in which the net ankle power was positive. Positive and negative work over the gait cycle, as well as the net work (i.e. positive + negative) during push-off were determined for all lower limb joints.

Walking energy cost.

Breath-by-breath VO2 and VCO2 values in minute three to six of both the rest and the walk test were used to calculate the mean steady-state energy consumption values (ECSrest and ECSwalk) [J·kg-1·m-1] [43]. The mean walking speed [m·min-1] was measured over the same time frame of the walk test. From these assessments, the net energy cost (EC) [J·kg-1·m-1] was calculated as (ECSwalk–ECSrest) / walking speed. To control for the influence of different body dimensions of children, net EC values were normalized according to the scheme by Schwartz et al. [44] to calculate the net non-dimensional energy cost, which was expressed as a percentage of speed-matched control cost (SMC-EC) [45].

AFO mechanical properties.

Angle-moment relation curves resulting from the BRUCE assessments were analyzed using custom-made software based on Matlab version R2011a (The Mathworks, Natick, USA). First, the vAFO’s neutral angle was determined, which is the angle of the vAFO when no force is exerted. Subsequently, we determined stiffness [Nm·deg-1], range of motion [deg] (i.e. the spring’s elastic range), and the threshold [Nm] (i.e. the exerted moment at the start of the spring’s elastic range) [29]. For the rigid configuration, only the stiffness was determined, as other variables were not applicable.

To calculate the vAFO’s contributions to the ankle work, first the contributions to the net ankle moment were determined. For the rigid vAFO, stiffness was multiplied by the vAFO’s deflection angle (i.e. the migration of the vAFO from its neutral angle) for each point in time. To align these deflection angles during gait to the angles as measured using BRUCE, it was assumed that the vAFO had negligible displacement during sway [20]. Considering the low stiffness towards plantar flexion of the stiff and flexible vAFO, and thus the possibility of plantar flexion movement, this assumption could not be preserved. As such, the alignment was done using the angle of the AFO-footwear combination when exceeding the spring’s threshold during gait. Using the moments exerted by the vAFO and the ankle’s angular velocity for each point in time, contributions to net ankle power and work [J·kg-1] over the gait cycle and during push-off could be calculated.

Statistics.

Statistical analyses were done with SPSS version 20 (SPSS Inc, Chicago, USA), using an alpha level of 0.05 for all tests of significance. Descriptive statistics (means and standard deviations (SD)) were used to summarize socio-demographic characteristics, disease characteristics, gait-related outcomes, and AFO mechanical properties. Differences in gait-related outcomes between conditions were analyzed with generalized estimating equation analyses, with conditions (i.e. shoes-only, rigid vAFO, stiff vAFO and flexible vAFO) as within-subject factor. Exchangeable correlation structures were assumed. Walking speed, as measured during the gait analyses, was added to the model as covariate [12].

Results

Fifteen children with spastic CP (11 boys, 4 girls) were included in the study (Fig 1). Social-demographic and disease characteristics of these children are presented in Table 1. Data from the physical examination are presented in Table 2. In 13 children, the effects of all three vAFO configurations were evaluated. In one child, only the flexible and stiff vAFOs were evaluated because this child refused to wear the rigid vAFO. Another child could not acclimatize to the flexible vAFO, because of too much foot deformation within the vAFO leading to pressure marks, and therefore only the stiff and rigid vAFOs were evaluated.

thumbnail
Table 2. Baseline passive range of motion and spasticity values of relevant joints and muscles of the most affected leg (n = 15).

https://doi.org/10.1371/journal.pone.0142878.t002

Gait biomechanics

During the gait analyses, the mean (SD) walking speed while walking with shoes-only was 1.09 (0.21) m·s-1. Speed was significantly lower while walking with vAFOs, i.e. 1.07 (0.24), 1.00 (0.21), and 1.05 (0.17) m·s-1 for the rigid, stiff and flexible configuration respectively (Wald χ2 = 10.3, p = 0.016).

Differences in knee joint angles and between walking with shoes-only and walking with the vAFO were comparable for all vAFOs. All vAFOs decreased the knee flexion angle at contralateral toe-off, midstance, and at timing of KEpk. Also the internal knee flexion-extension moment at midstance and at timing of KEpk were significantly improved by all vAFOs (Table 3; Fig 4C and 4D).

thumbnail
Fig 4. Mean (n = 15) of the most relevant gait parameters as a function of the gait cycle.

Vertical lines indicate timing of toe-off (with similar timing for stiff and flexible vAFOs).

https://doi.org/10.1371/journal.pone.0142878.g004

thumbnail
Table 3. Results of generalized estimating equation analyses for relevant gait parameters, with positive values representing flexion angles, internal extension moments, and power generation.

https://doi.org/10.1371/journal.pone.0142878.t003

At the ankle joint, we found that ankle RoM was significantly reduced by all vAFOs compared to walking with shoes-only, though ankle RoM was significantly less reduced by the stiff and flexible vAFO. Peak ankle power generation was reduced by the rigid vAFO, while it was preserved by the stiff and flexible vAFO (Table 3; Fig 4A, 4C and 4D). Ankle work was reduced most by the rigid vAFO. The vAFOs’ contribution to the ankle work over the gait cycle was smaller for the rigid vAFO, compared to the spring-like vAFOs, while no significant differences were found in the vAFOs’ contributions to ankle work during push-off between vAFOs (Table 4; Fig 5).

thumbnail
Fig 5. Mean (n = 15) net internal ankle moment and ankle power for walking with different degrees of vAFO stiffness, with mean vAFO contributions as a function of the gait cycle.

The area underneath the power curves (panel D-F) represents the net ankle work and vAFO work.

https://doi.org/10.1371/journal.pone.0142878.g005

thumbnail
Table 4. Results of generalized estimating equation analyses for hip, knee, and ankle work.

https://doi.org/10.1371/journal.pone.0142878.t004

Walking energy cost

Walking speed during the 6-minute walk test was comparable between all conditions. Compared to walking with shoes-only, the net EC was significantly reduced with 9.8%, 11.5%, and 8.2% by the rigid, stiff and flexible vAFO respectively. No significant differences were found between vAFOs (Table 5). On average, the overall (i.e. all vAFOs) reduction in net EC was 0.67 J·kg-1·m-1 (11%), with large individual differences. While some participants showed an improvement, i.e. reduction, in net EC with at least one of the vAFOs, others showed no response or even an increase of their energy cost while walking with the vAFO (Fig 6). When comparing the SMC-EC, only a significant reduction was found for the rigid and stiff vAFO compared to walking shoes-only (Table 5).

thumbnail
Fig 6. Overview of individual net energy cost responses.

The x-axis represents baseline (i.e. shoes-only) net energy cost values and the y-axis indicates the change in net energy cost as a result of walking with each vAFO. Vertically aligned dots thus represent the same participant. Abbreviations: vAFO, ventral shell ankle-foot orthosis

https://doi.org/10.1371/journal.pone.0142878.g006

thumbnail
Table 5. Results of generalized estimating equation analyses for walking speed and energy cost.

https://doi.org/10.1371/journal.pone.0142878.t005

AFO mechanical properties

The rigid vAFO was much stiffer (mean (SD) 3.8 (0,7) Nm·deg-1) towards dorsiflexion than the stiff (mean (SD) 1.6 (0.4) Nm·deg-1) and flexible vAFO (mean (SD) 0.7 (0,2) Nm·deg-1). vAFO properties of the stiff compared to the flexible vAFO were different towards dorsal flexion, as the stiff vAFO showed a smaller RoM and higher threshold. The mechanical properties towards plantar flexion were comparable between these two spring-like vAFOs (Table 6).

thumbnail
Table 6. Mechanical properties of the Ankle Foot Orthoses.

https://doi.org/10.1371/journal.pone.0142878.t006

Discussion

This study in children with spastic CP showed that, compared to walking shoes-only, rigid vAFOs and spring-like vAFOs comparably reduced the knee flexion angle and internal knee flexor moment during stance. Favorable effects on ankle RoM and power generation were found for the spring-like vAFOs, but not for the rigid vAFOs. Results further showed that all vAFOs improved gait efficiency compared to walking with shoes-only, but no significant differences were found between vAFOs.

This is the first clinical study investigating the effects of rigid versus spring-like vAFOs on gait in children with CP. Earlier, we evaluated the potential value of the spring-like vAFOs that were used in the current study. Results of that study suggested that the threshold of the springs within the vAFO could reduce knee flexion by preventing dorsiflexion in the beginning of the stance phase until approximately 0.5 Nm·kg-1. In the current study however, it appeared that the flexible and stiff vAFO were only able to prevent dorsiflexion until a net ankle moment of respectively 0.3 Nm·kg-1 and 0.4 Nm·kg-1. This was reflected in the ankle angle, which gradually increased towards dorsiflexion in the two spring-like vAFOs, but approached the maximal dorsiflexion angle already in early-stance (Fig 4A). However, for optimal performance, the spring should however be compressed into its elastics range at midstance. Accordingly, the vAFOs did not improve knee angles into normal values, although knee flexion in early stance was reduced by all vAFOs (Table 2; Fig 4B). Nonetheless, ankle and knee flexion-extension angles from midstance onwards were improved by all vAFOs compared to shoes-only. These improvements are comparable to the study of Rogozinski et al. [46], who investigated the efficacy of a similar type of AFO (i.e. ventral shell) in children with CP.

The comparable reductions of the knee flexion angles during stance between vAFOs were in contrast to our hypothesis. This might be explained by the hinge’s limited RoM, as the spring-like vAFOs can be expected to act rigidly when hitting the hinge’s dorsal stop [29]. In addition, the ankle RoM measured during walking with the rigid vAFO was still 7° (mainly movement in stance). This can be considered as slack, counteracting its extending effect on the knee angle. The rigid vAFO was however most effective in reducing the internal knee extensor moment in late stance (at TKEpk), though only significantly compared to the flexible vAFO. With forward CoP excursion being similar between vAFOs, this difference in knee moment might be explained by changes either in magnitude or direction of the ground reaction force in the sagittal plane (i.e. distance to the knee rotation center), possibly caused by altered trunk positions during walking [47,48]. Nonetheless, also the flexible and stiff vAFO normalized the internal knee extensor moment over the whole stance phase compared to walking with shoes-only (Fig 4D).

Differences between vAFOs were observed in ankle kinematics and kinetics. The ankle power generation was reduced by the rigid vAFO, while this was preserved by the spring-like vAFOs compared to walking with shoes-only (Table 2; Fig 4E). Nonetheless, peak ankle power generation was only half of reference values of typically developing children (Fig 4E). The potential beneficial effect of spring-like vAFOs on push-off function is in accordance with studies comparing different AFO designs in children with hemiplegia [49,50], showing that spring-like AFOs allow for the storage and return of energy without constraining remaining voluntary push-off. Similar to ankle power, a greater reduction in ankle work was found while walking with the rigid vAFO compared to the stiff and flexible vAFO. This is in accordance with studies in adult patient populations [20,51]. The vAFO’s energy return (i.e. vAFO work) was also smallest for the rigid vAFO, which is related to the limited range of motion. Contrary to our expectations, the energy return of the stiff and flexible vAFO was comparable (Table 3; Fig 5). The stiffness properties between these two spring-like vAFOs might not have been sufficiently different to reveal differences in vAFO work during walking in this study.

Compared to walking with shoes-only, the overall mean net EC was significantly lower when walking with vAFOs (-11%). This reduction in energy cost, was due to a decrease in energy consumption, as walking speed was not significantly changed by the vAFOs. This is similar to a study of Buckon et al. [10], who evaluated the effects of different vAFO designs made of polypropylene in children with spastic diplegia. Brehm et al [9] reported a decrease in net EC of only 6% in children with CP while walking with posterior leaf spring or rigid polypropylene vAFOs compared to barefoot walking. Sub-analyses in that study revealed that the mean reduction in net EC was much larger when comparing responders to non-responders. Although our study sample was too small to perform such sub-analyses, our results also indicate varying responses between subjects. Differences in the patients’ underlying impairments, such as spasticity, could explain the variety in gait biomechanics and walking energy cost. As our patient population had low levels of spasticity, possibly other factors might explain the variety in results. Unfortunately, our sample size was too small to analyze such underlying mechanisms. Nonetheless, our results suggest that most beneficial effects on net EC are seen in the children with highest baseline energy cost levels (Fig 6).

The lack of significant differences in net EC between the three vAFOs (Table 4) indicate that the potential benefit of preserving push-off power by the spring-like vAFOs may not necessarily enhance walking energy cost. Although we did not relate changes in biomechanical parameters to changes in net EC as a result of varying vAFO stiffness, Brehm et al. [9] found that changes in knee flexion during stance were significantly related to changes in net EC, while changes in push-off power were not. These results may suggest that, in children with CP who walk with excessive knee flexion, the normalization of knee kinematics and kinetics are dominant with regard to gait efficiency improvement. Alternatively, our study sample may have been too small to show differences in gait efficiency between vAFO stiffness levels. Additionally, the stiffness properties between vAFOs may not have been sufficiently distinct. Furthermore, the pre-defined stiffness levels were not matched to specific patient-related characteristics and underlying impairments. The nature of the optimal match between AFO stiffness and patient characteristics has not yet been unraveled, making it difficult to define the optimal stiffness level in relation to each individual patient. Additional studies, evaluating the effects of AFO stiffness levels outside the currently investigated range and in a larger group of children with CP, are needed to further study the relation between changes in gait biomechanics and changes in net EC. This may provide clues to improve and optimize AFO treatment aimed at enhancing gait performance in these children.

Accurate measurements of ankle kinematics in a shod condition are a challenge. In our study, we used VICTOR [36] to minimize the effects of probing on the shoe instead of the foot. The calculations of ankle kinematics and vAFO contributions were however based on the assumption that no movement occurred between vAFOs, shoes, and feet. Possibly, small movements may have occurred, interfering with the results, which can be considered a limitation. Secondly, our study population was homogeneous regarding levels of spasticity, passive range of motion, selective motor control, and gait pattern (i.e. excessive knee flexion), which limits the generalizability of results to other sub-groups within CP. However, such a gait-based selection of patients is essential to adequately evaluate the effects of vAFO mechanical properties on gait. The small sample size is a third limitation of the study, which could explain that some differences were not statistically significant. Despite these limitations, this is the first study providing accurate descriptions of AFO stiffness and its effects on gait in CP, some of which were evident and clinically important.

In conclusion, despite the homogeneity within our study sample of children with spastic CP, various responses to different degrees of vAFO stiffness were seen. Overall, both rigid vAFOs and spring-like vAFOs reduced the knee flexion angle and internal knee flexion moment comparably in the stance phase of gait, while favorable effects on ankle power generation were only found for the spring-like vAFOs. These favorable effects of spring-like vAFOs on push-off power did however not lead to greater reductions of walking energy cost. These findings might suggest that, in children with CP who walk with excessive knee flexion in stance, the optimal vAFO stiffness that maximizes gait efficiency is primarily defined by its effect on knee kinematics and kinetics during stance and less by its effect on ankle push-off power.

Supporting Information

S1 Dataset. Dataset containing demographic and disease characteristics of the participants.

https://doi.org/10.1371/journal.pone.0142878.s002

(SAV)

S2 Dataset. Complete dataset of outcome parameters used for statistical analysis.

https://doi.org/10.1371/journal.pone.0142878.s003

(SAV)

S1 Protocol. Complete protocol of the AFO-CP trial.

https://doi.org/10.1371/journal.pone.0142878.s004

(PDF)

Acknowledgments

The authors would like to thank Marjolein Piening for her assistance during the measurements. vAFOs were manufactured by OIM Noppe Orthopedietechniek, Noordwijkerhout, the Netherlands and Kamer Orthopedie, Amsterdam, the Netherlands.

Author Contributions

Conceived and designed the experiments: YLK AIB JCN JGB JH MAB. Performed the experiments: YLK. Analyzed the data: YLK JCN. Contributed reagents/materials/analysis tools: YLK. Wrote the paper: YLK AIB MAB JH. Critical reading of manuscript: JCN JGB.

References

  1. 1. Brehm MA, Becher J, Harlaar J. (2007) Reproducibility evaluation of gross and net walking efficiency in children with cerebral palsy. Dev Med Child Neurol 49: 45–48. pmid:17209976
  2. 2. Thomas SS, Buckon CE, Schwartz MH, Russman BS, Sussman MD, Aiona MD. (2009) Variability and minimum detectable change for walking energy efficiency variables in children with cerebral palsy. Dev Med Child Neurol 51: 615–621. pmid:19627334
  3. 3. Waters RL, Mulroy S. (1999) The energy expenditure of normal and pathologic gait. Gait Posture 9: 207–231. pmid:10575082
  4. 4. Maltais D, Bar-Or O, Galea V, Pierrynowski M. (2001) Use of orthoses lowers the O(2) cost of walking in children with spastic cerebral palsy. Med Sci Sports Exerc 33: 320–325. pmid:11224824
  5. 5. Kamp FA, Lennon N, Holmes L, Dallmeijer AJ, Henley J, Miller F. (2014) Energy cost of walking in children with spastic cerebral palsy: relationship with age, body composition and mobility capacity. Gait Posture 40: 209–214. pmid:24768085
  6. 6. Davids JR, Rowan F, Davis RB. (2007) Indications for orthoses to improve gait in children with cerebral palsy. J Am Acad Orthop Surg 15: 178–188. pmid:17341675
  7. 7. Rose J, Gamble JG, Burgos A, Medeiros J, Haskell WL. (1990) Energy expenditure index of walking for normal children and for children with cerebral palsy. Dev Med Child Neurol 32: 333–340. pmid:2332124
  8. 8. Campbell J, Ball J. (1978) Energetics of walking in cerebral palsy. Orthop Clin North Am 9: 374–377. pmid:662304
  9. 9. Brehm MA, Harlaar J, Schwartz M. (2008) Effect of ankle-foot orthoses on walking efficiency and gait in children with cerebral palsy. J Rehabil Med 40: 529–534. pmid:18758669
  10. 10. Buckon CE, Thomas SS, Jakobson-Huston S, Moor M, Sussman M, Aiona M. (2004) Comparison of three ankle-foot orthosis configurations for children with spastic diplegia. Dev Med Child Neurol 46: 590–598. pmid:15344518
  11. 11. Ballaz L, Plamondon S, Lemay M. (2010) Ankle range of motion is key to gait efficiency in adolescents with cerebral palsy. Clin Biomech (Bristol, Avon) 25: 944–948.
  12. 12. Schwartz MH, Rozumalski A, Trost JP. (2008) The effect of walking speed on the gait of typically developing children. J Biomech 41: 1639–1650. pmid:18466909
  13. 13. Collins SH, Kuo AD. (2010) Recycling energy to restore impaired ankle function during human walking. PLoS One 5: e9307. pmid:20174659
  14. 14. Farris DJ, Hampton A, Lewek MD, Sawicki GS. (2015) Revisiting the mechanics and energetics of walking in individuals with chronic hemiparesis following stroke: from individual limbs to lower limb joints. Journal of NeuroEngineering and Rehabilitation 12.
  15. 15. Chen G, Patten C. (2008) Joint moment work during the stance-to-swing transition in hemiparetic subjects. J Biomech 41: 877–883. pmid:18067898
  16. 16. Peterson CL, Kautz SA, Neptune RR. (2011) Muscle work is increased in pre-swing during hemiparetic walking. Clin Biomech (Bristol, Avon) 26: 859–866.
  17. 17. Steinwender G, Saraph V, Zwick EB, Steinwender C, Linhart W. (2001) Hip locomotion mechanisms in cerebral palsy crouch gait. Gait Posture 13: 78–85. pmid:11240355
  18. 18. Kuo AD, Donelan JM. (2010) Dynamic principles of gait and their clinical implications. Phys Ther 90: 157–174. pmid:20023002
  19. 19. Bregman DJ, van der Krogt MM, De G, V, Harlaar J, Wisse M, Collins SH. (2011) The effect of ankle foot orthosis stiffness on the energy cost of walking: A simulation study. Clin Biomech (Bristol, Avon).
  20. 20. Bregman DJ, Harlaar J, Meskers CG, De G, V. (2012) Spring-like Ankle Foot Orthoses reduce the energy cost of walking by taking over ankle work. Gait Posture 35: 148–153. pmid:22050974
  21. 21. Collins SH, Wiggin MB, Sawicki GS. (2015) Reducing the energy cost of human walking using an unpowered exoskeleton. Nature.
  22. 22. Bennett BC, Abel MF, Wolovick A, Franklin T, Allaire PE, Kerrigan DC. (2005) Center of mass movement and energy transfer during walking in children with cerebral palsy. Arch Phys Med Rehabil 86: 2189–2194. pmid:16271569
  23. 23. Singer ML, Kobayashi T, Lincoln LS, Orendurff MS, Foreman KB. (2014) The effect of ankle-foot orthosis plantarflexion stiffness on ankle and knee joint kinematics and kinetics during first and second rockers of gait in individuals with stroke. Clin Biomech (Bristol, Avon) 29: 1077–1080.
  24. 24. Kobayashi T, Leung AK, Akazawa Y, Hutchins SW. (2011) Design of a stiffness-adjustable ankle-foot orthosis and its effect on ankle joint kinematics in patients with stroke. Gait Posture 33: 721–723. pmid:21376602
  25. 25. Kobayashi T, Leung AK, Akazawa Y, Hutchins SW. (2013) The effect of varying the plantarflexion resistance of an ankle-foot orthosis on knee joint kinematics in patients with stroke. Gait Posture 37: 457–459. pmid:22921491
  26. 26. Bregman DJJ (2011) The optimal ankle foot orthosis: the influence of mechanical properties of Ankle Foot Orthoses on the walking ability of patients with central neurological disorders. Amsterdam: Vrije Universiteit.
  27. 27. Russell EE, Blanck RV, Harper NG, Hsu JR, Wilken JM. (2014) How does ankle-foot orthosis stiffness affect gait in patients with lower limb salvage? Clin Orthop Relat Res 472: 3026–3035. pmid:24817379
  28. 28. Kobayashi T, Singer ML, Orendurff MS, Gao F, Daly WK, Foreman KB. (2015) The effect of changing plantarflexion resistive moment of an articulated ankle-foot orthosis on ankle and knee joint angles and moments while walking in patients post stroke. Clin Biomech (Bristol, Avon) 30: 775–780.
  29. 29. Kerkum YL, Brehm MA, Buizer AI, van den Noort JC, Becher JG, Harlaar J. (2014) Defining the Mechanical Properties of a Spring-Hinged Ankle Foot Orthosis to Assess its Potential Use in Children With Spastic Cerebral Palsy. J Appl Biomech 30: 728–731. pmid:25010887
  30. 30. Buckon CE, Thomas SS, Jakobson-Huston S, Sussman M, Aiona M. (2001) Comparison of three ankle-foot orthosis configurations for children with spastic hemiplegia. Dev Med Child Neurol 43: 371–378. pmid:11409825
  31. 31. Wingstrand M, Hagglund G, Rodby-Bousquet E. (2014) Ankle-foot orthoses in children with cerebral palsy: a cross sectional population based study of 2200 children. BMC Musculoskelet Disord 15: 327. pmid:25274143
  32. 32. Kerkum YL, Harlaar J, Buizer AI, van den Noort JC, Becher JG, Brehm MA. (2013) Optimising Ankle Foot Orthoses for children with cerebral palsy walking with excessive knee flexion to improve their mobility and participation; protocol of the AFO-CP study. BMC Pediatr 13: 17. pmid:23374998
  33. 33. Palisano R, Rosenbaum P, Walter S, Russell D, Wood E, Galuppi B. (1997) Development and reliability of a system to classify gross motor function in children with cerebral palsy. Dev Med Child Neurol 39: 214–223. pmid:9183258
  34. 34. Kerkum YL, Brehm MA, van Hutten K, van den Noort JC, Harlaar J, Becher JG, et al. (2015) Acclimatization of the gait pattern to wearing an ankle-foot orthosis in children with spastic cerebral palsy. Clinical Biomechanics.
  35. 35. Bregman DJ, Rozumalski A, Koops D, de Groot V, Schwartz M, Harlaar J. (2009) A new method for evaluating ankle foot orthosis characteristics: BRUCE. Gait Posture 30: 144–149. pmid:19520576
  36. 36. Kerkum YL, Noort JCvd, Brehm MA, Buizer AI, Benistant M, Harlaar J. (2014) VICTOR: a new tool to measure shoe parameters that are important for ankle kinematics. Gait & Posture 39: S136–S137.
  37. 37. Cappozzo A, Catani F, Croce UD, Leardini A. (1995) Position and orientation in space of bones during movement: anatomical frame definition and determination. Clin Biomech (Bristol, Avon) 10: 171–178.
  38. 38. Brehm MA, Knol DL, Harlaar J. (2008) Methodological considerations for improving the reproducibility of walking efficiency outcomes in clinical gait studies. Gait Posture 27: 196–201. pmid:17467276
  39. 39. Zeni JA Jr., Richards JG, Higginson JS. (2008) Two simple methods for determining gait events during treadmill and overground walking using kinematic data. Gait Posture 27: 710–714. pmid:17723303
  40. 40. Wu G, Cavanagh PR. (1995) ISB recommendations for standardization in the reporting of kinematic data. J Biomech 28: 1257–1261. pmid:8550644
  41. 41. Zatsiorsky VM (2002) Kinetics of human motion. Champaign, Ill., [etc.]: Human Kinetics.
  42. 42. Schache AG, Baker R. (2007) On the expression of joint moments during gait. Gait Posture 25: 440–452. pmid:17011192
  43. 43. Garby L, Astrup A. (1987) The relationship between the respiratory quotient and the energy equivalent of oxygen during simultaneous glucose and lipid oxidation and lipogenesis. Acta Physiol Scand 129: 443–444. pmid:3577829
  44. 44. Schwartz MH, Koop SE, Bourke JL, Baker R. (2006) A nondimensional normalization scheme for oxygen utilization data. Gait Posture 24: 14–22. pmid:16112865
  45. 45. Brehm MA, Kempen JC, van der Kooi AJ, de Groot IJ, van den Bergen JC, Verschuuren JJ, et al. (2014) Age-related longitudinal changes in metabolic energy expenditure during walking in boys with Duchenne muscular dystrophy. PLoS One 9: e115200. pmid:25506914
  46. 46. Rogozinski BM, Davids JR, Davis RB III, Jameson GG, Blackhurst DW. (2009) The efficacy of the floor-reaction ankle-foot orthosis in children with cerebral palsy. J Bone Joint Surg Am 91: 2440–2447. pmid:19797580
  47. 47. Leteneur S, Simoneau E, Gillet C, Dessery Y, Barbier F. (2013) Trunk's natural inclination influences stance limb kinetics, but not body kinematics, during gait initiation in able men. PLoS One 8: e55256. pmid:23383128
  48. 48. Leteneur S, Gillet C, Sadeghi H, Allard P, Barbier F. (2009) Effect of trunk inclination on lower limb joint and lumbar moments in able men during the stance phase of gait. Clin Biomech (Bristol, Avon) 24: 190–195.
  49. 49. Desloovere K, Molenaers G, Van Gestel L, Huenaerts C, van Campenhout A, Callewaert B, et al. (2006) How can push-off be preserved during use of an ankle foot orthosis in children with hemiplegia? A prospective controlled study. Gait Posture 24: 142–151. pmid:16934470
  50. 50. Gestel Van L, Molenaers G, Huenaerts C, Seyler J, Desloovere K. (2008) Effect of dynamic orthoses on gait: a retrospective control study in children with hemiplegia. Dev Med Child Neurol 50: 63–67. pmid:18173633
  51. 51. Harper NG, Esposito ER, Wilken JM, Neptune RR. (2014) The influence of ankle-foot orthosis stiffness on walking performance in individuals with lower-limb impairments. Clin Biomech (Bristol, Avon) 29: 877–884.
  52. 52. Smits DW, van Groenestijn AC, Ketelaar M, Scholtes VA, Becher JG, Gorter JW. (2010) Selective motor control of the lower extremities in children with cerebral palsy: inter-rater reliability of two tests. Dev Neurorehabil 13: 258–265. pmid:20629592
  53. 53. Voorman JM, Dallmeijer AJ, Knol DL, Lankhorst GJ, Becher JG. (2007) Prospective longitudinal study of gross motor function in children with cerebral palsy. Arch Phys Med Rehabil 88: 871–876. pmid:17601467
  54. 54. van den Noort JC, Scholtes VA, Becher JG, Harlaar J. (2010) Evaluation of the catch in spasticity assessment in children with cerebral palsy. Arch Phys Med Rehabil 91: 615–623. pmid:20382296
  55. 55. Scholtes VA, Dallmeijer AJ, Becher JG. (2007) The effectiveness of multilevel botulinum toxin type A and comprehensive rehabilitation in children with cerebral palsy. In: VU University Medical Center, editor. The Spasticity Test: a clinical instrument to measure spasticity in children with cerebral palsy.Amsterdam.