Introduction

The ankle-brachial index (ABI) has been the foundation of non-invasive assessment of the arterial status of the lower limb for several decades. It is used not only to confirm or refute cases of suspected peripheral arterial disease (PAD), but also as a means of assessment of lower limb wound etiology and to determine if compression therapy is appropriate. Furthermore, the ABI has a burgeoning role as a tool for cardiovascular risk assessment as a substantial evidence base has demonstrated a relationship of abnormal ABI to adverse cardiovascular events1, 2; current guidelines (American College of Cardiologists/American Heart Association3; National Institute for Health and Care Excellence4) now recommend stringent cardiovascular risk reduction strategies for those with PAD as defined by an ABI0.9.

However, studies have demonstrated that the ABI is in fact infrequently and often incorrectly used in non-specialist healthcare settings such as primary care.5, 6 Lack of knowledge and skills to undertake the procedure using a handheld Doppler ultrasound probe and manual sphygmomanometer has been identified as a factor associated with this low use.7 In addition, the time consuming nature of Doppler ABI measurement and the need to rest subjects for at least 10 min before the procedure, also significantly limits its use in busy healthcare settings.5, 6

In recent years several automated ABI devices have been developed, which require minimal skills to use, and do not require a rest period before measurement, hence reducing the time needed to undertake the procedure. This is achieved by using multiple cuffs to measure arterial systolic pressures of all four limbs simultaneously. The majority of such automated devices use oscillometric technology to define the point when arterial occlusion occurs, the blood pressure at this point is then reported as the systolic pressure; while this method works satisfactorily for measuring brachial pressures, studies have shown that it becomes less accurate when measuring ankle pressures.8, 9

An alternative method of systolic blood pressure measurement is based on reperfusion plethysmography. A dual-chamber cuff applied to each limb consists of an upper occlusion chamber and a lower detection chamber. When the pressure of the upper occlusion chamber has exceeded arterial systolic pressure, the distal detection chamber detects a gradual decrease in limb volume as a result of blood redistribution in the absence of arterial blood inflow. As the pressure in the occlusion chamber is then incrementally reduced and reaches systolic pressure, arterial blood flow to the limb is restored, which is detected as a volume increase in the lower chamber. The pressure in the upper occlusion chamber at the point when this lower chamber volume increase occurs, is taken as the limb arterial systolic pressure.

The aim of this study was to compare the ABI measured with a Doppler ultrasound device (ABIDOP) with the ABI measured with an automated plethysmographic device (ABIPLE). An additional aim was to assess the time taken to measure the ABI via the two methods.

Methods

Data were combined from two studies (Figure 1); one was undertaken in a primary care setting and screened 368 individuals for PAD as defined by an ABI of 0.9. Participants had at least two risk factors for cardiovascular disease but no known cardiovascular disease or diabetes. The second study was undertaken in a secondary care vascular clinic and involved two health professionals measuring the ABIs of 12 individuals who had been referred with suspected PAD. Both studies were approved by the South East Wales Research Ethics Committee (REC No: 12/WA/0075, 12/WA/0242) and written informed consent was gained from each participant.

Figure 1
figure 1

Flow diagram illustrating diagnostic accuracy as per STARD (Standards for Reporting of Diagnostic Accuracy) standard.20 A full color version of this figure is available at the Hypertension Research journal online.

Each participant underwent bilateral ABI measurement using the plethysmographic device (Dopplex Ability (DA100PB), Huntleigh Healthcare, Cardiff, UK) and Doppler ultrasound (Doppler MD2, 8 MHz probe, Huntleigh Healthcare) with aneroid sphygmomanometery (Welch Allyn Inc, New York, NY, USA). The plethysmographic device was used in accordance with the manufacturers’ guidelines and was undertaken first as there was no need for a rest period before testing. The device measured the systolic pressures of all four limbs simultaneously before automatically calculating an ABI for each leg. In the event of a failed measurement, the procedure was repeated if this was acceptable to the participant and if the time schedule of the clinician permitted. The participant was then rested for 5 min before Doppler ultrasound ABI measurement, which was undertaken in accordance with the American Heart Association’s scientific statement for ABI measurement.10 All ABI measurements were undertaken by the same clinician (JHD), who is a registered nurse with significant experience in vascular assessment of the lower limb. JHD was blinded to the results of the automated device to ensure this did not influence the subjective Doppler measurements.

The time taken to perform bilateral ABI measurement for each participant was recorded; for the automated plethysmographic device this included the time taken to apply the cuffs to each limb, and for the Doppler method this included a 10 min rest period before the actual measurement procedure. Equipment used in study measurements was calibrated and maintained in accordance with the manufacturer’s guidelines.

Statistical analysis

Statistical analysis was undertaken using the IBM SPSS software (version 21, New York, NY, USA). Data are presented as mean with s.d. unless otherwise stated. The agreement between the two methods of ABI measurement was assessed by visual inspection of equality plots and Bland–Altman plots with calculation of 95% confidence interval of agreement.11 Sensitivity, specificity, positive predictive values and negative predictive value of the plethysmographic device were calculated, using the Doppler measurements as the reference standard. Data were also assessed to determine if there were any factors which (i) were associated with the automated device failed measurements and (ii) affected the agreement between the two methods; as continuous data were not normally distributed, this was achieved via the use of the Kruskal–Wallis test.

Categorical data were assessed via use of the χ2 test. A Wilcoxon signed rank test was used to assess the time taken to perform the ABI measurements. Significance was set at P<0.05.

Results

The flow chart of the study is shown in Figure 1. The characteristics of the study population according to each ABI measurement undertaken are shown in Table 1.

Table 1 Population demographics according to each limb ABI measurement

Failed measurements

Although an ABIDOP reading was attained for all remaining 724 measurements, there were 28 less ABIPLE measures. These failed measurements were associated with the presence of hypertension (P=0.015; Table 1). ABIDOP ranged from 0.44 to 1.51 (1.11±0.14 mean±s.d.), whereas the ABIPLE ranged from 0.47 to 1.49 (1.10±0.14). The prevalence of limb measurements indicating PAD (ABIDOP0.9) was 5.7% (n=41 limbs; Figure 2).

Figure 2
figure 2

PAD cases: distribution of ABIs. ABI, ankle-brachial index; PAD; peripheral arterial disease.

Agreement between the two measurement techniques

The equality plot of the two measurement methods is shown in Figure 3. The mean difference, or bias was 0.016±0.1, and the absolute mean difference was (0.07±0.07), 95% limits of agreement were ±0.2 (Figure 4). The ABIPLE was higher than the ABIDOP in 52% (n=361) of measurements, lower in 41% (n=286), and the same for the remaining 7% (n=49).

Figure 3
figure 3

Equality plot showing agreement between ABIDOP and ABIPLE. The identity (thin line) and regression (thick line) lines are shown. For the regression line the 95% CI limits are shown as dotted lines, y=0.35+0.697x, r=0.74. ABI, ankle-brachial index; CI, confidence interval.

Figure 4
figure 4

Bland–Altman plot showing agreement between ABIDOP and ABIPLE. ABI, ankle-brachial index.

There was a gradual increase in bias across the PAD status groups as the reference standard (Doppler ABI) increased (P<0.01) (Figure 5). In addition, participants with hypertension had a five times greater positive bias than participants who were not hypertensive (0.032 vs. 0.006; P=0.018; Figure 6), and similarly, the 95% limits of agreement were greater for the hypertensive group than the normotensive group (±0.23 vs. ±0.21, P<0.01).

Figure 5
figure 5

Study bias according to PAD status. A Kruskal–Wallis H-test (X2(3)=16.12) showed significant differences (P=0.001) between groups. A Bonferroni post hoc test revealed that the PAD group* were significantly different from the normal ABI group (P=0.04) and High ABI group (P<0.001). The Borderline PAD group** were significantly different from the High ABI group (P=0.045). ABI, ankle-brachial index; PAD; peripheral arterial disease.

Figure 6
figure 6

Study bias according to hypertensive status. Mann–Whitney U-test: U=39742.5, z=−2.363, P=0.018

The sensitivity of the plethysmographic device for diagnosis of PAD, as defined by an ABIDOP0.9, was 70%, specificity 96%, positive predictive value 52%, negative predictive value 98% with an overall accuracy 94%. Analysis of a receiver operator characteristic curve (Figure 7) revealed an area under the curve of 0.96 (95% confidence interval: 0.94–0.98, P<0.001). The optimal ABIPLE cutoff point for diagnosis of PAD (as defined by ABIDOP0.9) is 1.04, which provides a sensitivity of 98% and specificity of 75%.

Figure 7
figure 7

Receiver operating characteristic (ROC) curve for plethysmographic device in diagnosing PAD as defined by ABIDOP0.9. Area under curve=0.96 (95% CI: 0.94–0.98, P<0.001). ABI, ankle-brachial index; CI, confidence interval; PAD; peripheral arterial disease.

The ABIPLE measurement procedure was faster (P<0.01) than the ABIDOP measurement procedure (which included the rest period), at 7 min 55 s (±1:29) and 17 min 45 s (±1:05), respectively.

Discussion

Data suggest that the plethysmographic device has only moderate sensitivity (70%) for detection of PAD. However, this could be partly attributed to the fact that a large proportion of PAD cases within the sample (44%) had mild PAD (ABI range: 0.85–0.9; Figure 2) meaning that if the ABIPLE was only between 0.01 and 0.05 units higher than the ABIDOP, then a false negative outcome was concluded. It is therefore more useful to consider the receiver operator characteristic curve (Figure 7) to gain more insight into the accuracy of the plethysmographic device. An area under the curve of 0.96 indicates a high degree of accuracy in comparison to the Doppler method as a gold standard. The optimal cutoff point for diagnosis of PAD was <1.04, which is considerably higher than the threshold of 0.9 that is traditionally used for Doppler ABI measurements. However, this appears to be a common finding associated with the use of automated ABI devices; a systematic review and meta-analysis of 25 studies, which assessed the usefulness of oscillometric devices for ABI estimation compared with the conventional Doppler method also concluded that to increase the sensitivity for PAD, a higher threshold ABI<1.0 might be preferable.9 The implications of concomitant reduced specificity must, of course, be balanced against this improved sensitivity, as it would result in more patients being misdiagnosed as having PAD or more patients being referred for confirmatory tests such as duplex ultrasound. Both of these outcomes have ethical and financial repercussions.

Table 2 summarizes the results of existing studies that examined the agreement of automated ABI devices with Doppler ultrasound. Included devices were restricted to those which are specifically designed for ABI measurement, which are also portable and designed for use in non-specialist settings; larger devices designed for use within vascular laboratories were not included. Two devices (Vascular Explorer and Vicorder12) functioned on the principle of photo-plethysmography, whereas the remaining all used oscillometry to measure and calculate the ABI. Comparison of data is hindered by study heterogeneity, reporting omissions and differing reporting strategies. It can be seen that 95% limits of agreement of the Dopplex Ability are either narrower or in the same range as the reported limits of oscillometric devices.

Table 2 Agreement of automated ABI devices with Doppler ultrasound as the reference standard

The failed measurement rate (3.9%) of the automated device is lower than the reported failed measurement rate pertaining to the BOSO ABI device (24%13 and 9%14), but slightly higher than the failed measurement rate of the Watch BP office device (1.6%15 and 2.5%16). Notably, the ranges of ABIs measured by the Dopplex Ability and the Vascular Explorer and Vicorder devices12, which also function on the principle of plethysmography, were greater than the range measured by either of the oscillometric devices (Table 2).

Analysis of the Bland–Altman plot (Figure 4) demonstrated a higher level of agreement between the two measurement techniques at the lower end of the ABI spectrum and this was shown to be statistically significant (Figures 5, P<0.01). In contrast, oscillometric devices have often been shown to have lower levels of agreement with the Doppler ABI at the lower end of the ABI spectrum.9, 17

The agreement between the two methods was also found to be less in the presence of hypertension (P=0.018). Few studies have examined which factors affect the agreement between automated ABI devices and the Doppler method. A study by Takhashi et al. examined whether gender, age, smoking, alcohol, cardiovascular disease, diabetes, hyperlipidaemia and body mass index were associated with differences between oscillometric and the Doppler method in 113 subjects aged 61–88; they concluded that there were none.17 However, Wohlfahrt et al. measured the ABIs of 839 subjects aged >25 years using the Doppler method and an oscillometric device and noted that the difference between the methods increased significantly with increasing mean ABI (r=0.29, P<0.001).14

Important potential benefits of automated ABI systems include their negation of the need for both operator skill and a rest period before the measurement procedure, hence potentially making PAD screening more feasible and amenable. The utility of these devices could, however, be limited by financial considerations, with costs ranging from two to six times more than the typical cost of a handheld Doppler and sphygmomanometer. Several studies have used oscillometric devices that were designed for measuring brachial blood pressures to also measure ankle pressures, which were then used to calculate the ABI.8, 18 Such devices are not only readily available in the majority of healthcare settings but also cost considerably less than automated ABI systems reviewed in Table 2. However, a systematic review by Verbeck et al. found that such devices showed significantly higher (P<0.01) ABI differences from the Doppler value than the automated devices than were specifically designed for ABI measurement.8 Furthermore, as these standard blood pressure devices do not allow for simultaneous limb blood pressure measurements, the need to rest the patient before the procedure would remain, hence reducing any time savings associated with automated measurements.

Strengths and limitations

Although the study consisted of a large sample, only a small proportion (5.7%) of participants were found to have PAD. Hence, while there has been adequate investigation of how the plethysmographic device performs when measuring the ABIs of healthy subjects, further research is needed to assess its performance in severely diseased populations. The fact that a single clinician undertook all ABI measurements within the study meant that there were no issues relating to inter-observer variability. However, as the clinician had extensive experience of using the automated device, it is possible that she was able to obtain more accurate results. Furthermore, the subjective nature of Doppler ABI measurements results in susceptibility to observer bias and error, hence perhaps making them an imperfect reference standard in any case. Some authors have addressed this issue by using a superior diagnostic modality such as Duplex ultrasound or computed tomography angiography as the reference standard for such comparative studies.19

Perspectives

There are now more reasons than ever for generic, non-specialist healthcare professionals in settings such as primary care, to measure the ABIs of their patients. Compared with oscillometric ABI devices, the plethysmographic device used here appears to provide improved performance at the lower end of the ABI spectrum. These findings replicate the results of existing studies12 of photo-plethysmographic devices, hence suggesting that plethysmography may provide a superior measurement principle for the measurement of low ankle systolic pressures.

Further research is needed to determine if this plethysmographic device can provide sufficient diagnostic accuracy for diagnosis of PAD as a stand alone test. In the meantime, the device can be used as a fast and amenable method of identifying those patients who require further arterial assessment in the form of Doppler ABI measurement, Duplex scan or specialist vascular review. Results have suggested that the higher threshold of ABI<1.04 should be used for this purpose.