Introduction

Sudden cardiac death is a major cause of mortality, often the result of heart rhythm disturbances like ventricular tachycardias (VT) or ventrical fibrillations (VF). An increasingly important treatment for patients with these life-threatening arrhythmias is the automatic implantable cardioverter-defibrillator (ICD). This small, lithium battery-operated electronic device is surgically placed under the skin, usually near the left collar bone. One or more flexible insulated wires (leads) run from the ICD through the veins to the heart (Figs 1, 2). The ICD monitors the heart rate continuously and when a VT or VF is detected, a precisely calibrated electrical shock is delivered to terminate the arrhythmia and restore a normal heart rate. The use of this device has cut the death rate of patients with a history of a life-threatening ventricular arrhythmia considerably.1

Figure 1
figure 1

An ICD is surgically placed under the skin, usually near the left collar bone (courtesy of St. Jude Medical)

Figure 2
figure 2

An implanted ICD on X-ray (courtesy of P.H. Zaadstra and Walcheren Hospital)

As ICDs are devices with elaborate electrical signals, they are potentially sensitive to electromagnetic interference (EMI) from external sources which may satisfy arrhythmia criteria. Modern life exposes us all to an increasing number of potential sources of EMI. Manufacturers try to protect the ICDs from EMI as much as possible, using features like titanium casing, signal filtering, interference rejection circuits, noise reversion function and programmable parameters.2 However, EMI may still potentially suspend detection of VT and VF, and the undesired triggering of a shock is unpleasant or even life-threatening for the patient.

In a follow-up study of 341 patients with an ICD, EMI occurred five times in four patients (0.75 % per patient year).3 In vitro testing indicated that mobile phones could interfere with two ICDs, but during the in vivo phase of the study none of the 41 patients was affected.4 Large welding machines and electric motors did not interfere with the normal functioning of the ICDs tested,5 although one patient has been described where interference between a washing machine and an ICD resulted in inappropriate shock delivery.6

Dental equipment has been established as a source of light to moderate electromagnetic fields7 and some devices have been reported to interfere with the activity of pacemakers.8 Consequently, there is concern that EMI from dental equipment may also interfere with the correct function of ICDs. In particular, electrocautery is suspected to be a medical source of EMI.9,10,11 To our knowledge, no previous studies have been performed on the possible effects of dental devices on ICDs. Therefore, we investigated in vitro whether EMI with ICDs occurs during the operation of electrical dental equipment.

Methods

During the first part of the study, several examples of commonly used dental devices (Table 1) were tested for their ability to interfere with the correct function of an ICD (Guidant Vitality 2 EL type VR model T177; Fig. 3). The ICD and connected leads were placed in a plastic container with 1.5 L physiological saline to simulate the electrical resistance of the human body.8 ICD output was monitored by real-time telemetry with a Guidant Zoom 2920 programmer. All dental devices were operated continuously, as well as turned on and off for 90 seconds at intervals of 2.5 cm both parallel to and in-line with the ICD.

Table 1 Dental devices tested during the first part of the study
Figure 3
figure 3

The ICDs tested in this study

Since real-time telemetry between the ICD and the programmer may be compromised by EMI, even when the device function remains normal (Pinski and Trohman, 2002), we performed an additional off-line experiment. Three ICDs (Guidant Vitality 2 EL VR T177, Medtronic Marquis DR Model 7274 and Medtronic Marquis VR Model 7230; Fig. 3) were placed in a simulator (HKP DFA-2DC with HKP ARSI-2GA) to simulate the heart rate of the patient. Three different ultrasonic bath cleaners (Table 2) were operated continuously as well as on/off for 90 seconds as described above, after which the stored electrograms were reviewed off-line by telemetry with a Guidant Zoom 2920 programmer or a Medtronic Vitratron programming head. All experiments with dental devices were performed in triplicate.

Table 2 Ultrasonic bath cleaners tested during the second part of the study

Results

During the first part of the study, the dental chair and the attached electrosurgical unit, both handpieces, the ultrasonic tooth scaler and both amalgamators all failed to produce EMI at a distance of 2.5 cm. In contrast, the Branson 200 ultrasonic bath cleaner produced EMI up to a distance of 12.5 cm both during continuous use and intermittent operation. During operation of the ultrasonic bath cleaner, the ICD detected several episodes of ventricular rate noise. In addition, episodes of interference with the telemetry connection between ICD and programmer were also observed ('telemetry noise').

In the subsequent study, using stored electrograms, this ultrasonic bath cleaner also produced EMI in another ICD (Marquis DR Model 7274 ICD) during one of the three test runs. The other two types of ultrasonic bath cleaners tested failed to induce EMI in any of the ICDs tested (Guidant Vitality 2 EL VR T177, Medtronic Marquis DR Model 7274, Medtronic Marquis VR Model 7230).

Discussion

To our knowledge, this in vitro study is the first to investigate EMI with ICD during operation of dental equipment. The findings in this study indicate that EMI generated by most dental equipment does not interfere with normal functioning of the ICDs. The single exception was a Branson 200 ultrasonic bath cleaner, which induced EMI at a short distance in both the Guidant Vitality 2 EL VR T177 and the Medtronic Marquis DR Model 7274. The EMI resulted in several episodes of ventricular noise in the ICD. Although the observed ventricular signal did not satisfy VT of VF criteria of the ICD, it may temporarily suspend the correct ventricular arrhythmia detection. Since this effect was also observed in stored electrograms, it is not an effect on the telemetry between the ICD and the programmer.

The suggestion that some ultrasonic bath cleaners are a potential source of EMI is supported by the observation that two other types of ultrasonic bath cleaners inhibited the correct function of pacemakers up to a distance of 30 cm.8 Precaution in using ultrasonic bath cleaners around ICD patients seems advisable. However, this may be more of concern to dental healthcare workers with an implanted ICD, since ultrasonic bath cleaners are usually located more than 1 m from patient-care areas.

Conclusion

The findings of this study suggest that dental equipment, commonly used in the dental office, does not have significant effects on the ICDs tested. This is certainly the case when the general precaution is followed to keep electrical appliances at least 10-15 cm away from an ICD and its leads.10,12,13