Surgery for congenital heart disease
Predictors affecting durability of epicardial pacemaker leads in pediatric patients

https://doi.org/10.1016/j.jtcvs.2007.09.002Get rights and content
Under an Elsevier user license
open archive

Objectives

Despite pacemaker therapy in children and adolescents favoring an initial epicardial approach, predictors of lead failure have not been well clarified. The aim of this study was to assess the long-term outcomes and to determine predictors affecting lead durability in pediatric pacing therapy.

Methods

We reviewed the outcomes of 109 consecutive pacing leads implanted in 55 patients (median age, 5.2 years; range, 31 days–15.8 years), including 38 atrial and 71 ventricular leads. They consisted of 58 (53%) fishhooks, 37 (34%) screw-in leads, and 14 (13%) steroid-eluting suture-on leads. Seventy (64%) were implanted in patients with structural heart disease.

Results

The leads were followed for a median of 6.4 years (range, 3 days–22.9 years). Lead failure occurred in 29 leads (27%; median of 8.4 years after implantation). Exit block or elevation of pacing threshold was the most common cause (n = 18), but failures did not directly cause patient death. The overall 1-, 5-, 10-, and 15-year lead survivals were 100%, 89.0%, 72.5%, and 55.5%, respectively. Multivariate Cox analysis revealed concurrent structural heart disease (relative risk, 2.85; 95% confidence interval, 1.27–6.42; P = .011) to be the only significant predictor of lead failure.

Conclusions

Epicardial leads provide a reliable technique for managing rhythmic disturbance problems in the pediatric population. The only significant predictor of lead failure is the presence of structural heart disease.

CTSNet classification

20
21
24

Abbreviations and Acronyms

S-CHD
structural congenital cardiac disease

Cited by (0)