Clinical Research
Safety Issue
Safety of Transvenous Cardiac Resynchronization System Implantation in Patients With Chronic Heart Failure: Combined Results of Over 2,000 Patients From a Multicenter Study Program

https://doi.org/10.1016/j.jacc.2005.08.031Get rights and content
Under an Elsevier user license
open archive

Objectives

The purpose of this study was to evaluate the safety of implanting a cardiac resynchronization therapy (CRT) system.

Background

Clinicians and patients require data on the safety of the CRT implant procedure to estimate procedural risk.

Methods

We evaluated outcomes of transvenous CRT system implantation in 2,078 patients from the Multicenter InSync Randomized Clinical Evaluation (MIRACLE) study, the MIRACLE Implantable Cardioverter-Defibrillator (ICD) study, and the InSync III study. We compared the MIRACLE study to the InSync III study and the MIRACLE ICD study randomized phase to its general phase to evaluate the effect of new technologies.

Results

The implant attempt succeeded in 1,903 of 2,078 (91.6%) patients. Implant time decreased from 2.7 h in the MIRACLE study to 2.3 h in the InSync III study (p < 0.001), and from 2.8 h in the MIRACLE ICD study randomized phase to 2.4 h in the general phase (p < 0.001). The implant procedure produced 62 perioperative complications in 53 (9.3%) MIRACLE trial patients; 159 in 135 (21.1%) MIRACLE ICD study randomized phase patients and 71 in 62 (13.9%) general phase patients (p < 0.05 vs. randomized); and 41 in 37 (8.8%) InSync III study patients (p = NS vs. the MIRACLE study). We observed 73 postoperative complications in 62 (11.7%) MIRACLE trial patients, 77 in 68 (11.9%) MIRACLE ICD study randomized phase patients and 56 in 45 (11.0%) general phase patients (p = NS), and 37 in 34 (8.6%) InSync III study patients (p = NS). A total of 8% of patients required reoperation to treat lead dislodgement, extracardiac stimulation, or infection during follow-up.

Conclusions

Transvenous CRT system implantation appears safe, well-tolerated, has a high success rate, and improves with operator experience and the addition of new technologies.

Abbreviations and Acronyms

AERC
Adverse Events Review Committee
CRT
cardiac resynchronization therapy
CS
coronary sinus
ICD
implantable cardioverter-defibrillator
LV
left ventricle/ventricular
MIRACLE
Multicenter InSync Randomized Clinical Evaluation
NYHA
New York Heart Association
RA
right atrial/atrium
RV
right ventricle/ventricular
SCD
sudden cardiac death

Cited by (0)

This study was supported by Medtronic, Inc., Minneapolis, Minnesota.