Volume 20, Issue 1, August 2004, Pages 20-25
Gait & Posture

Reliability of the GAITRite® walkway system for the quantification of temporo-spatial parameters of gait in young and older people

https://doi.org/10.1016/S0966-6362(03)00068-7Get rights and content

Abstract

The purpose of this study was to evaluate the test–retest reliability of an instrumented walkway system (the GAITRite® mat) for the measurement of temporal and spatial parameters of gait in young and older people. Thirty young subjects (12 males, 18 females) aged between 22 and 40 years (mean 28.5, S.D. 4.8) and 31 older subjects (13 males, 18 females) aged between 76 and 87 years (mean 80.8, S.D. 3.1) walked at a self-selected comfortable walking speed across the pressure-sensor mat three times and repeated the process approximately 2 weeks later. Intra-class correlation coefficients (ICC), coefficients of variation (CV) and 95% limits of agreement were then determined. For both groups of subjects, the reliability of walking speed, cadence and step length was excellent (ICCs between 0.82 and 0.92 and CVs between 1.4 and 3.5%). Base of support and toe in/out angles, although exhibiting high ICCs, were associated with higher CVs (8.3–17.7% in young subjects and 14.3–33.0% in older subjects). It is concluded that the GAITRite® mat exhibits excellent reliability for most temporo-spatial gait parameters in both young and older subjects, however, base of support and toe in/out angles need to viewed with some caution, particularly in older people.

Introduction

Measurements of temporal and spatial parameters of gait are commonly used for the identification of gait disorders [1] and for the evaluation of therapeutic interventions such as exercise [2], [3]. A variety of techniques have been used for this purpose, including simple visual observation [4], [5], [6], stopwatches [7], [8] and paper walkways [9], [10], [11]. More recently, sophisticated gait analysis tools have become commercially available, including the in-shoe Clinical Stride Analyser® [12], [13] and the GAITRite® mat [14], [15], [16], [17]. Although the reliability of the Clinical Stride Analyser® is well established [12], [13], there is limited information available pertaining to the reliability of the GAITRite® system.

The GAITRite® mat is a portable walkway embedded with pressure sensors that detect footfalls as the subject walks the length of the mat. The software enables the documentation of a wide range of temporo-spatial gait parameters, including walking speed, cadence, step length, base of support and foot placement angles. A number of studies have been performed to evaluate the validity of these measurements against existing techniques. A single case study by McDonough et al. [14] evaluated the concurrent validity of the GAITRite® mat against chalk footsteps and a hand-held stopwatch. The results revealed that while there was good agreement for spatial variables between the GAITRite® mat and the chalk footprints, the association between the GAITRite® mat temporal parameters and timed measures with the stopwatch was somewhat lower. The authors attributed this poor association to the subjectivity involved in timing gait events with a stopwatch. A similar investigation by Selby-Silverstein and Besser [15] compared the GAITRite® mat to powdered footprints and an in-shoe pressure measurement tool (the Parotec® system), and reported moderate correlations for temporal variables. Cutlip et al. [16] compared the GAITRite® mat to the Peak Performance Technologies Motus 3.1® system, and reported strong associations between the systems for all parameters, however, significant differences between the systems were found for the measurement of step length and stride velocity. Recently, Bilney et al. [17] reported very high correlations between the GAITRite® mat and the Clinical Stride Analyzer®, and concluded that the GAITRite® system is a valid tool for gait analysis.

The reliability of the GAITRite® system has received only limited attention. Gretz et al. [18] reported good immediate retest and 2-week retest reliability (Intra-class correlation coefficients (ICCs)>0.75) for temporo-spatial measurements of normal adults and adults with Down's syndrome, but noted that base of support measurements were more variable than other parameters in both groups. Bilney et al. [17] reported high intraclass correlation coefficients for normal and fast-walking measurements obtained from three repeat trials recorded on the same day in 25 healthy subjects aged 21–71 years, however, the measurements were more variable between trials when subjects walked at a slow speed.

Although these results indicate that the GAITRite® mat provides both valid and reliable measurements of temporo-spatial gait parameters, the test–retest reliability of the system has not been evaluated in older people [17]. As ageing is associated with increased variability in gait patterns, particularly in elderly people who fall [19], [20], it is necessary to establish whether measurements obtained on different days using the GAITRite® mat are repeatable to determine whether the technique can be used to measure the effects of interventions in this population. Therefore, the aim of this study was to evaluate the test–retest reliability of GAITRite® system in healthy young subjects in addition to a sample of community-dwelling older people by obtaining measurements on two separate occasions, approximately 2 weeks apart.

Section snippets

Subjects

A convenience sample of thirty healthy young subjects (12 males, 18 females) aged between 22 and 40 years (mean 28.5, S.D. 4.8) were recruited from the staff and student population of the Prince of Wales Medical Research Institute. All were free of musculoskeletal or neurological pathology. Thirty-one community-dwelling people (13 males, 18 females) aged between 76 and 87 (mean 80.8, S.D. 3.1) were recruited from an existing database developed for previous research studies by our group. Subject

Differences between young and older subjects

The older subjects were significantly shorter than the younger subjects (t59=−2.65, P=0.01), see Table 1. A comparison of gait parameters between young and older subjects is shown in Table 2. Older subjects walked with a significantly slower velocity (t59=−5.77, P<0.001), and shorter step lengths for both right (t59=−7.18, P<0.001) and left limbs (t59=−7.43, P<0.001). However, there were no differences between young and older subjects for cadence, base of support or toe in/out angles.

Reliability

Mean

Discussion

Reliability is a fundamental requirement for any gait analysis system to ensure that any observed differences in gait performance between testing sessions reflect real changes in locomotor function, rather random or systematic error in the measurement technique. This study has shown that the GAITRite® mat generally provides highly reliable measurement of temporo-spatial gait parameters when young and older subjects are tested 2 weeks apart, and that the system can be used with confidence to

Conclusion

This study has shown that the GAITRite® produces highly reliable measurements of gait parameters, with the possible exception of base of support and toe in/out angle in older subjects, when measured on two occasions approximately 2 weeks apart. The system can be used with confidence to evaluate the effects of various interventions on walking speed, cadence and step length. The reliability of base of support and toe in/out angle measurement, though lower than the other parameters, may be

Acknowledgements

This project was funded by the NHMRC Health Research Partnerships Grant: Prevention of Injuries in Older People (Grant No. 209799) and an equipment grant from the Perpetual Foundation of Australia.

References (34)

  • D.E Krebs et al.

    Reliability of observational kinematic gait analysis

    Phys. Ther.

    (1985)
  • M Saleh et al.

    In defence of gait analysis. Observation and measurement in gait assessment

    J. Bone Jt. Surg.

    (1985)
  • J.C Wall et al.

    Use of a multimemory stopwatch to measure the temporal gait parameters

    J. Orthop. Sports Phys. Ther.

    (1997)
  • J.W Youdas et al.

    Measurements of temporal aspects of gait obtained with a multimemory stopwatch in persons with gait impairments

    J. Orthop. Sports Phys. Ther.

    (2000)
  • B Clarkson

    Absorbent paper method for recording foot placement during gait: suggestions from the field

    Phys. Ther.

    (1983)
  • D.K Heitmann et al.

    Balance performance and step width in noninstitutionalized, elderly female fallers and non-fallers

    Phys. Ther.

    (1989)
  • A.R Bird et al.

    The effect of pregnancy on footprint parameters

    J. Am. Podiatr. Med. Assoc.

    (1999)
  • Cited by (0)

    View full text