Abstract

Background

Diabetes mellitus (DM) patients show higher rates of repeat revascularisation even in the era of modern drug-eluting stents (DES). The concept of bioresorbable scaffolds is becoming captivating, as it might allow for repeat interventions, prolonging the time span during which patients can be treated by percutaneous coronary intervention (PCI).

Aims

We intend to evaluate the short- and long-term safety and efficacy of Absorb bioresorbable vascular scaffolds (Absorb BVS) in the treatment of coronary artery disease (CAD) in DM patients for any indication.

Methods

The ABSORB DM Benelux is an international prospective study in DM patients who have undergone PCI with ≥1 Absorb BVS. Major adverse cardiac events (MACE) at 1 year was the primary endpoint, defined as a composite of all-cause death, any myocardial infarction (MI) and ischaemia-driven target vessel revascularisation (TVR). Secondary endpoints were target lesion failure (TLF) and definite or probable scaffold thrombosis (ScT).

Results

Between April 2015 and March 2017, 150 DM patients and 188 non-complex lesions were treated. Device implantation was successful in 100%. MACE occurred in 14 (9.5%) patients, with all-cause death occurring in 4 (2.7%), any MI in 6 (4.1%) and ischaemia-driven TVR in 7 (4.8%) respectively. TLF was reported in 11 (7.5%). Definite and probable ScT was observed in 2 (1.4%).

Conclusion

Absorb BVS for treatment of anatomically low-risk patients with DM show acceptable safety and efficacy outcomes at 1 year. If these promising results are confirmed after a longer follow-up period, new-generation bioresorbable scaffolds combined with refinement of implantation techniques might open new horizons for CAD treatment in DM patients.
Opmerkingen

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1007/​s12471-019-1293-7) contains supplementary material, which is available to authorised users.
Trial ID: NTR5447

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author, T.M. Hommels, on reasonable request.
What’s new?
  • This is the first prospective study utilising Absorb bioresorbable vascular scaffolds for treatment of anatomically low-risk patients with diabetes mellitus for any indication.
  • Acceptable safety and efficacy outcomes were obtained at 1‑year follow-up.
  • A low incidence of scaffold thrombosis was observed with no occurrences of late thrombosis.
  • If these promising results are confirmed after a longer follow-up period, more performant bioresorbable scaffolds might open new horizons for treatment of coronary artery disease in diabetes mellitus patients.

Introduction

The incidence and prevalence of diabetes mellitus (DM) is increasing in both developed and developing countries [1, 2]. People with DM are between 2 and 4 times more likely to develop coronary artery disease (CAD) compared to non-DM patients [36]. Despite the major advances in percutaneous coronary intervention (PCI) with newer generations of drug-eluting stents (DES) accompanied by improved medical treatment, studies have continued to show a trend towards higher rates of major adverse cardiovascular events in DM patients compared to non-DM patients [7, 8].
Bioresorbable polymer drug-eluting scaffold systems enable the application of a short-term vessel scaffold (which subsequently dissolves) combined with drug delivery capability. The short-term results from the Absorb bioresorbable vascular scaffolds (Absorb BVS) clinical trial programs, at the time the study was designed, showed promising safety and efficacy outcomes for these devices, being non-inferior to those of the best-in-class durable polymer everolimus-eluting stents (EES) [915]. Considering the scaffold resorption, it was conceivable that vessel restoration following Absorb BVS implantation might be associated with more favourable long-term outcomes compared to metallic DES, mainly because inflammation induced by foreign bodies is only transient following Absorb BVS implantation while it is permanent after metallic DES implantation. Particularly in DM patients, where diabetes-related chronic peristrut inflammation triggers more aggressive restenosis, reduction of permanent inflammatory triggers may further improve clinical outcomes. Furthermore, repeat interventions at target lesions could be effectuated multiple times without a critical loss of vessel diameter, thus prolonging the time interval during which CAD in DM patients could still be managed by PCI. Henceforth, we designed the ABSORB DM Benelux Study to evaluate the short- and long-term safety and efficacy of the Absorb BVS in patients with DM.

Methods

The ABSORB DM Benelux Study is an international study in patients with DM and de novo lesions treated with the ABSORB family and conducted in The Netherlands, Belgium and Luxembourg. This clinical investigation is a prospective register and did not test any new device. The study was approved in February 2015, in accordance with the Declaration of Helsinki, by the Ethical Committees of each participating centre.

Study population

All patients aged ≥18 years with a history of DM undergoing PCI with implantation of ≥1 Absorb BVS for any indication, in a de novo lesion located in a native non-grafted artery, could be enrolled. The exclusion criteria were determined as: pregnancy; patients unable to provide (written) informed consent; known left ejection fraction <30%; life expectancy <3 years and inability to undergo dual antiplatelet therapy (DAPT) for at least 12 months.

Endpoints and definitions

The primary endpoint was the incidence of major adverse cardiac events (MACE) at 1 year, defined as a composite of all-cause death, any myocardial infarction (MI) and ischaemia-driven target vessel revascularisation (TVR). The secondary endpoints represented target lesion failure (TLF), defined as a composite of cardiac death (CD), target vessel MI and ischaemia-driven target lesion revascularisation (TLR), the incidence of definite or probable scaffold thrombosis (ScT) and the 1‑year incidence rates of the endpoint composites. Adverse event definitions are described in the Supplementary Table.

PCI procedure

The implanted devices are the bioresorbable polymer drug-eluting scaffold ABSORB BVS system and the ABSORB GT1 system (Abbott Vascular, Santa Clara, CA, USA). These devices are composed of poly-L-lactic acid and an everolimus-eluting polymer coating of poly-dl-lactic-acid, both of which are completely bioresorbable through a natural metabolic process within 3 years [16]. The average strut thickness is 150 µm. The device was available in diameters ranging from 2.5 to 3.5 mm with a length of 8, 12, 18, 24, or 28 mm. Implantation of an Absorb BVS was at the discretion of the operator. The vessel size, similar to other trials with this particular device, ranged from 2.50 to 3.75 mm. Predilatation and postdilatation were strongly recommended. Intracoronary imaging by means of optical coherence tomography (OCT) or intravascular ultrasound (IVUS) was encouraged but not mandatory. Treatment of bifurcations was not encouraged; however, in this case a provisional T‑stenting technique was advised. There were no limitations regarding lesion length; however, treatment of very calcified and tortuous lesions was not encouraged. If found necessary, additional implantation of metallic DES was accepted as a bailout procedure. Angiographic success was defined as a visually assessed <30% residual stenosis of the target lesion after successful device implantation. Procedural success was defined as angiographic success with no occurrence of events during the procedure. All patients received DAPT for at least 12 months.

Follow-up and assessment of adverse events

Clinical follow-up included clinical visits and telephone contact. All reported adverse cardiac events underwent assessment by an independent clinical event committee (Diagram BV, Zwolle, The Netherlands). Angiographic evaluations of baseline as well as repeat angiograms in patients with events were analysed by means of quantitative coronary angiography, Thrombolysis in Myocardial Infarction (TIMI) as well as presence of thrombus analysis by an independent core laboratory (Diagram BV, Zwolle, The Netherlands).

Statistical analysis

The baseline clinical and angiographic characteristics are presented using descriptive statistics. Categorical variables are summarised as frequency and percentages. Continuous variables are summarised as mean and standard deviation. The composite endpoints and clinical events are presented by using the Kaplan-Meier survival method with time-to-event analysis. In addition, a multivariate Cox regression model with adjustment for age, gender, PCI indication (acute coronary syndrome versus non-acute coronary syndrome) and insulin-treated DM was performed and presented with hazard ratio (HR) and 95% confidence interval (CI). Other regression models were performed for relevant factors such as multivessel or multiple lesion treatment (≥2 vessels/lesions), proximal versus distal segment implantation (proximal coronary location—segment number 1, 5, 6, 11 versus none), number of devices used for target lesion, total length of implanted devices, employment of intracoronary imaging and utilisation of pre- and postdilatation. A p-value <0.05 was considered to indicate formal statistical significance. The analyses were conducted with intention-to-treat. All statistical analyses were performed using SPSS version 25 (IBM Corp., Armonk, NY, USA).

Results

Between April 2015 and March 2017, a total of 150 DM patients and 188 lesions were treated by PCI with implantation of Absorb BVS. The patients were treated in 18 different centres by experienced Absorb BVS operators. Their baseline clinical characteristics are shown in Tab. 1. The patients had a mean age of 63.4 ± 10.4 years and were predominantly male (72%). DM type II was diagnosed in 93.3% and insulin-treated DM at hospital admission was 31.3%.
Table 1
Clinical characteristics of the patients at baseline
Baseline clinical characteristic
Patients (n = 150)
Age (years)—mean ± SD
 64.3 ± 10.4
Sex (male)—n (%)
108 (72.0)
Race (Caucasian)—n (%)
140 (93.3)
Body-mass index (kg/m2)—mean ± SD; n
 29.5 ± 5.1;148a
Risk factors—n (%)
 
Diabetes mellitus type 1
 10 (6.7)
Diabetes mellitus type 2
140 (93.3)
Insulin-dependent diabetes mellitus
 47 (31.3)
Diabetes mellitus treated with oral antidiabetic
117 (78.0)
HbA1c at hospitalisation (mmol/mol)—mean ± SD; n
 55.5 ± 11.5;42a
Arterial hypertension
104 (69.3)
Hypercholesterolaemia
100 (66.7)
Family history of cardiovascular disease
 59 (39.3)
Current smoker
 35 (23.3)
Medical history—n (%)
 
Previous ACS
 41 (27.3)
Previous PCI
 37 (24.7)
Previous CABG
  8 (5.3)
Previous CVA or TIA
 10 (8.7)
Severe chronic renal failureb
  4 (2.7)
Chronic pulmonary obstructive diseasec
 11 (7.3)
Clinical presentation—n (%)
 
Acute coronary syndrome
 73 (48.7)
ST-segment elevation myocardial infarction
 18 (12.0)
Non-ST-segment elevation myocardial infarction
 29 (19.3)
Unstable angina pectoris
 26 (17.3)
Non-acute coronary syndrome
 77 (51.3)
Stable angina pectoris
 59 (39.3)
Silent ischaemia
  8 (5.3)
Other
 10 (6.7)
Plus—minus values are means ± standard deviation
ACS acute coronary syndrome, PCI percutaneous coronary intervention, CABG coronary artery bypass grafting, CVA cerebrovascular accident, TIA transient ischaemic attack
aItalic numbers represent the known total from which the variable was calculated
bRenal insufficiency was defined as estimated glomerular filtration rate of less than 30 ml/min per 1.73m2 of body surface area (GFR <30 ml/min/1.73m2)
cChronic pulmonary obstructive disease was defined as ≥Gold class II
The angiographic characteristics are described in Tab. 2. The 188 target lesions were treated with a total of 214 implanted devices (ABSORB BVS 60.7%, ABSORB GT1 34.1% and metallic DES 5.1% in addition to implantation of ≥1 Absorb BVS). Device implantation was successful in 100% of the patients and the procedural success was 99.5% with a single patient developing a distal coronary dissection after implantation of an ABSORB BVS with good clinical evolution. Predilatation was performed in 93.3% and postdilatation in 75.5% of the procedures. In no case was postdilatation balloon size more than 0.5 mm larger than the scaffold size. Preimplantation, intracoronary imaging with OCT and IVUS was conducted in 7.4% and 1.0% respectively. Postimplantation, OCT was effectuated in 6.4% and IVUS in 1.6%, mostly for the purpose of apposition control. Postprocedural TIMI grade 3 was observed after all procedures.
Table 2
Angiographic characteristics of the patients at baseline
Baseline angiographic characteristics
 
Patient-level analysis
 
Number of patients
150
Number of treated target lesions—mean ± SD
  1.3 ± 0.5
Treated target lesions ≥2—n (%)
 30 (20.0)
Number of treated target vessels—mean ± SD
  1.1 ± 0.3
Treated target vessels ≥2—n (%)
 12 (8.0)
Devices implanted in proximal coronary segment—n (%)b
 57 (38.0)
Lesion-level analysis
 
Number of lesions
188
Coronary artery lesion distribution—n (%)
 
Right coronary artery
 57 (30.3)
Left anterior descending artery
 89 (47.3)
Circumflex artery
 40 (21.3)
Arterial or venous graft
  2 (1.1)
Coronary artery lesion characteristics
 
Visual estimated diameter stenosis—mean ± SD; nc
 85.5 ± 11.9;181a
Bifurcation—n (%)
 27 (14.4)
Device-level analysis
 
Number of devices
214
Device distribution—n (%)
 
ABSORB BVS
130 (60.7)
ABSORB GT1
 73 (34.1)
Metallic DES
 11 (5.1)
Number of devices at lesion—n (%)
 
1
168 (89.4)
2
 16 (8.5)
3
  2 (1.1)
4
  2 (1.1)
Number of devices per lesion—mean ± SD
  1.1 ± 0.5
Device diameter—mean ± SD
  3.0 ± 0.4
Inflation pressure—mean ± SD; nd
 14.3 ± 2.6;211a
Total treated length—mean ± SD
 29.7 ± 19.0
Procedure-level analysis
 
Results—n (%)
 
Visual diameter stenosis postprocedure <30%
185a (100)
Postprocedural TIMI grade 3
186a (100)
Angiographic success
188 (100)
Device implantation success
188 (100)
Procedural success
187 (99.5)
Peri-implantation procedures
 
FFR measurement—n (%)
 26 (13.8)
Preimplantation OCT or IVUS—n (%)
 14 (7.4)
Predilatation—n (%)
177 (94.1)
Predilatation balloon size—mean ± SD; n
  2.8 ± 0.8;176a
Predilatation pressure—mean ± SD; nd
 14.8 ± 4.0;174a
Postdilatation—n (%)
142 (75.5)
Postdilatation balloon size—mean ± SD
  3.2 ± 0.5
Postdilatation pressure—mean ± SDd
 17.3 ± 4.3
Postdilatation balloon size >0.5 mm larger than scaffold size—n (%)
  0
Postimplantation OCT or IVUS—n (%)
 15 (8.0)
Plus—minus values are means ± standard deviation
Length of lesions, devices and balloons were measured in millimetres (mm), as was the diameter of the devices
DES drug-eluting stents, PCI percutaneous coronary intervention, TIMI Thrombolysis in Myocardial Infarction with grade 3 referenced as completely restored flow, FFR fractional flow reserve, OCT optical coherence tomography, IVUS intravascular ultrasound
aItalic numbers represent the known total from which the variable was calculated
bProximal devices were defined as implantation at lesion segments 1, 5, 6, 11
cVisual estimated diameter stenosis was defined as a percentage
dDilatation and inflation pressures were measured in atmospheres (atm)

Clinical outcomes

All patients received a complete 1‑year follow-up. Three patients (2.0%) were lost to follow-up during this period. At 1‑year follow-up, 72.7% of the patients were still receiving DAPT. The clinical outcomes at 1 year are presented in Tab. 3 and corresponding Figs. 1 and 2. MACE occurred in 14 patients (9.5%), with all-cause death occurring in 4 (2.7%), any MI in 6 (4.1%) and ischaemia-driven TVR in 7 (4.8%) respectively. TLF was observed in 11 patients (7.5%), with CD occurring in 4 (2.7%), target vessel MI in 4 (2.7%) and ischaemia-driven TLR in 4 (2.8%) respectively. Although 2 deaths resulted from progression of malignancy, judging them as CD could not be avoided. Of the recorded target vessel MIs, 2 (1.4%) were periprocedural. Definite and probable ScT was observed in 2 (1.4%) patients. One 80-year-old male patient had a definite ScT after treatment of a single vessel and a single lesion with an ABSORB GT1 in the proximal circumflex artery for an acute coronary syndrome. He presented with an acute MI 8 days after the procedure and underwent successful revascularisation. At the index procedure predilatation was carried out, but no postdilatation or intracoronary imaging was performed. The second patient, a 62-year-old male, with treatment of a single vessel and a single lesion with an ABSORB GT1 in the proximal left anterior descending artery for a non-acute coronary syndrome indication, died of an unknown cause the day after discharge and was regarded as a probable ScT. At the index procedure, pre- and postdilatation were conducted, but intracoronary imaging was not performed. Finally, in a multivariate Cox regression analysis adjusting for age, gender and PCI indication, insulin-treated DM was the only variable that showed a trend toward predicting MACE (HR 2.76; 95% CI: 0.94–8.07; p = 0.06).
Table 3
Safety and efficacy outcomes at 1-year follow-up
Endpoints and clinical events—% (n)
Patients (n = 147)
Primary endpoint: MACEa
9.5 (14)
 All-cause death
2.7 (4)
 Any myocardial infarction
4.1 (6)
 Ischaemia-driven target vessel revascularisation
4.8 (7)
Target lesion failureb
7.5 (11)
 Cardiac death
2.7 (4)
 Target vessel myocardial infarction
2.7 (4)
 Periprocedural myocardial infarction
1.4 (2)
 Ischaemia-driven target lesion revascularisation
2.8 (4)
Definite or probable scaffold thrombosis
1.4 (2)
 Early: 0–30 days
1.4 (2)
  Acute: ≤24 h
0
  Subacute: >24 h–30 days
1.4 (2)
 Late: 31 days: ≤1 year
0
 Very late: >1 year
0
 Definite
0.7 (1)
 Probable
0.7 (1)
The clinical outcomes represented as endpoints and clinical events at 1‑year follow-up. Three patients were lost to follow-up. Endpoints and clinical events are presented by Kaplan-Meier estimates. Angiographic evaluations of baseline as well as repeat angiograms in patients with events were analysed by means of quantitative coronary angiography, Thrombolysis in Myocardial Infarction as well as presence of thrombus analysis by an independent core laboratory (Diagram BV, Zwolle, The Netherlands)
aMajor adverse cardiac events (MACE) were defined as a composite of all-cause death, any myocardial infarction and ischaemia-driven target vessel revascularisation
bTarget lesion failure was defined as a composite of cardiac death, target vessel myocardial infarction and ischaemia-driven target lesion revascularisation

Discussion

The Benelux ABSORB DM Study, the first dedicated prospective study with utilisation of Absorb BVS for percutaneous treatment of CAD in DM patients for any indication in non-complex anatomy shows acceptable safety and efficacy outcomes in this clinically high-risk patient population.
The efficacy outcomes observed in our study are similar to the 1‑year results of the ABSORB III trial with very comparable clinical outcomes for ischaemia-driven TVR and ischaemia-driven TLR [17]. Similarly, the TLF data were almost identical. Also, the reported ScT incidence was similar in both trials. Although ABSORB III included only 31.5% DM patients and no patients with ongoing MI, both studies share a relatively low lesion risk and therefore, as previously shown, simple lesions behave similarly in DM and non-DM patients after PCI [18].
Compared to 1‑year DM sub-analysis of the ABSORB trials, we observed a similar incidence of TLF, ischaemia-driven TVR and ischaemia-driven TLR. However, the incidences of the safety outcomes like ScT and target vessel MI were somewhat lower in our study [19]. While clinical and angiographic baseline characteristics were similar, the observed safety differences may be explained by the longer experience in the handling of Absorb BVS of the participating centres in our study. Indeed, postdilatation balloon size was never more than 0.5 mm larger than scaffold size, showing strict compliance with the procedure protocol.
Comparing the performance of the Absorb BVS to EES in DM patients of the SPIRIT V diabetic study, Absorb BVS has a lower incidence of MACE, TLF, ischaemia-driven TVR and TLR with comparable incidences of target vessel MI at 1‑year follow-up [20]. The patient groups had corresponding angiographic characteristics but there are non-negligible racial differences between both studies. Furthermore, our study showed similar safety and efficacy outcomes with EES used in DM patients in a pooled database of SPIRIT and COMPARE trials, as was also found in a propensity score matched comparison of the ABSORB EXTEND and the SPIRIT trials [7, 21].
The rationale for bioresorbable scaffolds derives from previous evidence that consistently reported that DM patients suffer worse outcomes with PCI as compared to coronary artery bypass grafting (CABG) [22, 23]. However, PCI is still performed in DM patients, particularly in those with single vessel disease as well as multivessel disease with a low SYNTAX score, as is consistent with the results from the SYNTAX trial [24]. Considering the accelerated nature of coronary atherosclerosis disease, CAD presents at a younger age in DM patients. Therefore, maintaining these patients free of ischaemia over decades becomes challenging. While CABG is indeed the treatment of choice for advanced multivessel disease in this patient population, it is known that venous grafts have a limited patency, giving PCI an important role in delaying the time at which these patients ultimately undergo CABG [25]. From this perspective, efforts to improve PCI outcomes in early-stage CAD in DM patients become paramount.
Furthermore, a large analysis of a pooled database of 18 clinical trials has shown that in the DES era, clinical outcomes after PCI in DM patients are highly dependent on lesion complexity at baseline with simple lesions being associated with similar efficacy outcomes to those in non-DM patients [18]. These data suggest that PCI may have favourable outcomes in a well-selected group of patients with DM, provided the extent of disease is less complex. However, progressive restenosis remains problematic with metallic DES, as the number of possible re-interventions is limited. Conversely, resorbable scaffolds are becoming attractive in this particular setting considering the theoretically larger number of possible re-interventions and thus extending the time span during which these patients could still be treated percutaneously.
On the basis of the above, the 1‑year results of our study, which showed that Absorb BVS has promising safety and efficacy outcomes post-PCI in DM patients with non-extensive CAD, stimulate further research in this direction. If maintained up to 3 years, the expected resorption period of this device, this treatment may herald brighter perspectives for PCI in DM patients. Whether this would be the case remains questionable, as the long-term follow-up safety outcomes from the AIDA and ABSORB III trials were unfavourable for Absorb BVS in comparison with EES [26, 27]. On the other hand, these trials also brought to light the importance of the scaffold implantation technique and the advantageous role of longer DAPT regimens [28, 29]. From this perspective, improved implantation techniques and newer, thinner strut resorbable devices combined with 3‑year DAPT regimens may result in improved safety outcomes. This trial was performed in The Netherlands, Belgium and Luxembourg. Following the recommendations of the Dutch Society of Cardiology, we also provided this recommendation to all participating centres in The Netherlands. All patients were recommended to continue DAPT if they were in the range of 3 years post index procedure and were not at high risk for bleeding events.

Limitations

This study has different limitations. First, it has the general intrinsic limitations of a prospective single-arm study lacking a comparative arm. Furthermore, the study population was smaller than foreseen due to the effectuated stop in clinical utilisation of the Absorb BVS. Despite our efforts, the multivariate analysis performed was limited by the number of events observed. There was no uniform procedure protocol for scaffold implantation; however, the different participating centres had years-long experience with Absorb BVS implantation. In addition, the outcomes were not adjusted for the small number of metallic DES that were implanted in addition to the Absorb BVS, in cases where Absorb BVS could not be implanted due to the unavailability of the required device size. Finally, a longer follow-up is mandatory to further evaluate the performance of the Absorb BVS in this specific group of patients.

Conclusion

The results of this multinational dedicated prospective study show that PCI with Absorb BVS for treatment of non-extensive CAD in DM patients, performed by experienced operators, is associated with acceptable safety and efficacy outcomes at 1 year when historically comparable to modern DES. If maintained for longer follow-up periods, covering the resorption phase of the device, our study results would further reinforce the potentially advantageous impact of bioresorbable scaffolds for CAD treatment in fast progressing atherosclerosis populations like DM patients, where PCI even with modern DES remains challenging. From this perspective, it could pave the way for further research with new-generation and more performant bioresorbable scaffolds.

Acknowledgements

The ABSORB DM Study team is grateful to Diagram BV Zwolle and all additional participating centres including their interventional cardiologists and study facilitators. These centres are Meander MC, Amersfoort, The Netherlands; CHR de la Citadelle, Liège, Belgium; CHU de Charleroi, Charleroi, Belgium; Clinique Saint-Luc Bouge, Namur, Belgium; Hagaziekenhuis, The Hague, The Netherlands; AZ Delta, Roeselare, Belgium; OLV Ziekenhuis, Aalst, Belgium; UZ Gent, Gent, Belgium; Amphia Ziekenhuis, Breda, The Netherlands; Haaglanden MC, The Hague, The Netherlands; OLVG, Amsterdam, The Netherlands, Ter Gooi Ziekenhuis, Hilversum, The Netherlands.

Conflict of interest

T.M. Hommels, R.S. Hermanides, S. Rasoul, B. Berta, A.J.J. IJsselmuiden, G.A.J. Jessurun, E. Benit, B. Pereira, G. De Luca and E. Kedhi declare that they have no competing interests.

Ethical standards

The prospective study was performed in accordance with the Declaration of Helsinki (DoH) and was registered in the Netherlands Trial Register (NTR) acknowledged by the World Health Organisation (WHO) and International Committee of Medical Journal Editors (ICMJE). Approval for ethical committees was sought in all participating centres and countries. This clinical investigation did not test any new device. The ABSORB BVS and the ABSORB GT1 are CE marked products and were used widely in daily clinical practice all over the world from 2009 to 2017. This clinical register does not involve a medicinal product and is therefore not subject to The European Clinical Trials Directive (2001/20/EC).
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail

Onze productaanbevelingen

Netherlands Heart Journal

Het Netherlands Heart Journal wordt uitgegeven in samenwerking met de Nederlandse Vereniging voor Cardiologie en de Nederlandse Hartstichting. Het tijdschrift is Engelstalig en wordt gratis beschikbaa ...

Literatuur
1.
go back to reference Shaw JE, Sicree RA, Zimmet PZ. Global estimates of the prevalence of diabetes for 2010 and 2030. Diabetes Res Clin Pract. 2010;87:4–14.CrossRef Shaw JE, Sicree RA, Zimmet PZ. Global estimates of the prevalence of diabetes for 2010 and 2030. Diabetes Res Clin Pract. 2010;87:4–14.CrossRef
2.
go back to reference Guariguata L, Whiting DR, Hambleton I, Beagley J, Linnenkamp U, Shaw JE. Global estimates of diabetes prevalence for 2013 and projections for 2035. Diabetes Res Clin Pract. 2014;103:137–49.CrossRef Guariguata L, Whiting DR, Hambleton I, Beagley J, Linnenkamp U, Shaw JE. Global estimates of diabetes prevalence for 2013 and projections for 2035. Diabetes Res Clin Pract. 2014;103:137–49.CrossRef
3.
go back to reference Gu K, Cowie C, Harris M. Mortality in adults with and without diabetes in a national cohort study of the US population, 1971–1993. Diabetes Care. IEEE Trans Med Imaging. 1998;21:1138:45. Gu K, Cowie C, Harris M. Mortality in adults with and without diabetes in a national cohort study of the US population, 1971–1993. Diabetes Care. IEEE Trans Med Imaging. 1998;21:1138:45.
4.
go back to reference Kannel WB, McGlee DL. Diabetes and glucose tolerance as risk factors for cardiovascular disease: the Framingham study. Diabetes Care. 1979;2:1201–1126.CrossRef Kannel WB, McGlee DL. Diabetes and glucose tolerance as risk factors for cardiovascular disease: the Framingham study. Diabetes Care. 1979;2:1201–1126.CrossRef
5.
go back to reference Lee W, Cape D, Cheung A, Zinman B. Impact of diabetes mellitus on coronary artery disease in woman and men: a meta-analysis of prospective studies. Diabetes Care. 2000;23:962–8.CrossRef Lee W, Cape D, Cheung A, Zinman B. Impact of diabetes mellitus on coronary artery disease in woman and men: a meta-analysis of prospective studies. Diabetes Care. 2000;23:962–8.CrossRef
6.
go back to reference Hammoud T, Tanguay JF, Bourassa MG. Management of coronary artery disease: therapeutic options in patients with diabetes. J Am Coll Cardiol. 2000;36:355–65.CrossRef Hammoud T, Tanguay JF, Bourassa MG. Management of coronary artery disease: therapeutic options in patients with diabetes. J Am Coll Cardiol. 2000;36:355–65.CrossRef
7.
go back to reference Stone GW, Kedhi E, Kereiakes DJ, et al. Differential clinical responses to everolimus-eluting and paclitaxel-eluting coronary stents in patients with and without diabetes mellitus. Circulation. 2011;124:893–900.CrossRef Stone GW, Kedhi E, Kereiakes DJ, et al. Differential clinical responses to everolimus-eluting and paclitaxel-eluting coronary stents in patients with and without diabetes mellitus. Circulation. 2011;124:893–900.CrossRef
8.
go back to reference Jensen LO, Maeng M, Thayssen P, et al. Influence of diabetes mellitus on clinical outcomes following primary percutaneous coronary intervention in patients with ST-segment elevation myocardial infarction. Am J Cardiol. 2012;109:629–35.CrossRef Jensen LO, Maeng M, Thayssen P, et al. Influence of diabetes mellitus on clinical outcomes following primary percutaneous coronary intervention in patients with ST-segment elevation myocardial infarction. Am J Cardiol. 2012;109:629–35.CrossRef
9.
go back to reference Diletti R, Onuma Y, Farooq V, et al. 6‑month clinical outcomes following implantation of the bioresorbable everolimus-eluting vascular scaffold in vessels smaller or larger than 2.5 mm. J Am Coll Cardiol. 2011;58:258–64.CrossRef Diletti R, Onuma Y, Farooq V, et al. 6‑month clinical outcomes following implantation of the bioresorbable everolimus-eluting vascular scaffold in vessels smaller or larger than 2.5 mm. J Am Coll Cardiol. 2011;58:258–64.CrossRef
10.
go back to reference Dudek D, Onuma Y, Ormiston JA, Thuesen L, Miquel-Herbert K, Serruys PW. Four-year clinical follow-up of the ABSORB everolimus-eluting bioresorbable vascular scaffold in patients with de novo coronary artery disease: the ABSORB trial. EuroIntervention. 2012;7:1060–1.CrossRef Dudek D, Onuma Y, Ormiston JA, Thuesen L, Miquel-Herbert K, Serruys PW. Four-year clinical follow-up of the ABSORB everolimus-eluting bioresorbable vascular scaffold in patients with de novo coronary artery disease: the ABSORB trial. EuroIntervention. 2012;7:1060–1.CrossRef
11.
go back to reference Onuma Y, Serruys PW, Ormiston JA, et al. Three-year results of clinical follow-up after a bioresorbable everolimus-eluting scaffold in patients with de novo coronary artery disease: the ABSORB trial. EuroIntervention. 2010;6:447–53.CrossRef Onuma Y, Serruys PW, Ormiston JA, et al. Three-year results of clinical follow-up after a bioresorbable everolimus-eluting scaffold in patients with de novo coronary artery disease: the ABSORB trial. EuroIntervention. 2010;6:447–53.CrossRef
12.
go back to reference Serruys PW, Onuma Y, Dudek D, et al. Evaluation of the second generation of a bioresorbable everolimus-eluting vascular scaffold for the treatment of de novo coronary artery stenosis: 12-month clinical and imaging outcomes. J Am Coll Cardiol. 2011;58:1578–88.CrossRef Serruys PW, Onuma Y, Dudek D, et al. Evaluation of the second generation of a bioresorbable everolimus-eluting vascular scaffold for the treatment of de novo coronary artery stenosis: 12-month clinical and imaging outcomes. J Am Coll Cardiol. 2011;58:1578–88.CrossRef
13.
go back to reference Serruys PW, Ormiston JA, Onuma Y, et al. A bioabsorbable everolimus-eluting coronary stent system (ABSORB): 2‑year outcomes and results from multiple imaging methods. Lancet. 2009;373:897–910.CrossRef Serruys PW, Ormiston JA, Onuma Y, et al. A bioabsorbable everolimus-eluting coronary stent system (ABSORB): 2‑year outcomes and results from multiple imaging methods. Lancet. 2009;373:897–910.CrossRef
14.
go back to reference Windecker S. The ABSORB cohort B trial: insights from longitudinal imaging follow-up from six months to three years. J Am Coll Cardiol. 2013;61:E1643–5.CrossRef Windecker S. The ABSORB cohort B trial: insights from longitudinal imaging follow-up from six months to three years. J Am Coll Cardiol. 2013;61:E1643–5.CrossRef
15.
go back to reference Whitbourn RJ. TCT-31 ABSORB EXTEND: an interim report on the 24-month clinical outcomes from the first 250 patients enrolled. J Am Coll Cardiol. 2013;62:B11.CrossRef Whitbourn RJ. TCT-31 ABSORB EXTEND: an interim report on the 24-month clinical outcomes from the first 250 patients enrolled. J Am Coll Cardiol. 2013;62:B11.CrossRef
16.
go back to reference Tenekecioglu E, Farooq V, Bourantes CV, et al. Bioresorbable scaffolds: a new paradigm in percutaneous coronary intervention. Bmc Cardiovasc Disord. 2016;16:38.CrossRef Tenekecioglu E, Farooq V, Bourantes CV, et al. Bioresorbable scaffolds: a new paradigm in percutaneous coronary intervention. Bmc Cardiovasc Disord. 2016;16:38.CrossRef
17.
go back to reference Ellis SG, Kereiakes DJ, Metzger DC, et al. Everolimus-eluting bioresorbable scaffolds for coronary artery disease. N Engl J Med. 2015;373:1905–15.CrossRef Ellis SG, Kereiakes DJ, Metzger DC, et al. Everolimus-eluting bioresorbable scaffolds for coronary artery disease. N Engl J Med. 2015;373:1905–15.CrossRef
18.
go back to reference Kedhi E, Généreux P, Palmerini T, et al. Impact of coronary lesion complexity on drug-eluting stent outcomes in patients with and without diabetes mellitus: analysis from 18 pooled randomized trials. J Am Coll Cardiol. 2014;63:2111–8.CrossRef Kedhi E, Généreux P, Palmerini T, et al. Impact of coronary lesion complexity on drug-eluting stent outcomes in patients with and without diabetes mellitus: analysis from 18 pooled randomized trials. J Am Coll Cardiol. 2014;63:2111–8.CrossRef
19.
go back to reference Kereiakes DJ, Ellis SG, Kimura T, et al. Efficacy and safety of the Absorb everolimus-eluting bioresorbable scaffold for treatment of patients with diabetes mellitus: results of the Absorb diabetic substudy. Jacc Cardiovasc Interv. 2017;10:42–9.CrossRef Kereiakes DJ, Ellis SG, Kimura T, et al. Efficacy and safety of the Absorb everolimus-eluting bioresorbable scaffold for treatment of patients with diabetes mellitus: results of the Absorb diabetic substudy. Jacc Cardiovasc Interv. 2017;10:42–9.CrossRef
20.
go back to reference Grube E, Chevalier B, Guagliumi G, et al. The SPIRIT V diabetic study: a randomized clinical evaluation of the XIENCE V everolimus-eluting stent vs the TAXUS Liberté paclitaxel-eluting stent in diabetic patients with de novo coronary artery lesions. Am Heart J. 2012;163:867–75.CrossRef Grube E, Chevalier B, Guagliumi G, et al. The SPIRIT V diabetic study: a randomized clinical evaluation of the XIENCE V everolimus-eluting stent vs the TAXUS Liberté paclitaxel-eluting stent in diabetic patients with de novo coronary artery lesions. Am Heart J. 2012;163:867–75.CrossRef
21.
go back to reference Campos CM, Caixeta A, Franken M, et al. Risk and timing of clinical events according to diabetic status of patients treated with everolimus-eluting bioresorbable vascular scaffolds versus everolimus-eluting stent: 2‑year results from a propensity score matched comparison of ABSORB EXTEND and SPIRIT trials. Catheter Cardiovasc Interv. 2018;91:387–95.CrossRef Campos CM, Caixeta A, Franken M, et al. Risk and timing of clinical events according to diabetic status of patients treated with everolimus-eluting bioresorbable vascular scaffolds versus everolimus-eluting stent: 2‑year results from a propensity score matched comparison of ABSORB EXTEND and SPIRIT trials. Catheter Cardiovasc Interv. 2018;91:387–95.CrossRef
22.
go back to reference Investigators B. The final 10-year follow-up results from the BARI randomized trial. J Am Coll Cardiol. 2007;49:1600:6. Investigators B. The final 10-year follow-up results from the BARI randomized trial. J Am Coll Cardiol. 2007;49:1600:6.
23.
go back to reference Farhouh ME, Domanski M, Sleeper LA, et al. Strategies for multivessel revascularization in patients with diabetes. N Engl J Med. 2012;367:2375–84.CrossRef Farhouh ME, Domanski M, Sleeper LA, et al. Strategies for multivessel revascularization in patients with diabetes. N Engl J Med. 2012;367:2375–84.CrossRef
24.
go back to reference Mohr FW, Morice MC, Kappetein AP, et al. Coronary artery bypass graft surgery versus percutaneous coronary intervention in patients with three-vessel disease and left main coronary disease: 5‑year follow-up of the randomised, clinical SYNTAX trial. Lancet. 2013;381:629–38.CrossRef Mohr FW, Morice MC, Kappetein AP, et al. Coronary artery bypass graft surgery versus percutaneous coronary intervention in patients with three-vessel disease and left main coronary disease: 5‑year follow-up of the randomised, clinical SYNTAX trial. Lancet. 2013;381:629–38.CrossRef
25.
go back to reference Goldman S, Zadina K, Moritz T, et al. Long-term patency of saphenous vein and left internal mammary artery grafts after coronary artery bypass surgery: results from a Department of Veterans Affairs Cooperative Study. J Am Coll Cardiol. 2004;44:2149–56.CrossRef Goldman S, Zadina K, Moritz T, et al. Long-term patency of saphenous vein and left internal mammary artery grafts after coronary artery bypass surgery: results from a Department of Veterans Affairs Cooperative Study. J Am Coll Cardiol. 2004;44:2149–56.CrossRef
26.
go back to reference Kereiakes DJ, Ellis SG, Metzger C. et al. 3‑year clinical outcomes with everolimus-eluting bioresorbable coronary scaffolds: the ABSORB III trial. J Am Coll Cardiol. 2017;70:2852–62.CrossRef Kereiakes DJ, Ellis SG, Metzger C. et al. 3‑year clinical outcomes with everolimus-eluting bioresorbable coronary scaffolds: the ABSORB III trial. J Am Coll Cardiol. 2017;70:2852–62.CrossRef
27.
go back to reference Wykrzykowska JJ, Kraak RP, Hofma SH, et al. Bioresorbable scaffolds versus metallic stents in routine PCI. N Engl J Med. 2017;376:2319–28.CrossRef Wykrzykowska JJ, Kraak RP, Hofma SH, et al. Bioresorbable scaffolds versus metallic stents in routine PCI. N Engl J Med. 2017;376:2319–28.CrossRef
28.
go back to reference Stone GW, Abizaid A, Onuma Y, et al. Effect of technique on outcomes following bioresorbable vascular scaffold implantation: analysis from the ABSORB trials. J Am Coll Cardiol. 2017;70:2863–74.CrossRef Stone GW, Abizaid A, Onuma Y, et al. Effect of technique on outcomes following bioresorbable vascular scaffold implantation: analysis from the ABSORB trials. J Am Coll Cardiol. 2017;70:2863–74.CrossRef
29.
go back to reference Collet C, Asano T, Miyazaki Y, et al. Late thrombotic events after bioresorbable scaffold implantation: a systematic review and meta-analysis of randomized clinical trials. Eur Heart J. 2017;38:2559–66.CrossRef Collet C, Asano T, Miyazaki Y, et al. Late thrombotic events after bioresorbable scaffold implantation: a systematic review and meta-analysis of randomized clinical trials. Eur Heart J. 2017;38:2559–66.CrossRef
Metagegevens
Titel
The 1‑year safety and efficacy outcomes of Absorb bioresorbable vascular scaffolds for coronary artery disease treatment in diabetes mellitus patients: the ABSORB DM Benelux study
Auteurs
T. M. Hommels
R. S. Hermanides
S. Rasoul
B. Berta
A. J. J. IJsselmuiden
G. A. J. Jessurun
E. Benit
B. Pereira
G. De Luca
E. Kedhi
Publicatiedatum
13-06-2019
Uitgeverij
Bohn Stafleu van Loghum
Gepubliceerd in
Netherlands Heart Journal / Uitgave 11/2019
Print ISSN: 1568-5888
Elektronisch ISSN: 1876-6250
DOI
https://doi.org/10.1007/s12471-019-1293-7

Andere artikelen Uitgave 11/2019

Netherlands Heart Journal 11/2019 Naar de uitgave