Abstract
Aims
Adenosine administration after pulmonary vein (PV) isolation using radiofrequency and cryoablation can cause acute recovery of conduction to the PVs and predicts atrial fibrillation (AF) recurrence. This study evaluated whether adenosine testing after second-generation balloon devices (cryothermal and laser) could reveal dormant PV reconduction and recurrence rate of AF.
Methods
Out of total 60 patients, 20 patients underwent PV isolation using laser balloon and for the remaining patients, the cryoballoon was used. Following PVI, waiting period of 30 min was obtained. Thereafter, a bolus 15–21 mg of adenosine was injected followed by rapid saline flush. The subsequent response was assessed for each vein using an in situ lasso catheter. Further ablation (if needed) using laser balloon and cryocatheter respectively was done, until no reconduction occurred after repeat adenosine.
Results
Acute PV isolation was achieved in all 80 PVs of 20 patients (100 %) using laser device and in 151 PVs (96.2 %) of 38 patients (95 %) using cryoballoon. However, in seven patients (35 %), 11 PVs (13.7 %) (4 LSPV, 2 LIPV, 4 RSPV, and 1 RIPV) showed dormant PV potentials after adenosine administration in laser group. Cryoballoon group showed dormant reconduction in four patients (10 %), four PVs (5 %) [one LSPV, one LIPV, and two RIPV]. The follow-up of 337 ± 92.4 days for cryoballoon and 267 ± 76.9 days for laser balloon group demonstrated similar success rates (85 %).
Conclusion
Adenosine testing after PV isolation using second-generation balloon based energy devices (laser and cryothermal) reveals dormant conduction in initially isolated PVs with similar long-term success rate.
Similar content being viewed by others
References
Kirchhof, P., Auricchio, A., Bax, J., Crijns, H., Camm, J., Diener, H. C., et al. (2007). Outcome parameters for trials in atrial fibrillation: executive summary. European Heart Journal, 28(22), 2803–2817.
Haissaguerre, M., Jais, P., Shah, D. C., Takahashi, A., Hocini, M., Quiniou, G., et al. (1998). Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. New England Journal of Medicine, 339(10), 659–666.
Cappato, R., Negroni, S., Pecora, D., Bentivegna, S., Lupo, P. P., Carolei, A., et al. (2003). Prospective assessment of late conduction recurrence across radiofrequency lesions producing electrical disconnection at the pulmonary vein ostium in patients with atrial fibrillation. Circulation, 108(13), 1599–1604.
Ouyang, F., Antz, M., Ernst, S., Hachiya, H., Mavrakis, H., Deger, F. T., et al. (2005). Recovered pulmonary vein conduction as a dominant factor for recurrent atrial tachyarrhythmias after complete circular isolation of the pulmonary veins: lessons from double Lasso technique. Circulation, 111(2), 127–135.
Furnkranz, A., Bordignon, S., Schmidt, B., Gunawardene, M., Schulte-Hahn, B., Urban, V., et al. (2013). Improved procedural efficacy of pulmonary vein isolation using the novel second-generation cryoballoon. Journal of Cardiovascular Electrophysiology, 24(5), 492–497.
Reddy, V. Y., Neuzil, P., Themistoclakis, S., Danik, S. B., Bonso, A., Rossillo, A., et al. (2009). Visually-guided balloon catheter ablation of atrial fibrillation: experimental feasibility and first-in-human multicenter clinical outcome. Circulation, 120(1), 12–20.
Metzner, A., Schmidt, B., Fuernkranz, A., Wissner, E., Tilz, R. R., Chun, K. R., et al. (2011). One-year clinical outcome after pulmonary vein isolation using the novel endoscopic ablation system in patients with paroxysmal atrial fibrillation. Heart Rhythm, 8(7), 988–993.
Dukkipati, S. R., Kuck, K. H., Neuzil, P., Woollett, I., Kautzner, J., McElderry, H. T., et al. (2013). Pulmonary vein isolation using a visually guided laser balloon catheter: the first 200-patient multicenter clinical experience. Circulation Arrhythmia and Electrophysiology, 6(3), 467–472.
Dukkipati, S. R., Neuzil, P., Kautzner, J., Petru, J., Wichterle, D., Skoda, J., et al. (2012). The durability of pulmonary vein isolation using the visually guided laser balloon catheter: multicenter results of pulmonary vein remapping studies. Heart Rhythm, 9(6), 919–925.
Metzner, A., Wissner, E., Schmidt, B., Chun, J., Hindricks, G., Piorkowski, C., et al. (2013). Acute and long-term clinical outcome after endoscopic pulmonary vein isolation: results from the first prospective, multicenter study. Journal of Cardiovascular Electrophysiology, 24(1), 7–13.
Arentz, T., Macle, L., Kalusche, D., Hocini, M., Jais, P., Shah, D., et al. (2004). “Dormant” pulmonary vein conduction revealed by adenosine after ostial radiofrequency catheter ablation. Journal of Cardiovascular Electrophysiology, 15(9), 1041–1047.
Tritto, M., De Ponti, R., Salerno-Uriarte, J. A., Spadacini, G., Marazzi, R., Moretti, P., et al. (2004). Adenosine restores atrio-venous conduction after apparently successful ostial isolation of the pulmonary veins. European Heart Journal, 25(23), 2155–2163.
Hachiya, H., Hirao, K., Takahashi, A., Nagata, Y., Suzuki, K., Maeda, S., et al. (2007). Clinical implications of reconnection between the left atrium and isolated pulmonary veins provoked by adenosine triphosphate after extensive encircling pulmonary vein isolation. Journal of Cardiovascular Electrophysiology, 18(4), 392–398.
Datino, T., Macle, L., Qi, X. Y., Maguy, A., Comtois, P., Chartier, D., et al. (2010). Mechanisms by which adenosine restores conduction in dormant canine pulmonary veins. Circulation, 121(8), 963–972.
Freilich, A., & Tepper, D. (1992). Adenosine and its cardiovascular effects. American Heart Journal, 123(5), 1324–1328.
Van Belle, Y. L., Janse, P. A., de Groot, N. M., Anne, W., Theuns, D. A., & Jordaens, L. J. (2012). Adenosine testing after cryoballoon pulmonary vein isolation improves long-term clinical outcome. Netherlands Heart Journal, 20(11), 447–455.
Miyazaki, S., Kobori, A., Hocini, M., Shah, A. J., Komatsu, Y., Taniguchi, H., et al. (2013). Clinical utility of adenosine-infusion test at a repeat atrial fibrillation ablation procedure. Heart Rhythm, 10(5), 629–635.
Calkins, H., Kuck, K. H., Cappato, R., Brugada, J., Camm, A. J., Chen, S. A., et al. (2012). 2012 HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design. Journal of Interventional Cardiac Electrophysiology, 33(2), 171–257.
Kuhne, M., Schaer, B., Ammann, P., Suter, Y., Osswald, S., & Sticherling, C. (2010). Cryoballoon ablation for pulmonary vein isolation in patients with paroxysmal atrial fibrillation. Swiss Medical Weekly, 140(15–16), 214–221.
Chun, K. R., Furnkranz, A., Metzner, A., Schmidt, B., Tilz, R., Zerm, T., et al. (2009). Cryoballoon pulmonary vein isolation with real-time recordings from the pulmonary veins. Journal of Cardiovascular Electrophysiology, 20(11), 1203–1210.
Furnkranz, A., Chun, K. R., Nuyens, D., Metzner, A., Koster, I., Schmidt, B., et al. (2010). Characterization of conduction recovery after pulmonary vein isolation using the “single big cryoballoon” technique. Heart Rhythm, 7(2), 184–190.
Schmidt, B., Gunawardene, M., Krieg, D., Bordignon, S., Furnkranz, A., Kulikoglu, M., et al. (2013). A prospective randomized single-center study on the risk of asymptomatic cerebral lesions comparing irrigated radiofrequency current ablation with the cryoballoon and the laser balloon. Journal of Cardiovascular Electrophysiology, 24(8), 869–874.
Verma, A., Kilicaslan, F., Pisano, E., Marrouche, N. F., Fanelli, R., Brachmann, J., et al. (2005). Response of atrial fibrillation to pulmonary vein antrum isolation is directly related to resumption and delay of pulmonary vein conduction. Circulation, 112(5), 627–635.
Reichlin, T., & Michaud, G. F. (2012). Our approach to maximizing the durability of pulmonary vein isolation during a paroxysmal atrial fibrillation ablation procedure. Journal of Cardiovascular Electrophysiology, 23(11), 1272–1276.
Cheung, J. W., Chung, J. H., Ip, J. E., Markowitz, S. M., Liu, C. F., Thomas, G., et al. (2012). Time course of adenosine-induced pulmonary vein reconnection after isolation: implications for mechanism of dormant conduction. Pacing and Clinical Electrophysiology, 35(5), 556–563.
Matsuo, S., Yamane, T., Date, T., Inada, K., Kanzaki, Y., Tokuda, M., et al. (2007). Reduction of AF recurrence after pulmonary vein isolation by eliminating ATP-induced transient venous re-conduction. Journal of Cardiovascular Electrophysiology, 18(7), 704–708.
Bordignon, S., Chun, K. R., Gunawardene, M., Schulte-Hahn, B., Nowak, B., Fuernkranz, A., et al. (2013). Endoscopic ablation systems. Expert Review of Medical Devices, 10(2), 177–183.
Acknowledgments
We would like to thank Suzanne Philippens (RN) for helping us in retrieving clinical follow-up data, Nousjka Vranken (BASc), and Willem Daessen, PhD for statistical analysis.
Funding
None.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Kumar, N., Blaauw, Y., Timmermans, C. et al. Adenosine testing after second-generation balloon devices (cryothermal and laser) mediated pulmonary vein ablation for atrial fibrillation. J Interv Card Electrophysiol 41, 91–97 (2014). https://doi.org/10.1007/s10840-014-9921-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10840-014-9921-z