Clinical outcome of left ventricular multipoint pacing versus conventional biventricular pacing in cardiac resynchronization therapy: a systematic review and meta-analysis

  • Published:
Heart Failure Reviews Aims and scope Submit manuscript

Abstract

Cardiac resynchronization therapy (CRT) is an effective treatment for selected patients with systolic heart failure. Unlike conventional biventricular pacing (BIP), the left ventricular multipoint pacing (MPP) can increase the number of left ventricular pacing sites via a quadripolar lead positioned in the coronary sinus. This synthetic study was conducted to integratively and quantitatively evaluate the clinical outcome of MPP in comparison with BIP. We systematically searched the databases of EMBASE, Ovid medline, and Cochrane Library through May 2018 for studies comparing the clinical outcome of MPP with BIP in the patients who accepted CRT. Hospitalization for reason of heart failure, left ventricular eject fraction (LVEF), CRT response, all-cause morbidity, and cardiovascular death rate was collected for meta-analysis. A total of 11 studies with 29,606 participants were included in this meta-analysis. Compared with BIP group, MPP decreased heart failure hospitalization (OR, 0.41; 95% CI, 0.33 to 0.50; P < 0.00001), improved LVEF (mean difference, 4.97; 95% CI, 3.11 to 6.83; P < 0.00001), increased CRT response (OR, 3.64; 95% CI, 1.68 to 7.87; P = 0.001), and decreased all-cause morbidity (OR, 0.41; 95% CI, 0.26–0.66; P = 0.0002) and cardiovascular death rate (OR, 0.21; 95% CI, 0.11–0.40; P < 0.00001). The published literature demonstrates that MPP was more effective than BIP in the heart failure patients who accepted cardiac resynchronization therapy.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Normand C, Linde C, Singh J, Dickstein K (2018) Indications for cardiac resynchronization therapy: A comparison of the major international guidelines. JACC Heart Fail 6(4):308–316. https://doi.org/10.1016/j.jchf.2018.01.022

    Article  PubMed  Google Scholar 

  2. Brignole M, Auricchio A, Baron-Esquivias G, Bordachar P, Boriani G, Breithardt OA, Cleland J, Deharo JC, Delgado V, Elliott PM, Gorenek B, Israel CW, Leclercq C, Linde C, Mont L, Padeletti L, Sutton R, Vardas PE, Zamorano JL, Achenbach S, Baumgartner H, Bax JJ, Bueno H, Dean V, Deaton C, Erol C, Fagard R, Ferrari R, Hasdai D, Hoes AW, Kirchhof P, Knuuti J, Kolh P, Lancellotti P, Linhart A, Nihoyannopoulos P, Piepoli MF, Ponikowski P, Sirnes PA, Tamargo JL, Tendera M, Torbicki A, Wijns W, Windecker S, Kirchhof P, Blomstrom-Lundqvist C, Badano LP, Aliyev F, Bansch D, Baumgartner H, Bsata W, Buser P, Charron P, Daubert JC, Dobreanu D, Faerestrand S, Hasdai D, Hoes AW, Le Heuzey JY, Mavrakis H, McDonagh T, Merino JL, Nawar MM, Nielsen JC, Pieske B, Poposka L, Ruschitzka F, Tendera M, Van Gelder IC, Wilson CM (2013) 2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European heart rhythm association (EHRA). Eur Heart J 34(29):2281–2329. https://doi.org/10.1093/eurheartj/eht150

    Article  PubMed  Google Scholar 

  3. Carluccio E, Biagioli P, Alunni G, Murrone A, Zingarini G, Coiro S, D’Antonio A, Mengoni A, Cerasa MF, Ambrosio G (2017) Non-cardiac factors for prediction of response to cardiac resynchronization therapy: the value of baseline, and of serial changes, in red cell distribution width. Int J Cardiol 243:347–353. https://doi.org/10.1016/j.ijcard.2017.05.123

    Article  PubMed  Google Scholar 

  4. Kydd AC, Khan FZ, Ring L, Pugh PJ, Virdee MS, Dutka DP (2014) Development of a multiparametric score to predict left ventricular remodelling and prognosis after cardiac resynchronization therapy. Eur J Heart Fail 16(11):1206–1213. https://doi.org/10.1002/ejhf.167

    Article  PubMed  Google Scholar 

  5. Martin CA, Gajendragadkar PR, Pugh PJ (2014) Unusual cause of poor response to cardiac resynchronisation therapy. Heart 100(6):514. https://doi.org/10.1136/heartjnl-2013-305340

    Article  PubMed  Google Scholar 

  6. Thompson N, Derval N (2016) Left ventricular endocardial stimulation in patients with a poor response to cardiac resynchronization therapy: What Is Next? JACC Clin Electrophysiol 2(7):810–811. https://doi.org/10.1016/j.jacep.2016.06.003

    Article  PubMed  Google Scholar 

  7. Wang J, Su Y, Bai J, Wang W, Qin S, Ge J (2014) Elevated pulmonary artery pressure predicts poor outcome after cardiac resynchronization therapy. J Interv Card Electrophysiol 40(2):171–178. https://doi.org/10.1007/s10840-014-9890-2

    Article  CAS  PubMed  Google Scholar 

  8. Werys K, Petryka-Mazurkiewicz J, Blaszczyk L, Misko J, Spiewak M, Malek LA, Mazurkiewicz L, Milosz-Wieczorek B, Marczak M, Kubik A, Dabrowska A, Piatkowska-Janko E, Sawionek B, Wijesurendra R, Piechnik SK, Bogorodzki P (2016) Cine dyscontractility index: a novel marker of mechanical dyssynchrony that predicts response to cardiac resynchronization therapy. J Magn Reson Imaging 44(6):1483–1492. https://doi.org/10.1002/jmri.25295

    Article  PubMed  Google Scholar 

  9. Auricchio A, Heggermont WA (2018) Technology advances to improve response to cardiac resynchronization therapy: what clinicians should know. Rev Esp Cardiol (Engl Ed) 71(6):477–484. https://doi.org/10.1016/j.rec.2018.01.006

    Article  Google Scholar 

  10. Bax JJ, Delgado V (2017) Cardiac resynchronization therapy with multipoint left ventricular lead pacing. JACC Clin Electrophysiol 3(13):1519–1522. https://doi.org/10.1016/j.jacep.2017.10.013

    Article  PubMed  Google Scholar 

  11. Moher D, Liberati A, Tetzlaff J, Altman DG (2009) Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. BMJ 339:b2535. https://doi.org/10.1136/bmj.b2535

    Article  PubMed  PubMed Central  Google Scholar 

  12. Stang A (2010) Critical evaluation of the Newcastle-Ottawa scale for the assessment of the quality of nonrandomized studies in meta-analyses. Eur J Epidemiol 25(9):603–605. https://doi.org/10.1007/s10654-010-9491-z

    Article  PubMed  Google Scholar 

  13. Niazi I, Baker J, Corbisiero R, Love C, Martin D, Sheppard R, Worley SJ, Varma N, Lee K, Tomassoni G (2017) Safety and efficacy of multipoint pacing in cardiac resynchronization therapy. The multipoint pacing trial. JACC Clin Electrophysiol 3(13):1510–1518. https://doi.org/10.1016/j.jacep.2017.06.022

    Article  PubMed  Google Scholar 

  14. Pappone C, Calović Z, Cuko A, McSpadden LC, Ryu K, Jordan CD, Saviano M, Baldi M, Pappone A, Dozza L, Giannelli L, Fragakis N, Vicedomini G, Santinelli V (2015) Multipoint left ventricular pacing provides additional echocardiographic benefit to responders and non-responders to conventional cardiac resynchronization therapy. Eur Heart J Suppl 17:A12–A17. https://doi.org/10.1093/eurheartj/suv005

    Article  Google Scholar 

  15. Sterlinski M, Sokal A, Lenarczyk R, Van Heuverswyn F, Rinaldi CA, Vanderheyden M, Khalameizer V, Francis D, Heynens J, Stegemann B, Cornelussen R (2016) In heart failure patients with left bundle branch block single lead multispot left ventricular pacing does not improve acute hemodynamic response to conventional biventricular pacing. A multicenter prospective, interventional, non-randomized study. PLoS One 11(4):e0154024. https://doi.org/10.1371/journal.pone.0154024

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  16. van Everdingen WM, Zweerink A, Salden OAE, Cramer MJ, Doevendans PA, Engels EB, van Rossum AC, Prinzen FW, Vernooy K, Allaart CP, Meine M (2018) Pressure-volume loop analysis of multipoint pacing with a quadripolar left ventricular lead in cardiac resynchronization therapy. JACC Clinical Electrophysiol 4:881–889. https://doi.org/10.1016/j.jacep.2018.02.005

    Article  Google Scholar 

  17. Zanon F, Marcantoni L, Baracca E, Pastore G, Giau G, Rigatelli G, Lanza D, Picariello C, Aggio S, Giatti S, Zuin M, Roncon L, Pacetta D, Noventa F, Prinzen FW (2018) Hemodynamic comparison of different multisites and multipoint pacing strategies in cardiac resynchronization therapies. J Interv Card Electrophysiol 7:1–9. https://doi.org/10.1007/s10840-018-0362-y

    Article  Google Scholar 

  18. Alonso P, Andrés A, Osca J, Cano O, José M, De Carranza ST, De Ros JO (2014) Improvement in hemodynamics and contractility with multipoint left ventricular pacing in cardiac resynchronization therapy. Rev Esp Cardiol (Engl Ed) 67(10):859–861. https://doi.org/10.1016/j.recesp.2014.06.008

    Article  Google Scholar 

  19. Cabrera Bueno F, Alzueta Rodriguez J, Olague de Ros J, Fernandez-Lozano I, Garcia Guerrero JJ, de la Concha JF, Hernandez Madrid A, Tolosana Viu JM, Osca Asensi J, Barrera Cordero A, Llorente Hernangomez E (2013) Improvement in hemodynamic response using a quadripolar LV lead. Pacing Clin Electrophysiol 36(8):963–969. https://doi.org/10.1111/pace.12172

    Article  PubMed  Google Scholar 

  20. Menardi E, Ballari GP, Goletto C, Rossetti G, Vado A (2015) Characterization of ventricular activation pattern and acute hemodynamics during multipoint left ventricular pacing. Heart Rhythm 12(8):1762–1769. https://doi.org/10.1016/j.hrthm.2015.04.029

    Article  PubMed  Google Scholar 

  21. Osca J, Alonso P, Cano O, Andres A, Miro V, Tello MJS, Olagüe J, Martínez L, Salvador A (2016) The use of multisite left ventricular pacing via quadripolar lead improves acute haemodynamics and mechanical dyssynchrony assessed by radial strain speckle tracking: initial results. Europace 18(4):560–567. https://doi.org/10.1093/europace/euv211

    Article  PubMed  Google Scholar 

  22. Valzania C, Eriksson MJ, Biffi M, Boriani G, Gadler F (2013) Acute changes in electromechanical parameters during different pacing configurations using a quadripolar left ventricular lead. J Interv Card Electrophysiol 38(1):61–69. https://doi.org/10.1007/s10840-013-9812-8

    Article  PubMed  Google Scholar 

  23. Zanon F, Baracca E, Pastore G, Marcantoni L, Fraccaro C, Lanza D, Picariello C, Aggio S, Roncon L, Dell’Avvocata F, Rigatelli G, Pacetta D, Noventa F, Prinzen FW (2015) Multipoint pacing by a left ventricular quadripolar lead improves the acute hemodynamic response to CRT compared with conventional biventricular pacing at any site. Heart Rhythm 12(5):975–981. https://doi.org/10.1016/j.hrthm.2015.01.034

    Article  PubMed  Google Scholar 

  24. Arias MA, Pachon M, Puchol A, Jimenez-Lopez J, Rodriguez-Padial L (2012) Acute and mid-term outcomes of transvenous implant of a new left ventricular quadripolar lead versus bipolar leads for cardiac resynchronization therapy: results from a single-center prospective database. Cardiol J 19(5):470–478

    Article  PubMed  Google Scholar 

  25. Forleo GB, Di Biase L, Panattoni G, Mantica M, Parisi Q, Martino A, Pappalardo A, Sergi D, Tesauro M, Papavasileiou LP, Santini L, Calò L, Tondo C, Natale A, Romeo F (2014) Improved implant and postoperative lead performance in CRT-D patients implanted with a quadripolar left ventricular lead. A 6-month follow-up analysis from a multicenter prospective comparative study. J Interv Card Electrophysiol 42(1):59–66. https://doi.org/10.1007/s10840-014-9956-1

    Article  PubMed  Google Scholar 

  26. Crossley GH, Biffi M, Johnson B, Lin A, Gras D, Hussin A, Cuffio A, Collier JL, El-Chami M, Li S, Holloman K, Exner DV (2015) Performance of a novel left ventricular lead with short bipolar spacing for cardiac resynchronization therapy: primary results of the Attain Performa quadripolar left ventricular lead study. Heart Rhythm 12(4):751–758. https://doi.org/10.1016/j.hrthm.2014.12.019

    Article  PubMed  Google Scholar 

  27. Philippon F, Liu L, Fung JW, Deharo JC, Anselme F, Delnoy PP, Crijns H, Morillo CA, Krahn AD, Gutleben K, Delumeau J, Molin F (2015) Left ventricular three-dimensional quadripolar lead acute clinical study: the LILAC study. Pacing Clin Electrophysiol 38(4):438–447. https://doi.org/10.1111/pace.12584

    Article  PubMed  Google Scholar 

  28. Rinaldi CA, Leclercq C, Kranig W, Kacet S, Betts T, Bordachar P, Gutleben KJ, Shetty A, Donal E, Keel A, Ryu K, Farazi TG, Simon M, Naqvi TZ (2014) Improvement in acute contractility and hemodynamics with multipoint pacing via a left ventricular quadripolar pacing lead. J Interv Card Electrophysiol 40(1):75–80. https://doi.org/10.1007/s10840-014-9891-1

    Article  PubMed  Google Scholar 

  29. Shetty AK, Duckett SG, Bostock J, Roy D, Ginks M, Hamid S, Rosenthal E, Razavi R, Rinaldi CA (2011) Initial single-center experience of a quadripolar pacing lead for cardiac resynchronization therapy. Pacing Clin Electrophysiol 34(4):484–489. https://doi.org/10.1111/j.1540-8159.2010.03003.x

    Article  PubMed  Google Scholar 

  30. Vado A, Menardi E, Rossetti G, Ballari G, Feola M, Bobbio M (2014) Single-center experience of a quadripolar pacing lead for cardiac resynchronization therapy. J Interv Card Electrophysiol 39(2):161–165. https://doi.org/10.1007/s10840-013-9849-8

    Article  PubMed  Google Scholar 

  31. Ziacchi M, Saporito D, Zardini M, Luzi M, Quartieri F, Morgagni G, De Maria E, Bertini M, Carinci V, Boriani G, Biffi M (2016) Left ventricular reverse remodeling elicited by a quadripolar lead: results from the multicenter Per4mer study. Pacing Clin Electrophysiol 39(3):250–260. https://doi.org/10.1111/pace.12792

    Article  PubMed  Google Scholar 

  32. Forleo GB, Di Biase L, Della Rocca DG, Panattoni G, Mantica M, Santamaria M, Pappalardo A, Panigada S, Santini L, Natale A, Romeo F (2012) Impact of previous myocardial infarction on outcomes of CRT patients implanted with a quadripolar left ventricular lead. Results from a multicentric prospective study. Int J Cardiol 160(2):145–146. https://doi.org/10.1016/j.ijcard.2012.05.077

    Article  PubMed  Google Scholar 

  33. Calo L, Martino A, de Ruvo E, Minati M, Fratini S, Rebecchi M, Lanzillo C, Fagagnini A, Borrelli A, De Luca L, Sciarra L (2014) Acute echocardiographic optimization of multiple stimulation configurations of cardiac resynchronization therapy through quadripolar left ventricular pacing: a tailored approach. Am Heart J 167(4):546–554. https://doi.org/10.1016/j.ahj.2013.12.028

    Article  PubMed  Google Scholar 

  34. Osca J, Alonso P, Cano O, Sanchez JM, Tejada D, Andres A, Sancho Tello MJ, Olague J (2015) The use of quadripolar left ventricular leads improves the hemodynamic response to cardiac resynchronization therapy. Pacing Clinical Electrophysiol 38(3):326–333. https://doi.org/10.1111/pace.12545

    Article  Google Scholar 

  35. Umar F, Taylor RJ, Stegemann B, Marshall H, Flannigan S, Lencioni M, De Bono J, Griffith M, Leyva F (2016) Haemodynamic effects of cardiac resynchronization therapy using single-vein, three-pole, multipoint left ventricular pacing in patients with ischaemic cardiomyopathy and a left ventricular free wall scar: the MAESTRO study. Europace 18(8):1227–1234. https://doi.org/10.1093/europace/euv396

    Article  PubMed  Google Scholar 

  36. Pappone C, Calovic Z, Vicedomini G, Cuko A, McSpadden LC, Ryu K, Romano E, Baldi M, Saviano M, Pappone A, Ciaccio C, Giannelli L, Ionescu B, Petretta A, Vitale R, Fundaliotis A, Tavazzi L, Santinelli V (2015) Multipoint left ventricular pacing in a single coronary sinus branch improves mid-term echocardiographic and clinical response to cardiac resynchronization therapy. J Cardiovasc Electrophysiol 26(1):58–63. https://doi.org/10.1111/jce.12513

    Article  PubMed  Google Scholar 

  37. Pappone C, Ćalović Ž, Vicedomini G, Cuko A, McSpadden LC, Ryu K, Romano E, Saviano M, Baldi M, Pappone A, Ciaccio C, Giannelli L, Ionescu B, Petretta A, Vitale R, Fundaliotis A, Tavazzi L, Santinelli V (2014) Multipoint left ventricular pacing improves acute hemodynamic response assessed with pressure-volume loops in cardiac resynchronization therapy patients. Heart Rhythm 11(3):394–401. https://doi.org/10.1016/j.hrthm.2013.11.023

    Article  PubMed  Google Scholar 

  38. Seger M, Hanser F, Dichtl W, Stuehlinger M, Hintringer F, Trieb T, Pfeifer B, Berger T (2014) Non-invasive imaging of cardiac electrophysiology in a cardiac resynchronization therapy defibrillator patient with a quadripolar left ventricular lead. Europace 16(5):743–749. https://doi.org/10.1093/europace/euu045

    Article  PubMed  Google Scholar 

  39. Behar JM, Bostock J, Zhu Li AP, Chin HM, Jubb S, Lent E, Gamble J, Foley PW, Betts TR, Rinaldi CA, Herring N (2015) Cardiac resynchronization therapy delivered via a multipolar left ventricular lead is associated with reduced mortality and elimination of phrenic nerve stimulation: long-term follow-up from a multicenter registry. J Cardiovasc Electrophysiol 26(5):540–546. https://doi.org/10.1111/jce.12625

    Article  PubMed  PubMed Central  Google Scholar 

  40. Forleo GB, Biase L, Bharmi R, Dalal N, Panattoni G, Pollastrelli A, Tesauro M, Santini L, Natale A, Romeo F (2015) Hospitalization rates and associated cost analysis of cardiac resynchronization therapy with an implantable defibrillator and quadripolar vs. bipolar left ventricular leads: a comparative effectiveness study. Europace 17(1):101–107. https://doi.org/10.1093/europace/euu290

    Article  PubMed  Google Scholar 

  41. Pappone C, Ćalović Ž, Vicedomini G, Cuko A, McSpadden LC, Ryu K, Jordan CD, Romano E, Baldi M, Saviano M, Pappone A, Vitale R, Catalano C, Ciaccio C, Giannelli L, Ionescu B, Petretta A, Fragakis N, Fundaliotis A, Tavazzi L, Santinelli V (2015) Improving cardiac resynchronization therapy response with multipoint left ventricular pacing: twelve-month follow-up study. Heart Rhythm 12(6):1250–1258. https://doi.org/10.1016/j.hrthm.2015.02.008

    Article  PubMed  Google Scholar 

  42. Bencardino G, Monaco A, Russo E, Colizzi C, Perna F, Pelargonio G, Narducci ML, Gabrielli FA, Lanza GA, Rebuzzi AG, Crea F (2016) Outcome of patients treated by cardiac resynchronization therapy using a quadripolar left ventricular lead. Circ J 80(3):613–618. https://doi.org/10.1253/circj.CJ-15-0932

    Article  PubMed  Google Scholar 

  43. Turakhia MP, Cao M, Fischer A, Nabutovsky Y, Sloman LS, Dalal N, Gold MR (2016) Reduced mortality associated with quadripolar compared to bipolar left ventricular leads in cardiac resynchronization therapy. JACC Clin Electrophysiol 2(4):426–433. https://doi.org/10.1016/j.jacep.2016.02.007

    Article  PubMed  Google Scholar 

  44. Zanon F, Marcantoni L, Baracca E, Pastore G, Lanza D, Fraccaro C, Picariello C, Conte L, Aggio S, Roncon L, Pacetta D, Badie N, Noventa F, Prinzen FW (2016) Optimization of left ventricular pacing site plus multipoint pacing improves remodeling and clinical response to cardiac resynchronization therapy at 1 year. Heart Rhythm 13(8):1644–1651. https://doi.org/10.1016/j.hrthm.2016.05.015

    Article  PubMed  Google Scholar 

  45. Behar JM, Chin HMS, Fearn S, Ormerod JOM, Gamble J, Foley PWX, Bostock J, Claridge S, Jackson T, Sohal M, Antoniadis AP, Razavi R, Betts TR, Herring N, Rinaldi CA (2017) Cost-effectiveness analysis of quadripolar versus bipolar left ventricular leads for cardiac resynchronization defibrillator therapy in a large, multicenter UK registry. JACC Clin Electrophysiol 3(2):107–116. https://doi.org/10.1016/j.jacep.2016.04.009

    Article  PubMed  PubMed Central  Google Scholar 

  46. Forleo GB, Santini L, Giammaria M, Potenza D, Curnis A, Calabrese V, Ricciardi D, D’Agostino C, Notarstefano P, Ribatti V, Morani G, Mantica M, Di Biase L, Bertaglia E, Calò L, Zanon F (2017) Multipoint pacing via a quadripolar left-ventricular lead: preliminary results from the Italian registry on multipoint left-ventricular pacing in cardiac resynchronization therapy (IRON-MPP). Europace 19(7):1170–1177. https://doi.org/10.1093/europace/euw094

    Article  PubMed  Google Scholar 

  47. Gu M, Jin H, Hua W, Fan XH, Ding LG, Wang J, Niu HX, Cai C, Zhang S (2017) Repetitive optimizing left ventricular pacing configurations with quadripolar leads improves response to cardiac resynchronization therapy. Medicine (Baltimore) 96(37):e8066. https://doi.org/10.1097/MD.0000000000008066

    Article  Google Scholar 

  48. Leyva F, Zegard A, Qiu T, Acquaye E, Ferrante G, Walton J, Marshall H (2017) Cardiac resynchronization therapy using quadripolar versus non-quadripolar left ventricular leads programmed to biventricular pacing with single-site left ventricular pacing: impact on survival and heart failure hospitalization. J Am Heart Assoc 6(10):e007026. https://doi.org/10.1161/JAHA.117.007026

    Article  PubMed  PubMed Central  Google Scholar 

  49. Leshem E, Suleiman M, Laish-Farkash A, Haim M, Geist M, Luria D, Glikson M, Goldenberg I, Michowitz Y (2018) Impact of quadripolar LV leads on heart failure hospitalization rates among patients implanted with CRT-D: data from the Israeli ICD registry. J Interv Card Electrophysiol 51(1):5–12. https://doi.org/10.1007/s10840-017-0305-z

    Article  PubMed  Google Scholar 

  50. Turagam MK, Afzal MR, Iskander S, Reddy M, Di Biase L, Natale A, Lakkireddy D (2016) A meta-analysis of quadripolar versus bipolar left ventricular leads on post-procedural outcomes. J Atr Fibrillation 9(2):7–11. https://doi.org/10.4022/jafib.1472

    Article  Google Scholar 

  51. Zaidi SIM, Panchal HB, Khan AA, Murtaza G, Rahman ZU, Baloch ZQ, Mungo N, Ramu V (2017) Outcomes in quadripolar vs bipolar left ventricular leads in cardiac resynchronization therapy (CRT): a meta-analysis of seven studies. Cardiology (Switzerland) 137:245–349. https://doi.org/10.1159/000477751

    Article  Google Scholar 

  52. Rao P, Faddis M (2017) Cardiac resynchronisation therapy: current indications, management and basic troubleshooting. Heart 103(24):2000–2007. https://doi.org/10.1136/heartjnl-2016-310656

    Article  PubMed  Google Scholar 

  53. Zhang B, Guo J, Zhang G (2017) Comparison of triple-site ventricular pacing versus conventional cardiac resynchronization therapy in patients with systolic heart failure: a meta-analysis of randomized and observational studies. J Arrhythm 34(1):55–64. https://doi.org/10.1002/joa3.12018

    Article  PubMed  PubMed Central  Google Scholar 

  54. Bordachar P, Gras D, Clementy N, Defaye P, Mondoly P, Boveda S, Anselme F, Klug D, Piot O, Sadoul N, Babuty D, Leclercq C (2017) Clinical impact of an additional left ventricular lead in cardiac resynchronization nonresponders: the V(3) trial. Heart Rhythm 15:870–876. https://doi.org/10.1016/j.hrthm.2017.12.028

    Article  PubMed  Google Scholar 

  55. Leclercq C, Lercher P, Lunati M, Rordorf R, Landolina M, Pappone C, Mangual J, McSpadden L, Badie N, Singh J (2018) Influence of etiology, QRS duration, and baseline systolic function on longterm CRT response rate with multipoint pacing: a multicenter experience. Europace 20:i215. https://doi.org/10.1093/europace/euy015

    Article  Google Scholar 

  56. Zegard A, Qiu T, Acquaye E, Marshall H, Leyva F (2017) Cardiac resynchronization therapy using quadripolar versus non-quadripolar left ventricular leads programmed to single-site left ventricular pacing impact on survival and heart failure hospitalization. Europace 19:iii120

    Article  Google Scholar 

  57. Tomassoni GF, Baker J, Corbisiero R, Love CJ, Martin DT, Sheppard RC, Worley SJ, Lee K, Lee LY, Cohorn C, Niazi I (2017) Impact of multipoint™ pacing (MPP) programming on clinical outcomes with a quadripolar CRT-D system in the MPP IDE study. Heart Rhythm 14(5):S38

    Google Scholar 

Download references

Funding

This work was supported by the National Key R&D Program of China (2017YFC1307800).

Author information

Authors and Affiliations

Authors

Corresponding authors

Correspondence to Yan Yao or Yu Jiang.

Ethics declarations

Conflict of interest

All authors declare that there is no conflict of interests.

Electronic supplementary material

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hu, F., Zheng, L., Ding, L. et al. Clinical outcome of left ventricular multipoint pacing versus conventional biventricular pacing in cardiac resynchronization therapy: a systematic review and meta-analysis. Heart Fail Rev 23, 927–934 (2018). https://doi.org/10.1007/s10741-018-9737-5

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10741-018-9737-5

Keywords

Navigation