Liver injury diagnosed on computed tomography after use of an automated cardiopulmonary resuscitation device

  • Case Report
  • Published:
Emergency Radiology Aims and scope Submit manuscript

Abstract

We report a case of an 89-year-old female with active extravasation and hemoperitoneum from a liver laceration demonstrated on multidetector computed tomography (CT), attributed to the use of an automated mechanical cardiopulmonary resuscitation (CPR) device. Although iatrogenic internal injuries related to manual CPR and CPR devices have previously been reported [1, 2], there has been no reported CT evidence of liver injury related to automated CPR devices to the authors’ knowledge. Imaging findings of complications related to the use of automated CPR devices are important to recognize and also help explain the possible mechanisms of injury. Liver injuries with active bleeding following CPR may have devastating consequences related to hemodynamic instability and may have an increased incidence when CPR is performed using an automated chest compression device.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1

References

  1. Wind J, Bekkers SCAM, van Hooren LJH (2009) Extensive injury after use of a mechanical cardiopulmonary resuscitation device. Amer J Emerg Med 27:1017.e1–1017.e2

    Article  CAS  Google Scholar 

  2. de Weerd Y, Kraaier K, Logtenberg M et al (2009) Successful bystander cardiopulmonary resuscitation complicated by liver rupture. Imaging Cardiol 17:33–34

    Google Scholar 

  3. Duchateau F-X, Gueye P, Curac S et al (2010) Effect of the AutoPulse automated band chest compression device on hemodynamics in out-of-hospital cardiac arrest resuscitation. Intensive Care Med 36:1256–1260

    Article  PubMed  Google Scholar 

  4. Timerman S, Cardoso LF, Ramires JA et al (2004) Improved hemodynamic performance with a novel chest compression device during treatment of in-hospital cardiac arrest. Resuscitation 61:273–280

    Article  PubMed  Google Scholar 

  5. Ong MEH, Ornaat JP, Edwards DP et al (2006) Use of an automated, load-distributing band chest compression device for out-of hospital cardiac arrest resuscitation. JAMA 295:2629–2637

    Article  PubMed  CAS  Google Scholar 

  6. Hallstrom A, Rea T, Sayre M et al (2006) Manual chest compression vs use of an automated chest compression device during resuscitation following out-of-hospital cardiac arrest. JAMA 295:2620–2628

    Article  PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jeremy R. Camden.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Camden, J.R., Carucci, L.R. Liver injury diagnosed on computed tomography after use of an automated cardiopulmonary resuscitation device. Emerg Radiol 18, 429–431 (2011). https://doi.org/10.1007/s10140-011-0949-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10140-011-0949-4

Keywords

Navigation