Skip to main content
Top
Gepubliceerd in: Netherlands Heart Journal 6/2011

01-06-2011 | Imaging in Cardiology

Hazardous twiddling of a subpectoral internal cardioverter defibrillator

Auteurs: A. J. H. H. M. van Oostrom, L. V. Boersma

Gepubliceerd in: Netherlands Heart Journal | Uitgave 6/2011

Log in om toegang te krijgen
share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail

Excerpt

A 72-year-old male presented with multiple inappropriate shocks 4 weeks post-implantation of a subpectoral dual chamber internal cardioverter defibrillator (ICD) with screw-in leads that were secured with sutures to the pectoral fascia. Chest X-ray showed retraction of both leads and lead-clustering nearby the can (Fig. 1, left panel) caused by a twiddler’s syndrome. Device interrogation demonstrated atrial fibrillation on both leads and loss of capture with high impedance of the ventricular lead. The patient underwent ICD and lead extraction showing extensive lead twisting (Fig. 1, right panel). The main causes of this rare and potentially fatal syndrome are device manipulation by the patient and oversized subcutaneous pockets [1, 2]. Our patient denied the former and is a rare case of a subpectoral-twiddler’s syndrome. Other causes are obesity, mental disorders and advanced age. Symptoms are related to device dysfunction, including stimulation of extra-cardiac muscles. Recurrence may be prevented by thorough information, limitation of the pocket size and suturing of the can.
Literatuur
1.
go back to reference Gasparini M, Regoli F, Ceriotti C, et al. Hiccups and dysphonic metallic voice, a unique presentation of Twiddler Syndrome. Circulation. 2006;114:e534–5.PubMedCrossRef Gasparini M, Regoli F, Ceriotti C, et al. Hiccups and dysphonic metallic voice, a unique presentation of Twiddler Syndrome. Circulation. 2006;114:e534–5.PubMedCrossRef
2.
go back to reference De Buitleir M, Canver CC. Twiddler’s syndrome complicating a transvenous defibrillator lead system. Chest. 1996;109:1391–4.PubMedCrossRef De Buitleir M, Canver CC. Twiddler’s syndrome complicating a transvenous defibrillator lead system. Chest. 1996;109:1391–4.PubMedCrossRef
Metagegevens
Titel
Hazardous twiddling of a subpectoral internal cardioverter defibrillator
Auteurs
A. J. H. H. M. van Oostrom
L. V. Boersma
Publicatiedatum
01-06-2011
Uitgeverij
Bohn Stafleu van Loghum
Gepubliceerd in
Netherlands Heart Journal / Uitgave 6/2011
Print ISSN: 1568-5888
Elektronisch ISSN: 1876-6250
DOI
https://doi.org/10.1007/s12471-011-0131-3

Andere artikelen Uitgave 6/2011

Netherlands Heart Journal 6/2011 Naar de uitgave

Rhythm Puzzle - Question

A simple broad complex tachycardia?