Failure of epicardial pacing leads in congenital heart disease: not uncommon and difficult to predict

Auteurs: M. C. Post, W. Budts, A. Van de Bruaene, R. Willems, B. Meyns, F. Rega, M. Gewillig

Gepubliceerd in: Netherlands Heart Journal | Uitgave 7-8/2011

Log in om toegang te krijgen
share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail

Abstract

Aims

We evaluate the incidence of epicardial lead failure and try to identify risk factors in patients with congenital heart disease.

Methods

All patients with a congenital heart defect and an epicardial pacing system, implanted within a timeframe of 25 years, were included in this study. Patients’ medical records and lead data were reviewed. Lead failure was defined as the primary endpoint.

Results

In total 198 active epicardial leads (atrial 40, ventricular 158) were implanted in 93 patients (median age at implantation 4.4 years (range 0–58.6)). During a total follow-up of 1235 lead-years, 29 lead failures (14.6%, 4 atrial, 25 ventricular) were documented in 22 patients (23.7%). Lead failure occurred at a median time period of 4.8 years (range 1.2–24.1) after implantation. Five-year freedom of lead failure was 88%. The only independent predictor for lead failure was the age at implantation (HR 0.44; 95%CI 0.20–0.97, p = 0.04), other characteristics failed to predict lead failure. Sudden cardiac death occurred in four patients (4.3%), in one a lead failure was documented.

Conclusion

A high incidence of epicardial lead failures is found in patients with congenital heart disease. Unfortunately, it is difficult to predict this potentially life-threatening complication.
Literatuur
1.
go back to reference Fortescue EB, Berul CI, Cecchin F, et al. Comparison of modern steroid-eluting epicardial and thin transvenous pacemaker leads in pediatric and congenital heart disease patients. J Interv Card Electrophysiol. 2005;14:27–36.PubMedCrossRef Fortescue EB, Berul CI, Cecchin F, et al. Comparison of modern steroid-eluting epicardial and thin transvenous pacemaker leads in pediatric and congenital heart disease patients. J Interv Card Electrophysiol. 2005;14:27–36.PubMedCrossRef
2.
go back to reference Beaufort-Krol GC, Mulder H, Nagelkerke D, et al. Comparison of longevity, pacing, and sensing characteristics of steroid-eluting epicardial versus conventional endocardial pacing leads in children. J Thorac Cardiovasc Surg. 1999;117:523–8.PubMedCrossRef Beaufort-Krol GC, Mulder H, Nagelkerke D, et al. Comparison of longevity, pacing, and sensing characteristics of steroid-eluting epicardial versus conventional endocardial pacing leads in children. J Thorac Cardiovasc Surg. 1999;117:523–8.PubMedCrossRef
3.
go back to reference Ector B, Willems R, Heidbuchel H, et al. Epicardial pacing: a single-centre study on 321 leads in 138 patients. Acta Cardiol. 2006;61:343–51.PubMedCrossRef Ector B, Willems R, Heidbuchel H, et al. Epicardial pacing: a single-centre study on 321 leads in 138 patients. Acta Cardiol. 2006;61:343–51.PubMedCrossRef
4.
go back to reference Cohen MI, Bush DM, Vetter VL, et al. Permanent epicardial pacing in pediatric patients: seventeen years of experience and 1200 outpatient visits. Circulation. 2001;103:2585–90.PubMed Cohen MI, Bush DM, Vetter VL, et al. Permanent epicardial pacing in pediatric patients: seventeen years of experience and 1200 outpatient visits. Circulation. 2001;103:2585–90.PubMed
5.
go back to reference Silvetti MS, Drago F, Grutter G, et al. Twenty years of paediatric cardiac pacing: 515 pacemakers and 480 leads implanted in 292 patients. Europace. 2006;8:530–6.PubMedCrossRef Silvetti MS, Drago F, Grutter G, et al. Twenty years of paediatric cardiac pacing: 515 pacemakers and 480 leads implanted in 292 patients. Europace. 2006;8:530–6.PubMedCrossRef
6.
go back to reference Murayama H, Maeda M, Sakurai H, et al. Predictors affecting durability of epicardial pacemaker leads in pediatric patients. J Thorac Cardiovasc Surg. 2008;135:361–6.PubMedCrossRef Murayama H, Maeda M, Sakurai H, et al. Predictors affecting durability of epicardial pacemaker leads in pediatric patients. J Thorac Cardiovasc Surg. 2008;135:361–6.PubMedCrossRef
7.
go back to reference ten Udink CF, Breur J, Boramanand N, et al. Endocardial and epicardial steroid lead pacing in the neonatal and paediatric age group. Heart. 2002;88:392–6.CrossRef ten Udink CF, Breur J, Boramanand N, et al. Endocardial and epicardial steroid lead pacing in the neonatal and paediatric age group. Heart. 2002;88:392–6.CrossRef
8.
go back to reference Tomaske M, Gerritse B, Kretzers L, et al. A 12-year experience of bipolar steroid-eluting epicardial pacing leads in children. Ann Thorac Surg. 2008;85:1704–11.PubMedCrossRef Tomaske M, Gerritse B, Kretzers L, et al. A 12-year experience of bipolar steroid-eluting epicardial pacing leads in children. Ann Thorac Surg. 2008;85:1704–11.PubMedCrossRef
9.
go back to reference Sachweh JS, Vazquez-Jimenez JF, Schondube FA, et al. Twenty years experience with pediatric pacing: epicardial and transvenous stimulation. Eur J Cardiothorac Surg. 2000;17:455–61.PubMedCrossRef Sachweh JS, Vazquez-Jimenez JF, Schondube FA, et al. Twenty years experience with pediatric pacing: epicardial and transvenous stimulation. Eur J Cardiothorac Surg. 2000;17:455–61.PubMedCrossRef
10.
go back to reference Horenstein MS, Hakimi M, Walters III H, et al. Chronic performance of steroid-eluting epicardial leads in a growing pediatric population: a 10-year comparison. Pacing Clin Electrophysiol. 2003;26:1467–71.PubMedCrossRef Horenstein MS, Hakimi M, Walters III H, et al. Chronic performance of steroid-eluting epicardial leads in a growing pediatric population: a 10-year comparison. Pacing Clin Electrophysiol. 2003;26:1467–71.PubMedCrossRef
11.
go back to reference Fortescue EB, Berul CI, Cecchin F, et al. Patient, procedural, and hardware factors associated with pacemaker lead failures in pediatrics and congenital heart disease. Heart Rhythm. 2004;1:150–9.PubMedCrossRef Fortescue EB, Berul CI, Cecchin F, et al. Patient, procedural, and hardware factors associated with pacemaker lead failures in pediatrics and congenital heart disease. Heart Rhythm. 2004;1:150–9.PubMedCrossRef
12.
go back to reference Cutler NG, Karpawich PP, Cavitt D, et al. Steroid-eluting epicardial pacing electrodes: six year experience of pacing thresholds in a growing pediatric population. Pacing Clin Electrophysiol. 1997;20:2943–8.PubMedCrossRef Cutler NG, Karpawich PP, Cavitt D, et al. Steroid-eluting epicardial pacing electrodes: six year experience of pacing thresholds in a growing pediatric population. Pacing Clin Electrophysiol. 1997;20:2943–8.PubMedCrossRef
13.
go back to reference Crossley GH, Brinker JA, Reynolds D, et al. Steroid elution improves the stimulation threshold in an active-fixation atrial permanent pacing lead. A randomized, controlled study. Model 4068 Investigators. Circulation. 1995;92:2935–9.PubMed Crossley GH, Brinker JA, Reynolds D, et al. Steroid elution improves the stimulation threshold in an active-fixation atrial permanent pacing lead. A randomized, controlled study. Model 4068 Investigators. Circulation. 1995;92:2935–9.PubMed
14.
go back to reference Johns JA, Fish FA, Burger JD, et al. Steroid-eluting epicardial pacing leads in pediatric patients: encouraging early results. J Am Coll Cardiol. 1992;20:395–401.PubMedCrossRef Johns JA, Fish FA, Burger JD, et al. Steroid-eluting epicardial pacing leads in pediatric patients: encouraging early results. J Am Coll Cardiol. 1992;20:395–401.PubMedCrossRef
Metagegevens
Titel
Failure of epicardial pacing leads in congenital heart disease: not uncommon and difficult to predict
Auteurs
M. C. Post
W. Budts
A. Van de Bruaene
R. Willems
B. Meyns
F. Rega
M. Gewillig
Publicatiedatum
01-08-2011
Uitgeverij
Bohn Stafleu van Loghum
Gepubliceerd in
Netherlands Heart Journal / Uitgave 7-8/2011
Print ISSN: 1568-5888
Elektronisch ISSN: 1876-6250
DOI
https://doi.org/10.1007/s12471-011-0158-5

Andere artikelen Uitgave 7-8/2011

Netherlands Heart Journal 7-8/2011 Naar de uitgave

Rhythm Puzzle - Answer

A young man with near-syncope