Comments on: ‘Randomised comparison of a balloon-expandable and self-expandable valve with quantitative assessment of aortic regurgitation using magnetic resonance imaging’

Auteurs: O. Soliman, M. Abdel-Wahab, P. Serruys

Gepubliceerd in: Netherlands Heart Journal | Uitgave 10/2020

share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail
insite
ZOEKEN
We read with interest the paper by Kooistra et al. [1], who reported on the superiority of a balloon-expandable valve (BEV, SAPIEN 3) over a self-expandable valve (SEV, CoreValve) in terms of aortic regurgitation (AR) and clinical outcome up to 1 year after transcatheter aortic valve implantation (TAVI) in a randomised comparison with quantitative AR assessment using magnetic resonance imaging (MRI).
There are several issues that might have led to bias in the study findings. First, MRI assessment of AR has some limitations, including higher cost, restricted availability, and limited feasibility due to the presence of pacemakers and device implants, compared to other techniques such as quantitative aortography with videodensitometric assessment. The latter has been validated in vitro and in vivo, proven accurate compared to MRI [2] and echocardiography, and is feasible in routine practice in almost every patient [3]. In contrast to MRI, quantitative aortography is readily available as an online tool in the catheterisation laboratory [3]. Comparison of several transcatheter heart valves (THVs) has shown a clear advantage of almost all new-generation THVs compared with an early-generation valve such as the CoreValve [4].
A second issue is the inherent selection bias due to exclusion of 22% of the cases from the final cohort, who were unable to undergo MRI examination in the presence of a pacemaker (PM). Patients with an SEV require PM implantation more often than those with a BEV. Furthermore, patients developing conduction disturbances often have deep implants, which might affect the occurrence and severity of paravalvular leakage (PVL) with both device types. Consequently, a balance between PM need and PVL would be an appropriate objective of a head-to-head device trial. Again, aortography allows assessment of PVL and of depth of implantation, simultaneously. Therefore, aortography could be a better option than MRI for device comparisons in this context.
Furthermore, atrial fibrillation, which was reported in almost one-third of the patients in this paper, reduces the accuracy of flow measurements on MRI. Thus, both phase contrast quantification of AR as well as volumetric quantification could be imprecise.
Most importantly, the choice of an earlier-generation SEV that is known to be inferior to the novel-generation BEV in terms of PVL and PM need represents an a priori advantageous bias toward the newer-generation device. A recent publication by Modolo et al. on 2258 patients with seven different devices, as well as other earlier publications, show the highest incidence of PVL after early-generation SEV [4]. Thus, it would be beneficial to compare contemporary device generations such as the Evolut series versus the SAPIEN 3 series. Unfortunately, the authors stopped enrolment in this study upon availability of the CoreValve Evolut series, an SEV with an improved design and better sealing properties against PVL [4].
Despite our critical comments, we must commend the authors for exploring ways to promote objective device head-to-head comparisons in the field of TAVI and for furthering the discussion about the effects of AR on outcome. However, we maintain our call for the use of quantitative aortography, a fully validated technique for AR assessment, as it essentially eliminates unnecessary costs and overcomes the limited availability and/or feasibility of MRI in daily patient care.

Conflict of interest

O. Soliman, M. Abdel-Wahab and P. Serruys declare that they have no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.
share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail

Onze productaanbevelingen

Netherlands Heart Journal

Het Netherlands Heart Journal wordt uitgegeven in samenwerking met de Nederlandse Vereniging voor Cardiologie en de Nederlandse Hartstichting. Het tijdschrift is Engelstalig en wordt gratis beschikbaa ...

Literatuur
1.
go back to reference Kooistra NHM, Abawi M, Voskuil M, Urgel K, Samim M, Nijhoff F, et al. Randomised comparison of a balloon-expandable and self-expandable valve with quantitative assessment of aortic regurgitation using magnetic resonance imaging. Neth Heart J. 2020;28(5):253–65.PubMedPubMedCentral Kooistra NHM, Abawi M, Voskuil M, Urgel K, Samim M, Nijhoff F, et al. Randomised comparison of a balloon-expandable and self-expandable valve with quantitative assessment of aortic regurgitation using magnetic resonance imaging. Neth Heart J. 2020;28(5):253–65.PubMedPubMedCentral
2.
go back to reference Abdel-Wahab M, Abdelghani M, Miyazaki Y, Holy EW, Merten C, Zachow D, et al. A novel angiographic quantification of aortic regurgitation after TAVR provides an accurate estimation of regurgitation fraction derived from cardiac magnetic resonance imaging. JACC Cardiovasc Interv. 2018;11(3):287–97.PubMed Abdel-Wahab M, Abdelghani M, Miyazaki Y, Holy EW, Merten C, Zachow D, et al. A novel angiographic quantification of aortic regurgitation after TAVR provides an accurate estimation of regurgitation fraction derived from cardiac magnetic resonance imaging. JACC Cardiovasc Interv. 2018;11(3):287–97.PubMed
4.
go back to reference Modolo R, Chang CC, Abdelghani M, Kawashima H, Ono M, Tateishi H, et al. Quantitative assessment of acute regurgitation following transcatheter aortic valve replacement: a multicenter pooled analysis of 2,258 valves. JACC Cardiovasc Interv. 2020;13(11):1303–11.PubMed Modolo R, Chang CC, Abdelghani M, Kawashima H, Ono M, Tateishi H, et al. Quantitative assessment of acute regurgitation following transcatheter aortic valve replacement: a multicenter pooled analysis of 2,258 valves. JACC Cardiovasc Interv. 2020;13(11):1303–11.PubMed
Metagegevens
Titel
Comments on: ‘Randomised comparison of a balloon-expandable and self-expandable valve with quantitative assessment of aortic regurgitation using magnetic resonance imaging’
Auteurs
O. Soliman
M. Abdel-Wahab
P. Serruys
Publicatiedatum
25-08-2020
Uitgeverij
Bohn Stafleu van Loghum
Gepubliceerd in
Netherlands Heart Journal / Uitgave 10/2020
Print ISSN: 1568-5888
Elektronisch ISSN: 1876-6250
DOI
https://doi.org/10.1007/s12471-020-01484-0

Andere artikelen Uitgave 10/2020

Netherlands Heart Journal 10/2020 Naar de uitgave