Beyond catheter tip and radiofrequency lesion delivery: the role of robotics in ablation of ventricular tachycardia

Auteurs: L.J. de Vries, F. Zijlstra, T. Szili-Torok

Gepubliceerd in: Netherlands Heart Journal | Uitgave 10/2015

share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail
insite
ZOEKEN
Catheter ablation of ventricular tachycardia (VT) has become an important and increasingly performed treatment [1]. The use of remote magnetic navigation (RMN) for this procedure has been shown to have several advantages compared with manual catheter navigation (MCN) [2]. Since RMN was introduced in 2003 the most obvious advantage of RMN when compared with MCN is safety. Increased safety for both operators and patients is achieved by shortening fluoroscopy time and by reducing complications due to catheter flexibility [3]. So far, there has not been a single report of myocardial perforation. Additionally, after an initial learning curve has been completed, the use of RMN most likely shortens procedure time [4], may prevent operator fatigue and helps to reach the parts of the heart that are less easily accessible during mapping.
In the current issue of the Netherlands Heart Journal, Wu et al. [5] present their meta-analysis on RMN vs. MCN for the ablation of VT. They included four non-randomised studies and conclude that acute and long-term success rates for VT ablation are equal between RMN and MCN. However, complication rates, procedure and fluoroscopy times are favourable for RMN.
Their conclusion adds to the aforementioned findings for improved safety and at least a non-inferior success rate. In fact, the only limitation of RMN that remains in experienced teams is the cost of the equipment which can be up to 2 million € [6]. However, the long-term benefits of this installation should be taken into account when assessing these expenses [6].
Looking at success rate, the authors have been cautious in their interpretation. Several studies have in fact demonstrated a better outcome compared with MCN in VT ablation [4, 7]. What could be a possible explanation for this variable outcome?
As RMN is a relatively new and evolving method, procedural outcomes can be further improved when sufficient experience is obtained. Only a limited group of operators could acquire the appropriate level of skill due to the cost of the system and a consequently limited availability. This could potentially result in an operator-dependent success rate variability and also brings up another point related to the financial aspect. Most of the centres able to afford the installation are academic centres, dealing with a very specific patient population. This means the procedures performed with RMN will be those in patients with a more complex pathology, which in turn may result in an underestimation of treatment success. Finally, because of the limited availability of RMN technology, present studies are mostly non-randomised. Non-significant trends in success rates favourable to RMN, as seen in Dinov et al. [8], may in the future prove significant in randomised trials.
Notably, the papers by Bauernfeind et al. and Szili-Torok et al., both used in the current report, showed superiority to MCN in acute success rates in structurally normal hearts but equal results in structural heart disease [4, 7]. This raises the issue of how RMN can improve success rates. Is it related to the improved manoeuvrability of the catheter tip: tip delivery, or due to the constant type of tissue tip contact which improves radiofrequency lesion formation: lesion delivery? The data from of the above-mentioned manuscripts as well as this meta-analysis indirectly suggest that tip delivery efficiency may be superior using RMN while lesion delivery should be equally good when compared with MCN. RMN is theoretically more suitable for reaching difficult locations because of the flexible nature of the catheter in combination with the possibility of more accurate positioning.
What are the future perspectives? An important step that could definitively make RMN favourable to MCN will be the utilisation of contact force measuring ablation catheters for magnetic navigation and ablation. Optimal electrode—tissue contact has been previously demonstrated to be of great importance for lesion formation [9] and has already been shown to improve clinical outcome in the treatment of atrial fibrillation [10]. When this technique becomes available for RMN and reaches its full potential (new combined parameters, involvement of cardiac imaging for real time visualisation of lesion formation) it might revolutionise treatment of ventricular and supraventricular arrhythmias with RMN and might prove pivotal in achieving long-term success in VT ablation.
Funding
None.
Conflict of interests
None declared.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
share
DELEN

Deel dit onderdeel of sectie (kopieer de link)

  • Optie A:
    Klik op de rechtermuisknop op de link en selecteer de optie “linkadres kopiëren”
  • Optie B:
    Deel de link per e-mail

Onze productaanbevelingen

Netherlands Heart Journal

Het Netherlands Heart Journal wordt uitgegeven in samenwerking met de Nederlandse Vereniging voor Cardiologie en de Nederlandse Hartstichting. Het tijdschrift is Engelstalig en wordt gratis beschikbaa ...

Literatuur
1.
go back to reference Aliot EM, Stevenson WG, Almendral-Garrote JM, et al. EHRA/HRS Expert Consensus on Catheter Ablation of Ventricular Arrhythmias: developed in a partnership with the European Heart Rhythm Association (EHRA), a Registered Branch of the European Society of Cardiology (ESC), and the Heart Rhythm Society (HRS); in collaboration with the American College of Cardiology (ACC) and the American Heart Association (AHA). Heart Rhythm. 2009;6:886–933.CrossRefPubMed Aliot EM, Stevenson WG, Almendral-Garrote JM, et al. EHRA/HRS Expert Consensus on Catheter Ablation of Ventricular Arrhythmias: developed in a partnership with the European Heart Rhythm Association (EHRA), a Registered Branch of the European Society of Cardiology (ESC), and the Heart Rhythm Society (HRS); in collaboration with the American College of Cardiology (ACC) and the American Heart Association (AHA). Heart Rhythm. 2009;6:886–933.CrossRefPubMed
2.
go back to reference Parreira L, Cavaco D, Reis-Santos K, et al. Remote magnetic navigation for mapping and ablation of right and left ventricular outflow tract arrhythmias. Rev Port Cardiol. 2013;32:489–95.PubMed Parreira L, Cavaco D, Reis-Santos K, et al. Remote magnetic navigation for mapping and ablation of right and left ventricular outflow tract arrhythmias. Rev Port Cardiol. 2013;32:489–95.PubMed
3.
go back to reference Proietti R, Pecoraro V, Di Biase L, et al. Remote magnetic with open-irrigated catheter vs. manual navigation for ablation of atrial fibrillation: a systematic review and meta-analysis. Europace. 2013;15:1241–48.CrossRefPubMed Proietti R, Pecoraro V, Di Biase L, et al. Remote magnetic with open-irrigated catheter vs. manual navigation for ablation of atrial fibrillation: a systematic review and meta-analysis. Europace. 2013;15:1241–48.CrossRefPubMed
4.
go back to reference Bauernfeind T, Akca F, Schwagten B, et al. The magnetic navigation system allows safety and high efficacy for ablation of arrhythmias. Europace. 2011;13:1015–21.CrossRefPubMedPubMedCentral Bauernfeind T, Akca F, Schwagten B, et al. The magnetic navigation system allows safety and high efficacy for ablation of arrhythmias. Europace. 2011;13:1015–21.CrossRefPubMedPubMedCentral
5.
go back to reference Wu Y, Wang W, Li K, et al. Remote magnetic navigation vs. manual navigation for ablation of ventricular tachycardia: a meta-analysis. Neth Heart J. 2015;23. doi:10.1007/s12471-015-0734-1. Wu Y, Wang W, Li K, et al. Remote magnetic navigation vs. manual navigation for ablation of ventricular tachycardia: a meta-analysis. Neth Heart J. 2015;23. doi:10.1007/s12471-015-0734-1.
6.
go back to reference Da Costa A, Lafond P, Romeyer-Bouchard C, et al. Remote magnetic navigation and arrhythmia ablation. Arch Cardiovasc Dis. 2012;105:446–53.CrossRefPubMed Da Costa A, Lafond P, Romeyer-Bouchard C, et al. Remote magnetic navigation and arrhythmia ablation. Arch Cardiovasc Dis. 2012;105:446–53.CrossRefPubMed
7.
go back to reference Szili-Torok T, Swagten B, Akca F, et al. Catheter ablation of ventricular tachycardias using remote magnetic navigation: a consecutive case-control study. J Cardiovasc Electrophysiol. 2012;23:948–54.CrossRefPubMed Szili-Torok T, Swagten B, Akca F, et al. Catheter ablation of ventricular tachycardias using remote magnetic navigation: a consecutive case-control study. J Cardiovasc Electrophysiol. 2012;23:948–54.CrossRefPubMed
8.
go back to reference Dinov B, Schönbauer R, Wojdyla-Hordynska A, et al. Long-term efficacy of single procedure remote magnetic catheter navigation for ablation of ischemic ventricular tachycardia: a retrospective study. J Cardiovasc Electrophysiol. 2012;23:499–505.CrossRefPubMed Dinov B, Schönbauer R, Wojdyla-Hordynska A, et al. Long-term efficacy of single procedure remote magnetic catheter navigation for ablation of ischemic ventricular tachycardia: a retrospective study. J Cardiovasc Electrophysiol. 2012;23:499–505.CrossRefPubMed
9.
go back to reference Yokoyama K, Nakagawa H, Shah DC, et al. Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ Arrhythm Electrophysiol. 2008;1:354–62.CrossRefPubMed Yokoyama K, Nakagawa H, Shah DC, et al. Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ Arrhythm Electrophysiol. 2008;1:354–62.CrossRefPubMed
10.
go back to reference Neuzil P, Reddy VY, Kautzner J, et al. Electrical reconnection after pulmonary vein isolation is contingent on contact force during initial treatment results from the EFFICAS I study. Circ Arrhythm Electrophysiol. 2013;6:327–33.CrossRefPubMed Neuzil P, Reddy VY, Kautzner J, et al. Electrical reconnection after pulmonary vein isolation is contingent on contact force during initial treatment results from the EFFICAS I study. Circ Arrhythm Electrophysiol. 2013;6:327–33.CrossRefPubMed
Metagegevens
Titel
Beyond catheter tip and radiofrequency lesion delivery: the role of robotics in ablation of ventricular tachycardia
Auteurs
L.J. de Vries
F. Zijlstra
T. Szili-Torok
Publicatiedatum
04-08-2015
Uitgeverij
Bohn Stafleu van Loghum
Gepubliceerd in
Netherlands Heart Journal / Uitgave 10/2015
Print ISSN: 1568-5888
Elektronisch ISSN: 1876-6250
DOI
https://doi.org/10.1007/s12471-015-0737-y

Andere artikelen Uitgave 10/2015

Netherlands Heart Journal 10/2015 Naar de uitgave